Results for: portugal flow
				Filters
				
					CMC Quick search Filters close;
				 
				
					
					
						
							CMC Area
							
							To filter, either use the tick boxes or click on the underscored item
						 
						
						
						
					 
			 
			
				
				
				
					In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
					Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
				
					Portugal, IPQ (Instituto Portugues da Qualidade)
					
						
    
        
            
                Fluid flow ,
				 Volume flowrate (or Mass flowrate); liquid
                :
                0.12 mL/h to 2000 mL/h
            
            
                Flow meter and flow generator (e.g. coriolis meters, syringe pumps)
            
            
                Relative expanded uncertainty
                :
                2.5 % to 0.11 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Temperature : ambient
            
            
                Pressure : up to 3 hPa
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/104.01, 104.02, 104.03, 104.61
            
            
                CMC ID :
                EURAMET-M-PT-00000C9I-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                1 mL to 1.00E4 mL
            
            
                Glassware equipment; Pipettes, flasks, burettes, cylinders
            
            
                Relative expanded uncertainty
                :
                1.0E-2 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Reference temperature : 20 °C
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.08, 103.07, 103.06, 103.09
            
            
                CMC ID :
                EURAMET-M-PT-00000C9U-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                DC current (up to 100 A) ,
				 DC current meters: low values
                :
                1.00E-8 A to 1.00E-4 A
            
            
                Ammeter, multimeter
            
            
                Relative expanded uncertainty
                :
                1.5E1 µA/A to 7.5E1 µA/A
            
            
                Uncertainty table
                
                    {"row_1": {"col_1":"Current", "col_2":"Expanded uncertainty<br>/ (µA/A)"}, "row_2": {"col_1":"10 nA to 100 nA", "col_2":"75"}, "row_3": {"col_1":"100 nA to 1 µA", "col_2":"25"}, "row_4": {"col_1":"1 µA to 10 µA", "col_2":"20"}, "row_5": {"col_1":"10 µA to 100 µA", "col_2":"15"}}
                    
                    
                 
             
            
            
                Source characterised using voltage and standard resistor
            
            
            
            
            
                Approved on 31 March 2022
            
            
                Institute service identifier :  IPQ/403.22
            
            
                CMC ID :
                EURAMET-EM-PT-00000MHH-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Liquid flow
                :
                3.00E-4 mL/h to 0.12 mL/h
            
            
                Flow generators (e.g. syringe pumps)
            
            
                Relative expanded uncertainty
                :
                2.7 % to 1.3 %
            
            
            
            
                Interferometric method
            
            
            
                Reference temperature : 20 ºC
            
            
                Liquid : Water
            
            
                Pressure : up to 3 hPa
            
            
            
                Approved on 09 July 2022
            
            
                Institute service identifier :  104.01, 104.02 and 104.03
            
            
                CMC ID :
                EURAMET-M-PT-00000NW9-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                DC resistance ,
				 DC resistance standards and sources: low values
                :
                1.00E-4 Ω to 1 Ω
            
            
                Standard resistor
            
            
                Relative expanded uncertainty
                :
                5 µΩ/Ω to 1.2 µΩ/Ω
            
            
                Uncertainty table
                
                    {"row_1": {"col_1":"<strong>Resistance<\/strong>", "col_2":"<strong>Expanded uncertainty/(µΩ/Ω)<\/strong>"}, "row_2": {"col_1":"0.1 mΩ to 10 mΩ", "col_2":"5"}, "row_3": {"col_1":"0.1 Ω", "col_2":"2"}, "row_4": {"col_1":"1 Ω", "col_2":"1.2"}, "row_5": {"col_1":"10 kΩ", "col_2":"1.2"}, "row_6": {"col_1":"10 Ω to 100 kΩ", "col_2":"1.5"}, "row_7": {"col_1":"100 kΩ to 1 MΩ", "col_2":"2"}, "row_8": {"col_1":"1 MΩ to 10 MΩ", "col_2":"3"}, "row_9": {"col_1":"10 MΩ to 100 MΩ", "col_2":"7"}, "row_10": {"col_1":"100 MΩ", "col_2":"20"}}
                    
                    
                 
             
            
            
                Comparison using DCC bridge
            
            
            
                Oil and air bath temperature : 18 °C to 25 °C
            
            
            
                Approved on 04 March 2021
            
            
                Institute service identifier :  IPQ/402.11, IPQ/402.12
            
            
                CMC ID :
                EURAMET-EM-PT-0000069X-3 
             
         
        
        
     
 
					 
					
						
    
        
            
                DC current (up to 100 A) ,
				 DC current sources: low values
                :
                1.00E-8 A to 1.00E-4 A
            
            
                Current generator, calibrator
            
            
                Relative expanded uncertainty
                :
                1.5E1 µA/A to 7.5E1 µA/A
            
            
                Uncertainty table
                
                    {"row_1": {"col_1":"Current", "col_2":"Expanded uncertainty<br>/ (µA/A)"}, "row_2": {"col_1":"10 nA to 100 nA", "col_2":"75"}, "row_3": {"col_1":"100 nA to 1 µA", "col_2":"25"}, "row_4": {"col_1":"1 µA to 10 µA", "col_2":"20"}, "row_5": {"col_1":"10 µA to 100 µA", "col_2":"15"}}
                    
                    
                 
             
            
            
                Comparison
            
            
            
            
            
                Approved on 31 March 2022
            
            
                Institute service identifier :  IPQ/403.11
            
            
                CMC ID :
                EURAMET-EM-PT-00000MHE-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                0.1 µL to 2.00E4 µL
            
            
                Micropipettes or piston pipettes
            
            
                Relative expanded uncertainty
                :
                5 % to 0.1 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Reference temperature : 20 °C
            
            
            
                Approved on 03 December 2024
            
            
                Institute service identifier :  IPQ/103.02
            
            
                CMC ID :
                EURAMET-M-PT-00000C9F-5 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume
                :
                0.1 µL to 2.00E4 µL
            
            
                Microchip
            
            
                Absolute expanded uncertainty
                :
                5 % to 0.1 %
            
            
            
            
                Gravimetric method
            
            
            
            
            
                Approved on 09 December 2024
            
            
                Institute service identifier :  103.01
            
            
                CMC ID :
                EURAMET-M-PT-00000PRF-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                1.00E-3 mL to 200 mL
            
            
                Piston operating apparatus (burettes, dispensers) and syringes
            
            
                Relative expanded uncertainty
                :
                2.0E-2 % to 1.0E-2 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Reference temperature : 20 °C
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.09, 103.12, 103.04
            
            
                CMC ID :
                EURAMET-M-PT-00000C9G-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                1 L to 20 L
            
            
                Standard test measures and overflow pipettes
            
            
                Relative expanded uncertainty
                :
                1.0E-2 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Reference temperature : 20 °C
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.21, 103.20, 103.13
            
            
                CMC ID :
                EURAMET-M-PT-00000C9D-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                Impedance (up to the MHz range) ,
				 Capacitance: low loss capacitors
                :
                1 pF to 1000 pF
            
            
                Standard capacitor (sealed, dry-nitrogen or fused silica dielectric)
            
            
                Relative expanded uncertainty
                :
                1.0E1 µF/F
            
            
            
            
                Comparison
            
            
            
                Frequency : 100 Hz, 400 Hz, 1000 Hz
            
            
            
                Approved on 06 August 2013
            
            
                Institute service identifier :  IPQ/404.21
            
            
                CMC ID :
                EURAMET-EM-PT-0000069Z-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                20 L to 1500 L
            
            
                Proving tanks and overflow pipettes
            
            
                Relative expanded uncertainty
                :
                1.0E-2 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Reference temperature : 20 °C
            
            
            
                Approved on 19 October 2016
            
            
                Institute service identifier :  IPQ/103.21, 103.20 and 103.13
            
            
                CMC ID :
                EURAMET-M-PT-00000C9V-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                1 L to 20 L
            
            
                Standard test measures
            
            
                Relative expanded uncertainty
                :
                2.0E-2 %
            
            
            
            
                Volumetric
            
            
            
                Liquid : water
            
            
                Ambient temperature : 10°C to 30 °C
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.21
            
            
                CMC ID :
                EURAMET-M-PT-00000C9E-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                20 L to 1.00E4 L
            
            
                Proving tanks
            
            
                Relative expanded uncertainty
                :
                2.0E-2 %
            
            
            
            
                Volumetric
            
            
            
                Liquid : water
            
            
                Ambient temperature : 10°C to 30 °C
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.21
            
            
                CMC ID :
                EURAMET-M-PT-00000C9C-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                DC voltage (up to 1100 V) ,
				 DC voltage sources: low values
                :
                1.00E-3 V to 10 V
            
            
                Calibrator
            
            
                Relative expanded uncertainty
                :
                3 µV/V to 5.0E1 µV/V
            
            
                Uncertainty table
                
                    {"row_1": {"col_1":"<strong>Voltage<\/strong>", "col_2":"<strong>Expanded uncertainty<br>/ (µV/V)<\/strong>"}, "row_2": {"col_1":"1 mV to 2 V", "col_2":"50"}, "row_3": {"col_1":"2 V to 10 V", "col_2":"3"}, "row_4": {"col_1":"10 V to 20 V", "col_2":"3"}, "row_5": {"col_1":"20 V to 1000 V", "col_2":"5"}}
                    
                    
                 
             
            
            
                Comparison to a reference with a resistive divider
            
            
            
            
            
                Approved on 06 August 2013
            
            
                Institute service identifier :  IPQ/401.12
            
            
                CMC ID :
                EURAMET-EM-PT-0000069T-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                DC resistance ,
				 DC resistance meters: low values
                :
                0.01 Ω to 1 Ω
            
            
                Resistance meter
            
            
                Relative expanded uncertainty
                :
                7 µΩ/Ω to 2 µΩ/Ω
            
            
                Uncertainty table
                
                    {"row_1": {"col_1":"Resistance", "col_2":"Expanded uncertainty<br>/ (µΩ/Ω)"}, "row_2": {"col_1":"10 mΩ", "col_2":"7"}, "row_3": {"col_1":"100 mΩ", "col_2":"5"}, "row_4": {"col_1":"1 Ω", "col_2":"1.5"}, "row_5": {"col_1":"10 Ω to 100 kΩ", "col_2":"2"}, "row_6": {"col_1":"1 MΩ", "col_2":"5"}, "row_7": {"col_1":"10 MΩ", "col_2":"7"}, "row_8": {"col_1":"100 MΩ", "col_2":"25"}}
                    
                    
                 
             
            
            
                Direct measurement
            
            
            
                Air temperature : 23 °C
            
            
            
                Approved on 05 April 2021
            
            
                Institute service identifier :  IPQ/409.52
            
            
                CMC ID :
                EURAMET-EM-PT-00000MHL-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                0.1 µL to 1000 µL
            
            
                Micropipettes or piston pipettes
            
            
                Relative expanded uncertainty
                :
                2.7 % to 0.4 %
            
            
            
            
                Photometric
            
            
            
                Liquid : Aqueous solutions
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.02
            
            
                CMC ID :
                EURAMET-M-PT-00000C9J-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Volume of liquid
                :
                1 mL to 100 mL
            
            
                Pycnometers
            
            
                Relative expanded uncertainty
                :
                3.0E-3 %
            
            
            
            
                Gravimetric
            
            
            
                Liquid : water
            
            
                Reference temperature : 20 °C
            
            
            
                Approved on 11 September 2019
            
            
                Institute service identifier :  IPQ/103.10
            
            
                CMC ID :
                EURAMET-M-PT-00000C9T-2 
             
         
        
        
     
 
					 
					
						
    
        
            
                DC current (up to 100 A) ,
				 DC current meters: low values
                :
                1.00E-13 A to 1.00E-9 A
            
            
                Picoammeter, electrometer
            
            
                Relative expanded uncertainty
                :
                1.5E2 µA/A to 2.0E3 µA/A
            
            
                Uncertainty table
                
                    {"row_1": {"col_1":"-", "col_2":"Expanded uncertainty<br>/ (µA/A)"}, "row_2": {"col_1":"1E-13", "col_2":"2000"}, "row_3": {"col_1":"1E-12", "col_2":"500"}, "row_4": {"col_1":"1E-11", "col_2":"200"}, "row_5": {"col_1":"1E-10", "col_2":"150"}, "row_6": {"col_1":"1E-9", "col_2":"150"}}
                    
                    
                 
             
            
            
                Charging of a capacitor
            
            
            
            
            
                Approved on 31 March 2022
            
            
                Institute service identifier :  IPQ/403.21
            
            
                CMC ID :
                EURAMET-EM-PT-00000MHG-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Fluid flow ,
				 Liquid flow
                :
                1.20E-3 mL/h to 0.12 mL/h
            
            
                Flowmeter
            
            
                Relative expanded uncertainty
                :
                5.2 % to 2.5 %
            
            
            
            
                Interferometric method
            
            
            
                Reference Temperature : 20 ºC
            
            
                Liquid : Water
            
            
                Pressure : up to 3 hPa
            
            
            
                Approved on 09 July 2022
            
            
                Institute service identifier :  104.61
            
            
                CMC ID :
                EURAMET-M-PT-00000NW8-1