Results for: moldova
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Moldova, Republic of, NMI (MD) (The National Metrology Institute of the Republic of Moldova)
Items for defining ITS-90 ,
Temperature
:
0.01 °C to 660.323 °C
Standard Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.3 mK to 3.6 mK
Calibration at fixed points
Temperature-controlled furnace : Heat-pipe
furnace
Approved on 31 March 2020
Institute service identifier : 340041
CMC ID :
EURAMET-T-MD-00000KHV-1
Items for defining ITS-90 ,
Temperature
:
660.323 °C
Standard Platinum Resistance Thermometer
Absolute expanded uncertainty
:
3.6 mK
Calibration at aluminium fixed point
Temperature-controlled furnace : Heat-pipe
furnace
Approved on 31 March 2020
Institute service identifier : 340040
CMC ID :
EURAMET-T-MD-00000KHQ-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
156.5985 °C
Indium cell
Absolute expanded uncertainty
:
1.2 mK
Direct comparison
Temperature-controlled furnace : 3-zone
Approved on 31 March 2020
Institute service identifier : 340038
CMC ID :
EURAMET-T-MD-000006CS-2
Items for defining ITS-90 ,
Temperature
:
0.01 °C
Standard Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.3 mK
Water triple point
Thermostat : ice bath
Approved on 31 March 2020
Institute service identifier : 340035
CMC ID :
EURAMET-T-MD-000006CV-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 300 °C
Noble metal thermocouple, type S, R and B
Absolute expanded uncertainty
:
0.4 °C
Comparison method
Stirred liquid bath : ethanol, water or silicon oil
Approved on 31 March 2020
Institute service identifier : 340013
CMC ID :
EURAMET-T-MD-00000KHZ-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
80 °C to 300 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty
:
3.0E-2 K
Direct comparison
Thermostat : silicon oil
Approved on 31 March 2020
Institute service identifier : 340011
CMC ID :
EURAMET-T-MD-000006D2-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty
:
6.0E-3 K
Direct comparison
Thermostat : ethanol
Approved on 31 March 2020
Institute service identifier : 340011
CMC ID :
EURAMET-T-MD-000006D0-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 80 °C
Digital thermometer thermistor
Absolute expanded uncertainty
:
4.0E-2 K
Comparison method
Thermostat : water
Approved on 31 March 2020
Institute service identifier : 340019
CMC ID :
EURAMET-T-MD-000006DI-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-80 °C to 10 °C
Digital thermometer IPRT
Absolute expanded uncertainty
:
4.0E-2 K
Comparison method
Cryostat : ethanol
Approved on 31 March 2020
Institute service identifier : 340019
CMC ID :
EURAMET-T-MD-000006DF-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 150 °C
Thermistor
Absolute expanded uncertainty
:
5.0E-2 K
Comparison method
Thermostat : silicon oil
Approved on 31 March 2020
Institute service identifier : 340042
CMC ID :
EURAMET-T-MD-000006D5-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 660 °C
Noble metal thermocouple, type S, R and B
Absolute expanded uncertainty
:
0.4 °C
Calibration at fixed points
Approved on 31 March 2020
Institute service identifier : 340045
CMC ID :
EURAMET-T-MD-00000KHY-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
300 °C to 660 °C
Digital thermometer IPRT
Absolute expanded uncertainty
:
0.3 K to 0.7 K
Comparison method
Approved on 31 March 2020
Institute service identifier : 340019
CMC ID :
EURAMET-T-MD-00000KIA-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
29.7646 °C
Thermocouple S and R
Absolute expanded uncertainty
:
0.4 °C
Calibration at gallium fixed point
Approved on 31 March 2020
Institute service identifier : 340043
CMC ID :
EURAMET-T-MD-000006D7-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
156.5985 °C
Digital thermometer thermocouple
Absolute expanded uncertainty
:
0.4 K
Calibration at indium fixed point
Temperature-controlled furnace : 3-zone
Approved on 31 March 2020
Institute service identifier : 340047
CMC ID :
EURAMET-T-MD-000006CJ-2
Fluid flow ,
Volume of liquid
:
2 L to 1000 L
Proving tanks
Relative expanded uncertainty
:
2.1E-2 %
Volumetric
Liquid : water
Reference temperature : 20 0C
Approved on 20 September 2023
Institute service identifier : 362005
CMC ID :
EURAMET-M-MD-00000OJH-1
Items for defining ITS-90 ,
Temperature
:
0.01 °C to 419.527 °C
Standard Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.3 mK to 2.7 mK
Calibration at fixed points
Temperature-controlled furnace : 3-zone
Approved on 31 March 2020
Institute service identifier : 340010
CMC ID :
EURAMET-T-MD-00000KHU-1
Items for defining ITS-90 ,
Temperature
:
29.7646 °C
Standard Platinum Resistance Thermometer
Absolute expanded uncertainty
:
0.5 mK
Calibration at gallium fixed point
Temperature-controlled furnace : 3-zone
Approved on 31 March 2020
Institute service identifier : 340035
CMC ID :
EURAMET-T-MD-000006CW-2
Items for defining ITS-90 ,
Temperature
:
231.928 °C
Standard Platinum Resistance Thermometer
Absolute expanded uncertainty
:
1.5 mK
Calibration at tin fixed point
Temperature-controlled furnace : 3-zone
Approved on 31 March 2020
Institute service identifier : 340009
CMC ID :
EURAMET-T-MD-000006CY-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
156.5985 °C
Thermocouple S and R
Absolute expanded uncertainty
:
0.4 °C
Calibration at indium fixed point
Temperature-controlled furnace : 3-zone
Approved on 31 March 2020
Institute service identifier : 340043
CMC ID :
EURAMET-T-MD-000006D8-2
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C
Thermistor
Absolute expanded uncertainty
:
6.0E-3 K
Comparison method
Thermostat : ethanol
Approved on 31 March 2020
Institute service identifier : 340042
CMC ID :
EURAMET-T-MD-000006D3-2