Results for: Russian Federation
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
1735 results
Reset all
CMC Area
To filter, either use the tick boxes or click on the underscored item
Chemical material
Chemical Analyte
Geographical location
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Russian Federation, RCAM-Rostest (Research Center for Applied Metrology - Rostest)
High voltage and current , AC high voltage ratio: phase displacement : 5.00E-5 rad to 1.00E-3 rad
High voltage transformer, voltage transformer bridge
Absolute expanded uncertainty : 5.0E-5 rad
Current comparator bridge and compressed gas capacitors
Voltage : 1 kV to 350 kV
Frequency : 50 Hz
Ratio : 10 to 10000
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/180
CMC ID : COOMET-EM-RU-00000Q7X-1
 
High voltage and current , Pulsed high voltage and current: switching impulse time parameters : 100 µs to 2500 µs
Switching impulse voltage measurement set-up, impulse calibrator, digital recorder
Relative expanded uncertainty : 3.0E-2
Comparison with reference measuring system
Peak voltage value : 1 kV to 500 kV
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/187
CMC ID : COOMET-EM-RU-00000Q87-1
 
High voltage and current , High DC voltage: meters : 1 kV to 100 kV
DC kilovolt meter, dedicated set-up for high voltage
Relative expanded uncertainty : 5.0E-5
Differential voltage meter, set of calibrated Zener diode chains
Temperature : 20 °C to 30 °C
Relative humidity : 40% to 80%
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/171
CMC ID : COOMET-EM-RU-00000Q88-1
 
Form , Roundness standards : 0 µm to 300 µm
Magnification stanart (e. g. fkick standard): roundness
Absolute expanded uncertainty : 2.8E1 nm to 3.0E3 nm
Q[28 , 10R] R in mm
The uncertainty is expressed in nm
Stylus-on-spindle roundness instrument
Diameter : 1 mm to 300 mm
Approved on 09 November 2016
Institute service identifier : COO-RCAM-Rostest-L-423/4
CMC ID : COOMET-L-RU-00000Q7T-1
 
Complex geometry , Complex geometry : 20 mm3 to 2000 mm3
CMM
Absolute expanded uncertainty : 1.4 µm to 4.4 µm
Q[1.35, 1.5L] L length in m
The uncertainty is expressed in µm
Ball plate calibration sphere
Temperature : 20 °C ± 0.3 °C
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/22
CMC ID : COOMET-L-RU-00000Q7S-1
 
Form , Flatness standards : 2.00E-3 µm to 2 µm
Optical flat: flatness
Absolute expanded uncertainty : 8.2 nm
Fizeau interferometer
Diameter : 50 mm to 120 mm
Approved on 30 September 2020
Institute service identifier : COO-L-RCAM-Rostest 411
CMC ID : COOMET-L-RU-00000Q7V-1
 
Pressure , Gauge pressure: gas medium : 4.00E4 Pa to 6.30E5 Pa
Pressure measuring device, standard pressure generator
Relative expanded uncertainty : 2.0E-4
Direct comparison with standard, crossfloat
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/5
CMC ID : COOMET-M-RU-00000Q8D-1
 
High voltage and current , Pulsed high voltage and current: switching impulse voltage parameters : 50 V to 2000 V
Switching impulse voltage measurement set-up, impulse calibrator, digital recorder
Relative expanded uncertainty : 5.0E-4 to 2.0E-3
Differential voltage meter, set of calibrated Zener diode-transistor chains
Impulse shape : according to IEC 60060
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/185
CMC ID : COOMET-EM-RU-00000Q85-1
 
High voltage and current , Pulsed high voltage and current: switching impulse voltage parameters : 1 kV to 800 kV
Switching impulse voltage measurement set-up
Relative expanded uncertainty : 3.0E-3
Differential voltage meter, set of calibrated Zener diode chains
Impulse shape : according to IEC 60060
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/186
CMC ID : COOMET-EM-RU-00000Q86-1
 
High voltage and current , High DC voltage: ratios : 1 to 1.00E5
High voltage resistive divider
Relative expanded uncertainty : 7.0E-5 to 3.0E-4
Measurement by means of differential voltage meter, set of calibrated Zener diode chains, DC voltmeter
Voltage : 1 kV to 800 kV
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/174
CMC ID : COOMET-EM-RU-00000Q8B-1
 
High voltage and current , Pulsed high voltage and current: lightning impulse voltage parameters : 50 kV to 1500 kV
Lightning impulse voltage measurement set-up
Relative expanded uncertainty : 5.0E-3
Comparison with reference measuring system
Impulse shape : according to IEC 60060
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/183
CMC ID : COOMET-EM-RU-00000Q83-1
 
High voltage and current , High voltage impedance: capacitance : 10 pF to 1000 pF
Compressed gas capacitor
Relative expanded uncertainty : 1.5E-4 to 1.0E-3
Comparison with reference
Voltage : 1 kV to 350 kV
Frequency : 50 Hz
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/175
CMC ID : COOMET-EM-RU-00000Q8C-1
 
Form , Roundness standards : 0 nm to 2000 nm
Sphere, hemisphere: roundness
Absolute expanded uncertainty : 2.8E1 nm
Stylus-on-spindle roundness instrument
Diameter : 1 mm to 100 mm
Approved on 09 November 2016
Institute service identifier : COO-RCAM-Rostest-L-423/4
CMC ID : COOMET-L-RU-00000Q7U-1
 
High voltage and current , Pulsed high voltage and current: lightning impulse time parameters : 0.8 µs to 60 µs
Lightning impulse voltage measurement set-up, impulse calibrator, digital recorder
Relative expanded uncertainty : 5.0E-2
Comparison with reference measuring system
Peak voltage value : 10 V to 500 kV
Impulse shape : according to IEC 60060
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/184
CMC ID : COOMET-EM-RU-00000Q84-1
 
High voltage and current , High DC voltage: meters : 400 kV to 800 kV
Dedicated set-up for high voltage
Relative expanded uncertainty : 3.0E-4
Comparison by means of differential voltage meter and set of calibrated Zener diode chains
Temperature : 20 °C to 30 °C
Relative humidity : 40% to 80%
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/173
CMC ID : COOMET-EM-RU-00000Q8A-1
 
High voltage and current , High voltage impedance: dissipation factor : 5.00E-5 to 1.00E-3
Compressed gas capacitor
Absolute expanded uncertainty : 5.0E-5
Comparison with reference
Frequency : 50 Hz
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/176
CMC ID : COOMET-EM-RU-00000Q7W-1
 
Complex geometry , CMM artifacts : 10 mm to 960 mm
Ball bar, centre spacing L
Absolute expanded uncertainty : 0.91 µm to 1.6 µm
Q[0.9, 0.75L] L in m
The uncertainty is expressed in µm
CMM with laser interferometer
Temperature : 20 °C ± 0.3 °C
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/21
CMC ID : COOMET-L-RU-00000Q7R-1
 
High voltage and current , AC high voltage: ratio error : 0 % to 0.02 %
High voltage transformer, voltage transformer bridge
Relative expanded uncertainty : 2.0E-4 to 5.0E-4
Current comparator bridge and compressed gas capacitors
Voltage : 1 kV to 350 kV
Frequency : 50 Hz
Ratio : 10 to 10000
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/179
CMC ID : COOMET-EM-RU-00000Q80-1
 
High voltage and current , High DC voltage: meters : 100 kV to 400 kV
Dedicated set-up for high voltage
Relative expanded uncertainty : 2.0E-4
Comparison by means of differential voltage meter and set of calibrated Zener diode chains
Temperature : 20 °C to 30 °C
Relative humidity : 40% to 80%
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/172
CMC ID : COOMET-EM-RU-00000Q89-1
 
High voltage and current , AC high voltage: peak values : 1 kV to 350 kV
AC high voltage meter, dedicated set-up for high voltage measurements (resistive and capacitive dividers)
Relative expanded uncertainty : 5.0E-4
Comparison with reference system
Frequency : 50 Hz
Approved on 21 October 2001
Institute service identifier : RCAM-Rostest/177
CMC ID : COOMET-EM-RU-00000Q7Y-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.