Results for: cuba
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
123 results
Reset all
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Cuba, CENTIS (Centro de Isótopos)
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.90E6 Bq/g to 8.10E7 Bq/g
Relative expanded uncertainty 3.6 %
HPGe spectrometer, balance
Re-188, point sources for HPGe measurements, active diameter < 5 mm (source-detector distance no less than 20 cm)
Reference standard : Calibrated Ge spectrometer using certified gamma reference point sources: Co-57 (No 96-157), Ba-133 (No 96-161), Eu-152 (No 96-171), Co-60 (No 96-153), Cs-137/Ba-137m (No 96-176), standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2017
CMC ID : COOMET-RI-CU-00000Q4Y-1
 
Ionizing radiation
Activity per unit mass , Reference material: soils/sediment : 0.4 Bq/g to 40 Bq/g
Relative expanded uncertainty 5 %
Low level gamma-ray spectrometry (HPGe spectrometer), balance
K-40, spiked sediment/soil, 50 g or 250 g
Reference standard : Calibrated γ-ray spectrometer using calibrated at CENTIS-DMR solution in spike sediment, set of standard weights
Source of traceability : BFKH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2029
CMC ID : COOMET-RI-CU-00000Q4M-1
 
Ionizing radiation
Activity per unit mass , Reference material: soils/sediment : 0.02 Bq/g to 10 Bq/g
Relative expanded uncertainty 5 % to 3 %
Gamma-ray spectrometry (HPGe spectrometer), balance
Cs-137, spiked sediment/soil, 50 g or 250 g
Reference standard : Calibrated γ-ray spectrometer using calibrated at CENTIS-DMR solution in spike sediment, set of standard weights
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2029
CMC ID : COOMET-RI-CU-00000Q4L-1
 
Ionizing radiation
Activity per unit mass , Reference material: soils/sediment : 0.02 Bq/g to 10 Bq/g
Relative expanded uncertainty 4 %
Gamma-ray spectrometry (HPGe spectrometer), balance
Cs-134, spiked sediment/soil, 50 g or 250 g
Reference standard : Calibrated γ-ray spectrometer using calibrated at CENTIS-DMR solution in spike sediment, set of standard weights
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2029
CMC ID : COOMET-RI-CU-00000Q4I-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.70E5 Bq/g to 8.10E7 Bq/g
Relative expanded uncertainty 3.8 %
Well type ionization chamber, balance
Co-60, 5 ml nuclide solution in a glass ampoule
Reference standard : Certified gamma reference point source of Co-60 (No 96-153) by 4πβ-γ coincidence counting, standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2012
CMC ID : COOMET-RI-CU-00000Q4Q-1
 
Ionizing radiation
Activity , Single nuclide solid point source (gamma emitting) : 3800 Bq to 4.20E5 Bq
Relative expanded uncertainty 1.6 %
Gamma ray spectrometer
Am-241, polyvinil acetate foils, active diameter < 5 mm
Reference standard : Certified gamma reference point source of Am-241 (No 96-165) by 4πα-γ coincidence counting
Source of traceability : MKEH (Bq)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2028
CMC ID : COOMET-RI-CU-00000Q4D-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.50E5 Bq/g to 8.10E7 Bq/g
Relative expanded uncertainty 1.8 %
HPGe spectrometer, balance
Co-57, point sources, active diameter < 5 mm
Reference standard : Certified gamma reference point source of Co-57 traceable to a secondary standard ionization chamber calibrated using a 4 pi [ e-X]-gamma coincidence counting, set of standard weights
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 23 September 2011
Institute service identifier : COO-RAD-CENTIS-DMR-2041
CMC ID : COOMET-RI-CU-00000Q52-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.90E6 Bq/g to 1.50E8 Bq/g
Relative expanded uncertainty 3.8 %
Well type ionization chamber, balance
Tl-201, 5 ml nuclide solution in a glass ampoule
Reference standard : Calibrated Ge spectrometer using certified gamma reference point sources: Am-241 (No 96-165), Co-57 (No 96-157), Ba-133 (No 96-161), Eu-152 (No 96-171), standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2002
CMC ID : COOMET-RI-CU-00000Q54-1
 
Ionizing radiation
Activity per unit mass , Reference material: biological material : 0.02 Bq/g to 10 Bq/g
Relative expanded uncertainty 9.6 %
Gamma-ray spectrometry (HPGe spectrometer), balance
Eu-152, cylindrical container, sample density from 0.9 g cm-3 to 1.1 g cm-3
Reference standard : Calibrated γ-ray spectrometer using a sediment sample spiked with a calibrated Eu-152 solution, set of standard weights
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 23 September 2011
Institute service identifier : COO-RAD-CENTIS-DMR-2048
CMC ID : COOMET-RI-CU-00000Q5U-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.70E5 Bq/g to 8.10E7 Bq/g
Relative expanded uncertainty 3.8 %
Well type ionization chamber, balance
Cs-137, 5 ml nuclide solution in a glass ampoule
Reference standard : Certified gamma reference point source of Cs-137/Ba-137m (No 96-176) by 4πβ-γ tracer coincidence counting, standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2008
CMC ID : COOMET-RI-CU-00000Q56-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.70E5 Bq/g to 8.10E7 Bq/g
Relative expanded uncertainty 3.8 %
Well type ionization chamber, balance
Eu-152, 5 ml nuclide solution in a glass ampoule
Reference standard : Certified gamma reference point source of Eu-152 (No 96-171) by 4π(β+X+e)-γ coincidence counting, standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2014
CMC ID : COOMET-RI-CU-00000Q5C-1
 
Ionizing radiation
Activity per unit mass , Mixed Nuclide Solution (Low Level) : 0.1 Bq/g to 20 Bq/g
Relative expanded uncertainty 2.5 %
Gamma-ray spectrometry (HPGe spectrometer), balance
Cs-137, 150 ml to 500 ml of solution in plastic cylindrical containers
Reference standard : Calibrated γ-ray spectrometer using calibrated at CENTIS-DMR solutions, set of standard weights
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2030
CMC ID : COOMET-RI-CU-00000Q5O-1
 
Ionizing radiation
Activity per unit mass , Mixed Nuclide Solution (Low Level) : 0.1 Bq/g to 15 Bq/g
Relative expanded uncertainty 3 %
Gamma-ray spectrometry (HPGe spectrometer), balance
Co-60, 150 ml to 500 ml of solution in plastic cylindrical containers
Reference standard : Calibrated γ-ray spectrometer using calibrated at CENTIS-DMR solutions, set of standard weights
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2030
CMC ID : COOMET-RI-CU-00000Q5N-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.00E5 Bq/g to 8.00E7 Bq/g
Relative expanded uncertainty 1.5 %
Liquid scintillation counting (CIEMAT/NIST), balance
P-32, scintillation vial
Reference standard : Liquid scintillation counter (CIEMAT/NIST), set of standard weights
Source of traceability : CENTIS-DMR (Bq), LNE and PTB (H-3 tracer), INIMET (g)
Approved on 23 September 2011
Institute service identifier : COO-RAD-CENTIS-DMR-2033
CMC ID : COOMET-RI-CU-00000Q59-1
 
Ionizing radiation
Activity , Sealed well type ionization chamber used in medicine : 3.00E6 Bq to 1.60E9 Bq
Relative expanded uncertainty 5 %
Comparison
Y-90, nuclide solution in glass ampoules and plastic syringes of commom use in nuclear medicine, different volumes
Reference standard : Secondary standard ionization chamber
Source of traceability : CENTIS-DMR (Bq), INIMET (g)
Approved on 23 September 2011
Institute service identifier : COO-RAD-CENTIS-DMR-2040
CMC ID : COOMET-RI-CU-00000Q46-1
 
Ionizing radiation
Activity , Single nuclide solid point source (gamma emitting) : 3000 Bq to 2.40E5 Bq
Relative expanded uncertainty 1.3 %
Gamma ray spectrometer
Co-60, polyvinil acetate foils, active diameter < 5 mm
Reference standard : Certified gamma reference point source of Co-60 (No 96-153) by 4πβ-γ coincidence counting
Source of traceability : MKEH (Bq)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2026
CMC ID : COOMET-RI-CU-00000Q4F-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.90E6 Bq/g to 8.10E7 Bq/g
Relative expanded uncertainty 3.8 %
Well type ionization chamber, balance
I-131, 5 ml nuclide solution in a glass ampoule
Reference standard : Calibrated Ge spectrometer using certified gamma reference point sources: Am-241 (No 96-165), Co-57 (No 96-157), Ba-133 (No 96-161), Eu-152 (No 96-171), Co-60 (No 96-153), Cs-137/Ba-137m (No 96-176), standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2006
CMC ID : COOMET-RI-CU-00000Q5E-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.90E6 Bq/g to 1.10E9 Bq/g
Relative expanded uncertainty 2.6 %
HPGe spectrometer, balance
Tc-99m, point sources for HPGe measurements, active diameter < 5 mm (source-detector distance no less than 20 cm)
Reference standard : Calibrated Ge spectrometer using certified gamma reference point sources: Am-241 (No 96-165), Co-57 (No 96-157), Ba-133 (No 96-161), Eu-152 (No 96-171), standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2003
CMC ID : COOMET-RI-CU-00000Q4Z-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.90E6 Bq/g to 1.50E8 Bq/g
Relative expanded uncertainty 3 %
HPGe spectrometer, balance
Tl-201, point sources for HPGe measurements, active diameter < 5 mm (source-detector distance no less than 20 cm)
Reference standard : Calibrated Ge spectrometer using certified gamma reference point sources: Am-241 (No 96-165), Co-57 (No 96-157), Ba-133 (No 96-161), Eu-152 (No 96-171), standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2001
CMC ID : COOMET-RI-CU-00000Q58-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1.90E6 Bq/g to 1.10E9 Bq/g
Relative expanded uncertainty 3.8 %
Well type ionization chamber, balance
Tc-99m, 5 ml nuclide solution in a glass ampoule
Reference standard : Calibrated Ge spectrometer using certified gamma reference point sources: Am-241 (No 96-165), Co-57 (No 96-157), Ba-133 (No 96-161), Eu-152 (No 96-171), standard weighing set
Source of traceability : MKEH (Bq), INIMET (g)
Approved on 13 April 2005
Institute service identifier : COO-RAD-CENTIS-DMR-2004
CMC ID : COOMET-RI-CU-00000Q4T-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.