Poland, INTiBS (Institute of Low Temperature and Structure Research)
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
173 K to 273 K
Platinum Resistance Thermometer (PRT)
Absolute expanded uncertainty
:
1.0E1 mK
Calibration by comparison at measurement points
Cryostat : Isothermal
Approved on 10 November 2015
Institute service identifier : INTiBS/32.1
CMC ID :
EURAMET-T-PL-000007TD-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
78 K to 173 K
Platinum Resistance Thermometer (PRT)
Absolute expanded uncertainty
:
1.6E1 mK
Calibration by comparison at measurement points
Cryostat : Isothermal
Approved on 10 November 2015
Institute service identifier : INTiBS/32
CMC ID :
EURAMET-T-PL-000007TC-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-100 °C to 0 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
1.0E-2 °C
Calibration by comparison at measurement points
Cryostat : Isothermal
Approved on 07 September 2016
Institute service identifier : INTiBS/67.1
CMC ID :
EURAMET-T-PL-000007V3-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-196 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty
:
1.6E-2 °C
Calibration by comparison
Approved on 10 November 2015
Institute service identifier : INTiBS/62
CMC ID :
EURAMET-T-PL-000007UY-1
Items for defining ITS-90 ,
Temperature
:
273.16 K
Water triple point cell for capsule SPRT
Absolute expanded uncertainty
:
8.0E-2 mK
Comparison with a cell
Approved on 10 November 2015
Institute service identifier : INTiBS/68
CMC ID :
EURAMET-T-PL-000007U6-1
Items for defining ITS-90 ,
Temperature
:
273.16 K
Capsule SPRT
Absolute expanded uncertainty
:
0.2 mK
Measurement at triple point of water
Approved on 10 November 2015
Institute service identifier : INTiBS/69
CMC ID :
EURAMET-T-PL-000007U8-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-196 °C to -100 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
0.3 °C to 0.2 °C
Calibration by comparison at measurement points
Cryostat : Isothermal
Approved on 07 September 2016
Institute service identifier : INTiBS/67.2
CMC ID :
EURAMET-T-PL-000007V4-1
Items for defining ITS-90 ,
Temperature
:
54.3584 K
Oxygen triple point cell for capsule SPRT
Absolute expanded uncertainty
:
0.43 mK
Comparison with reference cell
Approved on 10 November 2015
Institute service identifier : INTiBS/3
CMC ID :
EURAMET-T-PL-000007UA-1
Items for defining ITS-90 ,
Temperature
:
83.8058 K
SPRT - capsule type
Absolute expanded uncertainty
:
0.34 mK
Measurement at triple point of argon
Approved on 20 May 2010
Institute service identifier : INTiBS/19
CMC ID :
EURAMET-T-PL-000007UF-1
Items for defining ITS-90 ,
Temperature
:
24.5561 K
Neon triple point cell for capsule SPRT
Absolute expanded uncertainty
:
0.34 mK
Comparison with reference cell
Approved on 10 November 2015
Institute service identifier : INTiBS/2
CMC ID :
EURAMET-T-PL-000007U9-1
Items for defining ITS-90 ,
Temperature
:
83.8058 K
Argon triple point cell for capsule SPRT
Absolute expanded uncertainty
:
0.33 mK
Comparison with reference cell
Approved on 10 November 2015
Institute service identifier : INTiBS/4
CMC ID :
EURAMET-T-PL-000007UB-1
Items for defining ITS-90 ,
Temperature
:
234.3156 K
SPRT - capsule type
Absolute expanded uncertainty
:
0.38 mK
Measurement at triple point of mercury
Approved on 20 May 2010
Institute service identifier : INTiBS/20
CMC ID :
EURAMET-T-PL-000007UG-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-196 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
1.6E-2 °C
Calibration by comparison
Approved on 10 November 2015
Institute service identifier : INTiBS/66
CMC ID :
EURAMET-T-PL-000007V1-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-100 °C to 0 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
0.2 °C to 0.1 °C
Calibration by comparison at measurement points
Cryostat : Isothermal
Approved on 07 September 2016
Institute service identifier : INTiBS/67.3
CMC ID :
EURAMET-T-PL-000007V5-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-196 °C to -100 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
1.6E-2 °C
Calibration by comparison at measurement points
Cryostat : Isothermal
Approved on 07 September 2016
Institute service identifier : INTiBS/67
CMC ID :
EURAMET-T-PL-000007V2-1
Items for defining ITS-90 ,
Temperature
:
234.3156 K
Mercury triple point cell for capsule SPRT
Absolute expanded uncertainty
:
0.32 mK
Comparison with reference cell
Approved on 10 November 2015
Institute service identifier : INTiBS/6
CMC ID :
EURAMET-T-PL-000007UC-1
Items for defining ITS-90 ,
Temperature
:
54.3584 K
SPRT - capsule type
Absolute expanded uncertainty
:
0.43 mK
Measurement at triple point of oxygen
Approved on 20 May 2010
Institute service identifier : INTiBS/18
CMC ID :
EURAMET-T-PL-000007UE-1
Items for defining ITS-90 ,
Temperature
:
24.5561 K
SPRT - capsule type
Absolute expanded uncertainty
:
0.47 mK
Measurement at triple point of neon
Approved on 20 May 2010
Institute service identifier : INTiBS/17
CMC ID :
EURAMET-T-PL-000007UD-1