Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241 and 2006 COOMET.RI(II)-K2.Am-241

 

MEASURAND   SIR equivalent activity of 241Am

 

Degrees of equivalence represented by Di = (xI - x) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Equivalent activity of 241Am
Di / MBq
Participating institutes

Results in absolute values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241 and 2006 COOMET.RI(II)-K2.Am-241

 

MEASURAND   SIR equivalent activity of 241Am

 

Degrees of equivalence represented by Di = (xI - x) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
VNIIM -3 14
POLATOM 4 25
MKEH -18 11
LNE-LNHB -13 13
BelGIM 4 48
CENTIS-DMR -13 27
Participating institutes
BIPM.RI(II)-K1.Am-241
COOMET.RI(II)-K2.Am-241

Results in absolute values

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Am-241
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Am-241
Time of measurements 1977 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
Results last updated on 04 July 2024
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP

BFKH

Government Office of the Capital City Budapest, Hungary, EURAMET

comparison carried out with MKEH (Hungarian Trade Licensing Office)

comparison carried out with MKEH (Hungarian Trade Licensing Office)

CMI

Czech Metrology Institute, Czechia, EURAMET

comparison carried out with CMI-IIR (Czech Metrology Institute - Inspectorate for Ionizing Radiation)

CNEA

National Commission of Atomic Energy, Argentina, SIM

Report in progress, Draft B

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

NPL

National Physical Laboratory, United Kingdom, EURAMET

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

PTKMR-BATAN

Center for Technology of Radiation Safety and Metrology, National Nuclear Energy Agency, Indonesia, APMP

comparison carried out with P3KRBiN (Pusat Penelitian & Pengembangan Keselamatan Radiasi & Biomedika Nuklir)

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241 and 2006 COOMET.RI(II)-K2.Am-241

 

MEASURAND   SIR equivalent activity of 241Am

 

BIPM.RI(II)-K1.Am-241

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 2055.8 MBq, with a standard uncertainty uR = 2.8 MBq

(see Section 4.1 of the BIPM.RI(II)-K1.Am-241 Final report dated November 2007)

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (xi - xR) and Ui  its expanded uncertainty (k = 2) both expressed in MBq. With n the number of laboratories, 

Ui = 2[(1 - 2/n)ui2 + (1/n2ui2]1/2 when each laboratory has contributed to the calculation of xR.

 

Linking COOMET.RI(II)-K2.Am-241 to BIPM.RI(II)-K1.Am-241

The value xi is the equivalent activity for laboratory i participant in COOMET.RI(II)-K2.Am-241 having been normalized using the SIR measurement of the VNIIM solution of the COOMET.RI(II)-K2 comparison.

The degree of equivalence of laboratory i participant in COOMET.RI(II)-K2.Am-241 with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and U, its expanded uncertainty (k = 2), both expressed in MBq.

The approximation Ui = 2(ui2 + uR2)1/2 is used in the Matrix of equivalence as none of these laboratories contributed to the key comparison reference value.

These statements make it possible to extend the BIPM.RI(II)-K1.Am-241 graph of equivalence to all participants in the COOMET.RI(II)-K2.Am-241 comparison.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241 and 2006 COOMET.RI(II)-K2.Am-241

 

MEASURAND   SIR equivalent activity of 241Am

 

xi   measured mass fraction reported by laboratory iLabi

ui   combined standard uncertainty of xi

 

BIPM.RI(II)-K1.Am-24

  Labi xi ui Date of measurement
  MBq MBq
  VNIIM 2052.6 7.7 2006-08-03
  POLATOM 2060.0 12.0 2009-01-28
  MKEH 2038.0 4.3 2009-02-25
  LNE-LNHB 2042.7 5.8 2011-05-05

 

COOMET.RI(II)-K2.Am-241

  Labi xi ui Date of measurement
  MBq MBq
  BelGIM 2060 24 2006-06-01
  CENTIS-DMR 2043 13 2006-06-01