Results for: Norway
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
124 results
Reset all
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Norway, DSA (Norwegian Radiation and Nuclear Safety Authority)
Ionizing radiation
Personal dose equivalent/rate, superficial , Dosimeter : 1.00E-4 Sv/h to 0.12 Sv/h
Relative expanded uncertainty 6 %
Irradiation or calibration on a ISO water slab or ISO rod
X-ray, 10 kV to 50 kV, 10 kV to 40 kV ISO 4037-1 narrow beam
Reference standard : Secondary standard ionization chamber
Source of traceability : VSL
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1021
CMC ID : EURAMET-RI-NO-00000JRH-1
 
Ionizing radiation
Personal dose equivalent/rate, penetrating , Dosemeter : 3.00E-4 Sv/h to 0.015 Sv/h
Relative expanded uncertainty 5 %
Irradiation or calibration on a ISO water slab or ISO rod
Cs-137, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1019
CMC ID : EURAMET-RI-NO-00000JRG-1
 
Ionizing radiation
Absorbed dose/rate to water , Radiotherapy dosemeter : 0.4 Gy to 50 Gy
Relative expanded uncertainty 1 %
Calibration in water phantom
Co-60, IAEA 398
Reference standard : Two secondary standard ionization chambers
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1004
CMC ID : EURAMET-RI-NO-00000JRK-1
 
Ionizing radiation
Kerma/rate , Radiodiagnostic dosemeter and dosimeter : 3.00E-4 Gy/h to 0.1 Gy/h
Relative expanded uncertainty 1.8 %
Calibration free in air
X-ray, 50 kV to 420 kV, 60 kV to 300 kV ISO 4037-1 narrow beam
Reference standard : Secondary standard ionization chamber
Source of traceability : VSL
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1009
CMC ID : EURAMET-RI-NO-00000JRP-1
 
Ionizing radiation
Kerma/rate , Radiotherapy, radiodiagnostic dosemeter : 1.00E-5 Gy/s to 3.00E-3 Gy/s
Relative expanded uncertainty 1 %
Calibration free in air
X-ray, 50 kV to 420 kV, 180 kV, 250 kV, Rapport BIPM 01/04
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1007
CMC ID : EURAMET-RI-NO-00000JRO-1
 
Ionizing radiation
Directional dose equivalent/rate , Radioprotection dosemeter : 3.00E-4 Sv/h to 0.015 Sv/h
Relative expanded uncertainty 5 %
Calibration free in air
Cs-137, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1016
CMC ID : EURAMET-RI-NO-00000JRZ-1
 
Ionizing radiation
Ambient dose equivalent/rate , Radioprotection dosemeter : 0.18 Sv/h to 2 Sv/h
Relative expanded uncertainty 5 %
Calibration free in air
Co-60, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1012
CMC ID : EURAMET-RI-NO-00000JRV-1
 
Ionizing radiation
Kerma/rate , Radiotherapy dosemeter : 0.4 Gy to 50 Gy
Relative expanded uncertainty 0.9 %
Calibration free in air
Co-60, IAEA 277
Reference standard : Two secondary standard ionization chambers
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1003
CMC ID : EURAMET-RI-NO-00000JRR-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Dosimeter : 1.00E-4 Sv/h to 0.12 Sv/h
Relative expanded uncertainty 6 %
Irradiation or calibration on a ISO water slab or ISO rod
X-ray, 50 kV to 420 kV, 60 kV to 300 kV ISO 4037-1 narrow beam
Reference standard : Secondary standard ionization chamber
Source of traceability : VSL
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1022
CMC ID : EURAMET-RI-NO-00000JRI-1
 
Ionizing radiation
Kerma/rate , Radiodiagnostic dosemeter and dosimeter : 1.00E-3 Gy/h to 0.3 Gy/h
Relative expanded uncertainty 1.6 %
Calibration free in air
X-ray, 10 kV to 50 kV, 10 kV to 40 kV ISO 4037-1 narrow beam
Reference standard : Secondary standard ionization chamber
Source of traceability : VSL
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1008
CMC ID : EURAMET-RI-NO-00000JRM-1
 
Ionizing radiation
Kerma/rate , Radiotherapy, radiodiagnostic dosemeter : 1.00E-4 Gy/s to 0.015 Gy/s
Relative expanded uncertainty 1.1 %
Calibration free in air
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, Rapport BIPM 01/04
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1005
CMC ID : EURAMET-RI-NO-00000JRL-1
 
Ionizing radiation
Directional dose equivalent/rate , Radioprotection dosemeter : 0.18 Sv/h to 2 Sv/h
Relative expanded uncertainty 5 %
Calibration free in air
Co-60, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1015
CMC ID : EURAMET-RI-NO-00000JRY-1
 
Ionizing radiation
Absorbed dose/rate to water , Radiotherapy dosemeter : 2.00E-3 Gy/s to 0.01 Gy/s
Relative expanded uncertainty 1 %
Calibration in water phantom
Co-60, IAEA 398
Reference standard : Two secondary standard ionization chambers
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1002
CMC ID : EURAMET-RI-NO-00000JRJ-1
 
Ionizing radiation
Ambient dose equivalent/rate , Radioprotection dosemeter : 3.00E-4 Sv/h to 0.1 Sv/h
Relative expanded uncertainty 6 %
Calibration free in air
X-ray, 50 kV to 420 kV, 60 kV to 300 kV ISO 4037-1 narrow beam
Reference standard : Secondary standard ionization chamber
Source of traceability : VSL
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1014
CMC ID : EURAMET-RI-NO-00000JRU-1
 
Ionizing radiation
Kerma/rate , Radiotherapy dosemeter : 2.00E-3 Gy/s to 0.01 Gy/s
Relative expanded uncertainty 0.9 %
Calibration free in air
Co-60, IAEA 277
Reference standard : Two secondary standard ionization chambers
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1001
CMC ID : EURAMET-RI-NO-00000JRQ-1
 
Ionizing radiation
Personal dose equivalent/rate, penetrating , Dosimeter : 3.00E-4 Sv/h to 0.1 Sv/h
Relative expanded uncertainty 6 %
Irradiation or calibration on a ISO water slab or ISO rod
X-ray, 50 kV to 420 kV, 60 kV to 300 kV ISO 4037-1 narrow beam
Reference standard : Secondary standard ionization chamber
Source of traceability : VSL
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1020
CMC ID : EURAMET-RI-NO-00000JRE-1
 
Ionizing radiation
Personal dose equivalent/rate, penetrating , Dosimeter : 0.18 Sv/h to 2 Sv/h
Relative expanded uncertainty 5 %
Irradiation or calibration on a ISO water slab or ISO rod
Co-60, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1018
CMC ID : EURAMET-RI-NO-00000JRF-1
 
Ionizing radiation
Kerma/rate , Radiotherapy, radiodiagnostic dosemeter : 1.00E-4 Gy/s to 0.03 Gy/s
Relative expanded uncertainty 1 %
Calibration free in air
X-ray, 50 kV to 420 kV, 100 kV, 135 kV, Rapport BIPM 01/04
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1006
CMC ID : EURAMET-RI-NO-00000JRN-1
 
Ionizing radiation
Kerma/rate , Radioprotection dosemeter : 3.00E-4 Gy/h to 0.015 Gy/h
Relative expanded uncertainty 2 %
Calibration free in air
Cs-137, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1011
CMC ID : EURAMET-RI-NO-00000JRT-1
 
Ionizing radiation
Kerma/rate , Radioprotection dosemeter : 0.18 Gy/h to 2 Gy/h
Relative expanded uncertainty 2 %
Calibration free in air
Co-60, ISO 4037-1
Reference standard : Secondary standard ionization chamber
Source of traceability : BIPM
Approved on 24 January 2012
Institute service identifier : EUR-RAD-DSA-1010
CMC ID : EURAMET-RI-NO-00000JRS-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.