Recommendation-CIPM-1992-3

Recommendation 3 of the 81st CIPM (1992)

Revision of the mise en pratique of the definition of the metre

The Comité International des Poids et Mesures,

recalling

  • that in 1983 the 17th Conférence Générale des Poids et Mesures (CGPM) adopted a new definition of the metre;
  • that in the same year the CGPM invited the Comité International des Poids et Mesures (CIPM)
    • to draw up instructions for the practical realization of the metre;
    • to choose radiations which can be recommended as standards of wave1ength for the interferometric measurement of length and draw up instructions for their use;
    • to pursue studies undertaken to improve these standards and in due course to extend or revise these instructions;
  • that in response to this invitation CIPM made a number of Recommendations in 1983 concerning the practical realization of the metre (the 'mise en pratique');

considering

  • that science and technology continue to demand improved accuracy in the realization of the metre;
  • that since 1983 work in nationallaboratories, BIPM and elsewhere has substantially improved the reproducibility of radiations which are suitable for the practical realization of the metre;
  • that such work has also substantially reduced the uncertainty in the determined values of the frequencies and wavelengths of some of these radiations;

decides that the list of recommended radiations given by the CIPM in 1983 (Recommendation 1 (CI-1983)) be replaced by the list of recommended radiations given below.

 

LIST OF RECOMMENDED RADIATIONS, 1992

 

This list replaces the one published in BIPM Proc.-Verb. Corn. Int. Poids et Mesures, 1983,51, 25-28 and Metrologia, 1984, 19, 165-166.

In this list, the values of the frequency f and of the wavelength λ should be related exactly by the relation λ f = c, with c = 299 792 458 m/s but the values of λ are rounded.

The data and analysis used for the compilation of this list are set out in the associated Appendix: Source Data for the List of Recommended Radiations, 1992 and its Annotated Bibliography*.

It should be noted that for several of the listed radiations, few independent values are available, so that the estimated uncertainties may not, therefore, reflect all sources of variability.

Each of the listed radiations can be replaced, without degrading the accuracy, by a radiation corresponding to another component of the same transition or by another radiation, when the frequency difference is known with sufficient accuracy. It should be also noted that to achieve the uncertainties given here it is not sufficient just to meet the specifications for the listed parameters. In addition, it is necessary to follow the best good practice concerning methods of stabilization as described in numerous scientific and technical publications. References to appropriate articles, illustrating accepted good practice for a particular radiation, may be obtained by application to a member laboratory of the CCDM, or to the BIPM.

1. Radiations of Stabilized Lasers

1.1. Absorbing molecule CH4, transition V3, P (7), component F2(2).

1.1.1. The values
f = 88 376 181 600,18 kHz
λ = 3 392 231 397,327 fm
with an estimated relative standard uncertainty of 3 × 10–12 apply to the radiation of a He-Ne laser stabilized to the central component [(7-6) transition] of the resolved hyperfine-structure triplet, the mean of recoil splitting, for effectively stationary molecules, i.e. the values are corrected for second order Doppler shift.

1.1.2. The values
f = 88 376 181 600,5 kHz
λ = 3 392 231 397,31 m
with an estimated relative standard uncertainty of 2,3 × 10–11 apply to the radiation of a He-Ne laser stabilized to the centre of the unresolved hyperfine structure of a room temperature methane cell, within or external to the laser, subject to the following conditions:

  • methane pressure ≤ 3 Pa
  • mean one-way axial intracavity surface power density+ ≤ 104 W m-2
  • radius of wavefront curvature  ≥ 1 m
  • inequality of power between counter-propagating waves ≤ 5 %
  • detector placed at the output facing the laser tube.

* The appendix referred to above is published in the Report of the 8th meeting of the CCDM (1992).

1.2. Absorbing atom 40Ca, transition 3P1 - 1S0; ΔmJ = 0.

The values
f = 455 986 240,5 MHz
λ = 657 459 439,3 fm
with an estimated relative standard uncertainty of 4,5 × 10–10 apply to the radiation of a laser stabilized with a thermal atomic beam.

1.3. Absorbing molecule 127I2, transition 8-5, P(10), component a9 (or g).

The values
f = 468 218 332,4 MHz
λ = 640 283 468,7 fm
with an estimated relative standard uncertainty of 4,5 × 10–10 apply to the radiation of a He-Ne laser stabilized with an internal iodine cell having a cold finger temperature of (16 ± 1) °C and a frequency modulation width, peak to peak, of (6 ± 1) MHz.

1.4. Absorbing molecule 127I2, transition 11-5, R(127), component al3 (or i).

The values
f= 473 612 214 705 kHz
λ = 632 991 398,22 fm
with an estimated relative standard uncertainty of 2,5 × 10–11 apply to the radiation of a He-Ne laser with an internal iodine ceIl, subject to the conditions:

  • cell-wall temperature: (25 ± 5) °C
  • cold finger temperature: (15 ± 0,2) °C
  • frequency modulation width, peak to peak: (6 ± 0,3) MHz
  • one-way intracavity beam power+: (10 ± 5) mW, for an absolute value of the power shift coefficient ≤ 1,4 kHzlmW.

These conditions are by themselves insufficient to ensure that the stated standard uncertainty will be achieved. It is also necessary for the optical and electronic control systems to be operating with the appropriate technical performance. The iodine cell may also be operated under relaxed conditions, leading to the larger uncertainty specified in Appendix M2 of the CCDM Report (1992).

1.5. Absorbing molecule 127I2, transition 9-2, R(47), component a7 (or o).

The values
f = 489 880 354,9 MHz
λ = 611 970 770,0 fm
with an estimated relative standard uncertainty of 3 × 10–10 apply to the radiation of a He-Ne laser stabilized with an iodine ceIl, within or external to the laser, having a cold finger temperature of (- 5 ± 2) °C.

1.6. Absorbing molecule 127I2, transition 17-1, P(62), component a1.

The values
f = 520 206 808,4 MHz
λ = 576 294 760,4 fm
with an estimated relative standard uncertainty of 4 × 10–10 apply to the radiation of a dye laser (or frequency-doubled He-Ne laser) stabilized with an iodine ceIl, within or external to the laser, having a cold-finger temperature of (6 ± 2) °C.

1.7. Absorbing molecule 127I2, transition 26-0, R(12), component a9

The values
f = 551579482,96 MHz
λ = 543516333,1 fm
with an estimated relative standard uncertainty of 2,5 × 10–10 apply to the radiation of a frequency stabilized He-Ne laser with an external iodine cell having a cold-finger temperature of (0 ± 2) °c.

1.8. Absorbing molecule 127I2, transition 43-0, P(13), component a3 (or s).

The values
f = 582 490 603,37 MHz
λ = 514 673 466,4 fm
with an estimated relative standard uncertainty of 2,5 × 10–10 apply to the radiation of an Ar+ laser stabilized with an iodine cell external to the laser, having a cold-finger temperature of (- 5 ± 2) °C.

Note
+ The one-way intracavity beam power is obtained by dividing the output power by the transrnittance of the output mirror.

2. Radiations of Spectral Lamps

2.1. Radiation corresponding to the transition between the levels 2plO and 5d5 of the atom of 86Kr.

The value
λ = 605 780 210,3 fm
with an estimated overall relative uncertainty of ± 4 × 10–9 [equivalent to three times the relative standard uncertainty of 1,3 × 10–9] applies to the radiation emitted by a lamp operated under the conditions recommended by the CIPM (BlPM Proc.-Verb. Corn. Int. Poids et Mesures, 1960, 28, 71-72 and BlPM Comptes Rendus lle Conf. Gén. Poids et Mesures, 1960, 85)].

2.2. Radiations of atoms 86Kr, 198Hg and 114Cd recommended by the CIPM in 1963 (BIPM Corn. Cons. Déf. Mètre, 1962, 3, 18-19 and BIPM Proc. Verb. Com. Int. Poids et Mesures, 1963, 52, 26-27), with the indicated values for the wavelengths and the corresponding uncertainties.

DOI : 10.59161/CIPM1992REC3E

The reader should note that the official version of this Resolution is the French text