Results for: Zambia
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Zambia, ZMA (Zambia Metrology Agency)
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
220 °C to 280 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
1 °C
Comparison in dry block calibrator
The uncertainty claimed is maximum for the whole range. Sensor type is Type N thermocouple. Calculated values of inhomogeneity included in the CMC calculation.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH6-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
280 °C to 450 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.93 °C
Comparison in dry block calibrator
Nominal Hysterisis is included in the uncertainity calculation. The uncertainty claimed is maximum for the whole range.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH0-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
220 °C to 280 °C
Base metal thermocouple
Absolute expanded uncertainty
:
1 °C
Comparison in dry block calibrator
Type N thermocouple. Calculated values of inhomogeneity included in the CMC calculation.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH1-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-40 °C to 0 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty
:
5.0E-2 °C
Comparison in liquid bath
Nominal Hysterisis is included in the uncertainity calculation. The uncertainty claimed is maximum for the whole range.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CGX-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
280 °C to 450 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
1.3 °C
Comparison in dry block calibrator
The uncertainty claimed is maximum for the whole range. Sensor type is Type N thermocouple. Calculated values of inhomogeneity included in the CMC calculation.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH7-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 220 °C
Temperature sensor with display unit
Absolute expanded uncertainty
:
1 °C
Comparison in liquid bath
The uncertainty claimed is maximum for the whole range. Sensor type is Type N thermocouple. Calculated values of inhomogeneity included in the CMC calculation.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH5-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 220 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.5 °C
Comparison in liquid bath
Nominal Hysterisis is included in the uncertainity calculation. The uncertainty claimed is maximum for the whole range.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CGY-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-40 °C to 100 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty
:
0.2 °C
Comparison in liquid bath
Resolution = 0.2 °C upto 50 °C and 0.1 °C above 50 °C. Liquid Type-Mercury.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH3-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
100 °C to 200 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty
:
0.13 °C
Comparison in liquid bath
Resolution= 0.1 °C. Liquid Type-Mercury.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH4-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
220 °C to 280 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.81 °C
Comparison in dry block calibrator
Nominal Hysterisis is included in the uncertainity calculation. The uncertainty claimed is maximum for the whole range.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CGZ-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
280 °C to 450 °C
Base metal thermocouple
Absolute expanded uncertainty
:
1 °C
Comparison in dry block calibrator
Type N thermocouple. Calculated values of inhomogeneity included in the CMC calculation.
Approved on 06 November 2018
Institute service identifier : ZMASIL
CMC ID :
AFRIMETS-T-ZM-00000CH2-1