Results for: Laboratorio Central Oficial de Electrotecnica
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
16 results
Reset all
CMC Area
To filter, either use the tick boxes or click on the underscored item
Geographical location
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Spain, LCOE (Laboratorio Central Oficial de Electrotecnica)
High voltage and current , High AC voltage: RMS value : 1 kV to 500 kV
RMS voltage responding measuring system, AC voltmeter, AC voltage source
Relative expanded uncertainty : 5.0E2 µV/V
Comparison with reference system
Frequency : 50 Hz, 60 Hz
Approved on 07 February 2019
Institute service identifier : LCOE/62
CMC ID : EURAMET-EM-ES-00000E51-1
 
High voltage and current , High DC voltage: ratios : 1.00E-6 to 1
DC high voltage dividers, DC high voltage probes
Relative expanded uncertainty : 5.0E-5 to 1.0E-4
{"row_1":{"col_1":"Voltage (kV)","col_2":"Expanded uncertainty (dimensionless)"},"row_2":{"col_1":"1 kV","col_2":"50 E-6"},"row_3":{"col_1":"1 kV to 100 kV","col_2":"100E-6"}}
100E-06, (50E-06 for 1 kV level)
Comparison with reference divider
Voltage : 1 kV to 100 kV
Approved on 23 June 2022
Institute service identifier : LCOE/61 update
CMC ID : EURAMET-EM-ES-00000E50-2
 
High voltage and current , High AC voltage: peak values : 1 kV to 500 kV
Peak voltage responding measuring system, AC voltmeter, AC voltaqe source
Relative expanded uncertainty : 0.5 mV/V to 1 mV/V
{"row_1":{"col_1":"","col_2":"Expanded uncertainty (mV/V)"},"row_2":{"col_1":"1 kV to 200 kV","col_2":"0.5"},"row_3":{"col_1":"200 kV to 500 kV","col_2":"1"}}
Comparison with reference system
Frequency : 50 Hz, 60 Hz
Approved on 07 February 2019
Institute service identifier : LCOE/64
CMC ID : EURAMET-EM-ES-00000E52-2
 
High voltage and current , Impulse voltage and current: lightning impulse front time : 0.8 µs to 2 µs
High voltage divider, high voltage measuring system, EMC surge generator
Relative expanded uncertainty : 2 % to 3 %
{"row_1":{"col_1":"","col_2":"Expanded uncertainty"},"row_2":{"col_1":"1 kV to 600 kV","col_2":"2.0E-2"},"row_3":{"col_1":"600 kV to 700 kV","col_2":"3.0E-2"}}
Expanded uncertainty is expressed in per unit, it means 2% up to 600 kV and 3% from 600 kV to 700 kV
Comparison with high voltage reference measuring system
IEC 60060-2
Impulse shape : according to IEC 60060-1; 1 kV to 700 kV
Approved on 07 February 2019
Institute service identifier : LCOE/68
CMC ID : EURAMET-EM-ES-00000E55-3
 
High voltage and current , Impulse voltage and current: lightning impulse time to half value : 40 µs to 60 µs
High voltage divider, high voltage measuring system, EMC surge generator
Relative expanded uncertainty : 1 %
Comparison with high voltage reference measuring system
IEC 60060-2
Impulse shape : according to IEC 60060-1; 1 kV to 700 kV
Approved on 07 February 2019
Institute service identifier : LCOE/69
CMC ID : EURAMET-EM-ES-00000E56-3
 
High voltage and current , High DC voltage: average value : 1 kV to 200 kV
High DC voltage measuring systems, DC voltage sources, DC voltmeters
Relative expanded uncertainty : 5.0E1 µV/V to 1.0E2 µV/V
100, (50 for 1 kV level) The uncertainty is expressed in µV/V
Comparison with reference system
Approved on 07 February 2019
Institute service identifier : LCOE/59
CMC ID : EURAMET-EM-ES-00000E4Z-2
 
High voltage and current , Parameters for other impulse voltage types: other impulse measuring system : 1 kV to 100 kV
High voltage measuring system for Very Fast Transients: voltage scale factor
Relative expanded uncertainty : 1.1E1 mV/V
Comparison with high voltage reference measuring system
Time to steady state, ts, : 6 ns to 25 ns
Calibration of the scale factor of very fast transients measuring systems using a high voltage step (see figure 6 of AMD 1 of IEC 60270 for parameters definition)
Approved on 12 April 2021
Institute service identifier : 73
CMC ID : EURAMET-EM-ES-00000MHP-1
 
High voltage and current , Electric discharge: apparent charge : 0.2 pC to 1.00E5 pC
Partial discharge calibrator: apparent charge q
Relative expanded uncertainty : 1.5E1 mC/C
Comparison with digital integration
Approved on 12 April 2021
Institute service identifier : 76
CMC ID : EURAMET-EM-ES-00000MHS-1
 
High voltage and current , Parameters for other impulse voltage types: other impulse measuring system : 6 ns to 25 ns
High voltage measuring system for Very Fast Transients: time to steady state, ts
Absolute expanded uncertainty : 1.5 ns
Comparison with high voltage reference measuring system
Step voltage magnitude : 1 kV to 100 kV
Calibration of the time to steady state of very fast transients measuring systems using a high voltage step (see figure 6 of AMD 1 of IEC 60270 for parameters definition)
Approved on 12 April 2021
Institute service identifier : 74
CMC ID : EURAMET-EM-ES-00000MHQ-1
 
High voltage and current , Electric discharge: apparent charge : 0.1 pC to 0.2 pC
Partial discharge calibrator: apparent charge q
Relative expanded uncertainty : 5.0E1 mC/C
Comparison with analogue integrator
Approved on 12 April 2021
Institute service identifier : 75
CMC ID : EURAMET-EM-ES-00000MHR-1
 
High voltage and current , High AC voltage: voltage transformer ratio error : 0 to 0.02
Voltage transformer, voltage divider
Absolute expanded uncertainty : 8.0E-5 to 1.0E-4
{"row_1":{"col_1":"","col_2":"Expanded uncertainty"},"row_2":{"col_1":"1 kV to 60 kV","col_2":"80E-06"},"row_3":{"col_1":"60 kV to 500 kV","col_2":"100E-06"}}
The expanded uncertainties given in this table are relative to the measured value and thus dimensionless
Bridge
Primary voltage : 1 kV to 500 kV
Secondary voltage : 1 V to 500 V
Frequency : 50 Hz, 60 Hz
Approved on 23 June 2022
Institute service identifier : LCOE/65 update
CMC ID : EURAMET-EM-ES-00000E47-3
 
High voltage and current , High AC voltage: voltage transformer phase displacement : 0 crad to 2 crad
Voltage transformer, voltage divider
Absolute expanded uncertainty : 8.7E-3 crad to 1.2E-2 crad
{"row_1":{"col_1":"<br>","col_2":"1 kV to 60 kV","col_3":"60 kV to 500 kV"},"row_2":{"col_1":"1E-06 crad to 2 crad","col_2":"0.0087","col_3":"0.012"}}
The expanded uncertainties given in this table are expressed in crad
Bridge
Primary voltage : 1 kV to 500 kV
Secondary voltage : 1 V to 500 V
Frequency : 50 Hz, 60 Hz
Approved on 23 June 2022
Institute service identifier : LCOE/66 update
CMC ID : EURAMET-EM-ES-00000E4E-2
 
High voltage and current , High AC current: ratio error : 0 to 0.02
Current transformer
Absolute expanded uncertainty : 5.0E-5 to 1.0E-4
{"row_1":{"col_1":"% I<sub>R</sub>","col_2":"Expanded uncertainty"},"row_2":{"col_1":"1 % ≤ I<sub>R</sub> &lt; 20 %","col_2":"1.10<sup>-4</sup>"},"row_3":{"col_1":"20 % ≤ I<sub>R</sub> ≤ 200 %","col_2":"5.10<sup>-5</sup>"}}
The expanded uncertainties given in this table are relative to the measured value and thus dimensionless. IR = rated current of the current transformer to be calibrated.
Comparison with reference standard
Primary current : 0.01 A to 8 kA
Secondary current : 0.01 A to 10 A
Frequency : 50 Hz, 60 Hz
Approved on 12 April 2021
Institute service identifier : 71
CMC ID : EURAMET-EM-ES-00000E4R-4
 
High voltage and current , High AC current: ratio: phase displacement : 0 mrad to 20 mrad
Current transformer
Absolute expanded uncertainty : 0.11 mrad
Comparison with reference standard
Primary current : 0.01 A to 8 kA
Secondary current : 0.01 A to 10 A
Frequency : 50 Hz, 60 Hz
According to IEC 61869-2
Approved on 07 February 2019
Institute service identifier : LCOE/72
CMC ID : EURAMET-EM-ES-00000E4S-1
 
High voltage and current , Impulse voltage and current: switching impulse test voltage value : 1 kV to 100 kV
High voltage divider, high voltage measuring system, EMC surge generator
Relative expanded uncertainty : 0.1 %
Comparison with high voltage reference measuring system
IEC 60060-2
Impulse shape : according to IEC 60060-1; 1kV to 100 kV
Approved on 07 February 2019
Institute service identifier : LCOE/70
CMC ID : EURAMET-EM-ES-00000E57-3
 
High voltage and current , Impulse voltage and current: lightning impulse test voltage value : 1 kV to 700 kV
High voltage divider, high voltage measuring system, EMC surge generator
Relative expanded uncertainty : 0.5 %
Comparison with high voltage reference measuring system
IEC 60060-2
Impulse shape : according to IEC 60060-1
Approved on 07 February 2019
Institute service identifier : LCOE/67
CMC ID : EURAMET-EM-ES-00000E54-2
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.