Results for: Panama
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Panama, CENAMEP AIP (Centro Nacional de Metrología de Panama AIP)
Mass ,
Mass
:
500 mg
Mass standard
Absolute expanded uncertainty
:
2.5E-2 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-500mg
CMC ID :
SIM-M-PA-0000056C-1
Mass ,
Mass
:
1 mg
Mass standard
Absolute expanded uncertainty
:
6.0E-3 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-1mg
CMC ID :
SIM-M-PA-00000564-1
Mass ,
Mass
:
10 g
Mass standard
Absolute expanded uncertainty
:
6.0E-2 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-10g
CMC ID :
SIM-M-PA-0000056G-1
Frequency ,
Frequency
:
100 kHz
Local frequency standard
Relative expanded uncertainty
:
6.0E-13 Hz/Hz
Direct frequency measurement with a signal mixer
Averaging time : 1 d
Amplitude : 0.5 V to 5 V
Approved on 25 October 2011
Institute service identifier : CMP-F0-S005
CMC ID :
SIM-TF-PA-000005VY-1
Frequency ,
Frequency
:
5 MHz
Local frequency standard
Relative expanded uncertainty
:
6.0E-13 Hz/Hz
Direct frequency measurement with a signal mixer
Averaging time : 1 d
Amplitude : 0.5 V to 5 V
Approved on 25 October 2011
Institute service identifier : CMP-F0-S005
CMC ID :
SIM-TF-PA-000005W0-1
Mass ,
Mass
:
100 g
Mass standard
Absolute expanded uncertainty
:
0.15 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-100g
CMC ID :
SIM-M-PA-0000056J-1
Mass ,
Mass
:
200 g
Mass standard
Absolute expanded uncertainty
:
0.3 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-200g
CMC ID :
SIM-M-PA-0000056K-1
Mass ,
Mass
:
50 g
Mass standard
Absolute expanded uncertainty
:
0.1 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-50g
CMC ID :
SIM-M-PA-0000056I-1
Mass ,
Mass
:
1 kg
Mass standard
Absolute expanded uncertainty
:
1.5 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-1kg
CMC ID :
SIM-M-PA-0000056M-1
AC current ,
AC current up to 100 A: meters
:
0.1 A to 2 A
Multimeter
Relative expanded uncertainty
:
2.0E2 µA/A to 3.3E2 µA/A
Uncertainty table
{"row_1":{"col_1":"","col_2":"60 Hz","col_3":"1 kHz","col_4":"5 kHz"},"row_2":{"col_1":"100 mA","col_2":"300","col_3":"225","col_4":"300"},"row_3":{"col_1":"200 mA","col_2":"225","col_3":"200","col_4":"275"},"row_4":{"col_1":"1 A","col_2":"325","col_3":"325","col_4":"-"},"row_5":{"col_1":"2 A","col_2":"315","col_3":"315","col_4":"-"}}
Direct comparison
Frequency : 60 Hz to 5 kHz
Approved on 04 April 2017
Institute service identifier : CENAMEP AIP/CMP-E4-S001
CMC ID :
SIM-EM-PA-000008O3-1
DC current (up to 100 A) ,
DC current meters: intermediate values
:
1.00E-4 A to 2 A
Multimeter
Relative expanded uncertainty
:
5.5E1 µA/A to 1.2E2 µA/A
Uncertainty table
{"row_1":{"col_1":"","col_2":"Expanded uncertainty in µA/A"},"row_2":{"col_1":"100 µA to 200 µA","col_2":"120"},"row_3":{"col_1":"> 0.2 mA to 1 mA","col_2":"80"},"row_4":{"col_1":"> 1 mA to 100 mA","col_2":"55"},"row_5":{"col_1":"> 100 mA to 200 mA","col_2":"70"},"row_6":{"col_1":"> 0.2 A to 2 A","col_2":"100"}}
Direct comparison
Approved on 04 April 2017
Institute service identifier : CENAMEP AIP/CMP-E4-S001
CMC ID :
SIM-EM-PA-000008O1-1
Frequency ,
Frequency
:
1 Hz to 2.25E8 Hz
General frequency source
Relative expanded uncertainty
:
2.0E-11 Hz/Hz
Direct frequency measurement with a signal mixer
Averaging time : 100 s
Amplitude : 0.5 V to 5 V
Number of measurements : 360
Approved on 25 October 2011
Institute service identifier : CMP-F0-S006
CMC ID :
SIM-TF-PA-000005W2-1
Frequency ,
Frequency
:
1.00E6 Hz to 3.00E9 Hz
Frequency counter
Relative expanded uncertainty
:
5.0E-11 Hz/Hz
Using frequency synthetizers with an external timebase from atomic standard
Averaging time : 10 s
Approved on 25 October 2011
Institute service identifier : CMP-F0-S007
CMC ID :
SIM-TF-PA-000005W5-1
Frequency ,
Frequency
:
1 MHz
Local frequency standard
Relative expanded uncertainty
:
6.0E-13 Hz/Hz
Direct frequency measurement with a signal mixer
Averaging time : 1 d
Amplitude : 0.5 V to 5 V
Approved on 25 October 2011
Institute service identifier : CMP-F0-S005
CMC ID :
SIM-TF-PA-000005VZ-1
Mass ,
Mass
:
20 g
Mass standard
Absolute expanded uncertainty
:
8.0E-2 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-20g
CMC ID :
SIM-M-PA-0000056H-1
Time interval ,
Time interval
:
10 s to 8.64E4 s
Time difference meter
Absolute expanded uncertainty
:
5.0E-2 s
Direct measurement comparison with reference to local standard
Averaging time : 10 s to 86400 s
The service refers to the calibration of time interval measurement devices such as stopwatches, timers, digital and analog clocks for example.
Approved on 25 October 2011
Institute service identifier : CMP-F0-S002
CMC ID :
SIM-TF-PA-000005VW-1
Mass ,
Mass
:
100 mg
Mass standard
Absolute expanded uncertainty
:
1.5E-2 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-100mg
CMC ID :
SIM-M-PA-0000056A-1
Mass ,
Mass
:
200 mg
Mass standard
Absolute expanded uncertainty
:
2.0E-2 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-200mg
CMC ID :
SIM-M-PA-0000056B-1
Mass ,
Mass
:
50 mg
Mass standard
Absolute expanded uncertainty
:
1.2E-2 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-50mg
CMC ID :
SIM-M-PA-00000569-1
Mass ,
Mass
:
10 mg
Mass standard
Absolute expanded uncertainty
:
8.0E-3 mg
Direct comparison
Temperature : ((19 to 24) ± 1) °C
Humidity : ((40 to 60) ± 5) %
Approved on 08 September 2006
Institute service identifier : SRV-CMP-M1-F1-10mg
CMC ID :
SIM-M-PA-00000567-1