Results for: paraguay
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Paraguay, INTN (Instituto Nacional de Tecnologia y Normalizacion)
Mass ,
Conventional mass
:
1 mg to 100 mg
Mass standard
Absolute expanded uncertainty
:
3.0E-3 mg to 5.0E-3 mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GHZ-1
Pressure ,
Gauge pressure: oil medium
:
2.00E5 Pa to 1.00E8 Pa
Pressure gauge
Absolute expanded uncertainty
1.98E-04p + 2.14E+02 The uncertainty is expressed in Pa
Direct comparison
Temperature : (20 ± 2) °C
p pressure in Pa
Approved on 27 January 2017
Institute service identifier : INTN/ONM - Presión
CMC ID :
SIM-M-PY-00000GHY-1
Fluid flow ,
Volume of liquid
:
1 mL to 5 mL
Volumetric test measure (one mark pipette)
Absolute expanded uncertainty
:
2.0E-3 mL to 6.0E-3 mL
0.001V + 0.001
V in mL
Gravimetric determination
ISO 4787
Fluid : Pure water
Temperature : (18 to 25) °C
Approved on 23 September 2023
Institute service identifier : INTN/ONM-Volumen
CMC ID :
SIM-M-PY-00000NDD-1
Mass ,
Conventional mass
:
10 g to 100 g
Mass standard
Absolute expanded uncertainty
:
2.0E-2 mg to 5.0E-2 mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GI2-1
Mass ,
Conventional mass
:
20 kg
Mass standard
Absolute expanded uncertainty
:
3.0E1 mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GI5-1
Fluid flow ,
Volume of liquid
:
10 mL to 100 mL
Volumetric test measure (pycnometer)
Absolute expanded uncertainty
:
3.6E-3 mL to 8.1E-3 mL
0.00005V + 0.0031
V in mL
Gravimetric determination
Fluid : Pure water
Temperature : (18 to 25) °C
Approved on 30 September 2023
Institute service identifier : INTN/ONM-Volumen
CMC ID :
SIM-M-PY-00000NDG-1
Mass ,
Conventional mass
:
1 g to 10 g
Mass standard
Absolute expanded uncertainty
:
1.0E-2 mg to 2.0E-2 mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GI1-1
Mass ,
Conventional mass
:
1 kg to 10 kg
Mass standard
Absolute expanded uncertainty
:
0.5 mg to 1.6E1 mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GI4-1
Mass ,
Conventional mass
:
0.1 kg to 1 kg
Mass standard
Absolute expanded uncertainty
0.050 to 0.50 The uncertainty is expressed in mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GI3-1
Fluid flow ,
Volume of liquid
:
10 mL to 100 mL
Volumetric test measure (one mark pipette)
Absolute expanded uncertainty
:
7.0E-3 mL to 1.6E-2 mL
0.0001V + 0.006
V in mL
Gravimetric determination
ISO 4787
Fluid : Pure water
Temperature : (18 to 25) °C
Approved on 30 September 2023
Institute service identifier : INTN/ONM-Volumen
CMC ID :
SIM-M-PY-00000NDE-1
Fluid flow ,
Volume of liquid
:
1 mL to 1000 mL
Volumetric test measure (flask)
Absolute expanded uncertainty
:
3.7E-3 mL to 0.1 mL
0.0001V + 0.0036
V in mL
Gravimetric determination
ISO 4787
Fluid : Pure water
Temperature : (18 to 25) °C
Approved on 23 September 2023
Institute service identifier : INTN/ONM-Volumen
CMC ID :
SIM-M-PY-00000NDC-1
Mass ,
Conventional mass
:
0.1 g to 1 g
Mass standard
Absolute expanded uncertainty
:
5.0E-3 mg to 1.0E-2 mg
Direct comparison
Temperature : 23 °C
Humidity : 40 % to 60 %
Pressure : 1000 hPa
Approved on 18 June 2019
Institute service identifier : INTN/ONM-Masa
CMC ID :
SIM-M-PY-00000GI0-1
Fluid flow ,
Volume of liquid
:
10 mL to 100 mL
Volumetric test measure (graduated pipette and burette)
Absolute expanded uncertainty
:
8.1E-3 mL to 5.3E-2 mL
0.0005 V + 0.0031
V in mL
Gravimetric determination
ISO 4787
Fluid : Pure water
Temperature : (18 to 25) °C
Approved on 30 September 2023
Institute service identifier : INTN/ONM-Volumen
CMC ID :
SIM-M-PY-00000NDF-1
Fluid flow ,
Volume of liquid
:
20 µL to 1.00E4 µL
Volumetric test measure (piston pipettes)
Absolute expanded uncertainty
:
0.15 µL to 2.5E1 µL
0.0025V + 0.10
V in μL
Gravimetric determination
ISO 8655-6
Fluid : Pure water
Temperature : (18 to 23) °C
Approved on 30 September 2023
Institute service identifier : INTN/ONM-Volumen
CMC ID :