CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 10 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 20 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

Red diamonds: participants in CCM.P-K6
Green triangles: participants in APMP.M.P-K6 only
Blue circles: participants in APMP.M.P-K6.1 only

Orange squares: participants in EURAMET.M.P-K8

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 30 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

From 1st March 2009, NMi-VSL has been named VSL

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 40 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

Red diamonds: participants in CCM.P-K6
Green triangles: participants in APMP.M.P-K6 only
Blue circles: participants in APMP.M.P-K6.1 only

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL PRESSURE : p = 50 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

Red diamonds: participants in CCM.P-K6

Orange squares: participants in EURAMET.M.P-K8
 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 60 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

Red diamonds: participants in CCM.P-K6
Green triangles: participants in APMP.M.P-K6 only
Blue circles: participants in APMP.M.P-K6.1 only

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~
70 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2 for participants in CCM.P-K6

From 1st March 2009, NMi-VSL has been named VSL

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2 for participants in EURAMET.M.P-K8

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 80 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

Red diamonds: participants in CCM.P-K6
Green triangles: participants in APMP.M.P-K6 only
Blue circles: participants in APMP.M.P-K6.1 only

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6  and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 90 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

From 1st March 2009, NMi-VSL has been named VSL

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 100 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

Red diamonds: participants in CCM.P-K6
Green triangles: participants in APMP.M.P-K6 only
Blue circles: participants in APMP.M.P-K6.1 only
Orange squares: partcipants in EURAMET.M.P-K8

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 110 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

From 1st March 2009, NMi-VSL has been named VSL

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 120 kPa

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2),
expressed in mm2

From 1st March 2009, NMi-VSL has been named VSL

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 10 kPa

 

 

     
Lab i Di Ui
  / mm2 / mm2
METAS -0.0451 0.075
PTB 0.0002 0.0055
NIST -0.0002 0.0036
NIM 0.0029 0.0041
NPL -0.0021 0.0100
NRC 0.0014 0.0040

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 20 kPa

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value, expressed in mm2

* participants in APMP.M.P-K6 only
**participants in APMP.M.P-K6.1 only

Orange : participants in EURAMET.M.P-K8

Matrix of equivalence inside APMP.M.P-K6.1:

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0010 0.011
METAS -0.0261 0.050
PTB 0.0000 0.0041
NIST -0.0011 0.0026
NIM 0.0018 0.0032
NPL -0.0005 0.0048
NRC 0.0002 0.0027

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 30 kPa

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0015 0.011
METAS -0.0205 0.035
PTB 0.0000 0.0038
NIST -0.0013 0.0028
NIM 0.0015 0.0033
NPL -0.0002 0.0038
NRC 0.0000 0.0029

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 40 kPa

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value, expressed in mm2

 

* participants in APMP.M.P-K6 only
**participants in APMP.M.P-K6.1 only

Matrix of equivalence inside APMP.M.P-K6.1:

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0017 0.011
METAS -0.0234 0.033
PTB 0.0005 0.0033
NIST -0.0011 0.0027
NIM 0.0012 0.0033
NPL 0.0000 0.0030
NRC -0.0002 0.0027

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

 

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL PRESSURE : p = 50 kPa

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0012 0.011
METAS -0.0167 0.026
PTB 0.0005 0.0032
NIST -0.0013 0.0026
NIM 0.0000 0.0032
NPL 0.0000 0.0025
NRC -0.0010 0.0025
METAS* -0.0020 0.0061
LNE* -0.0041 0.0055
PTB* 0.0006 0.0033
NSAI NML* -0.0018 0.0117
MIKES* -0.0043 0.0069
VSL* -0.0001 0.0100
NIS* -0.0049 0.0078
INRIM* -0.0033 0.0073
UME* -0.0030 0.0059
EIM* -0.0038 0.0086
CMI* -0.0026 0.0061
BEV* -0.0046 0.0098
MIRS/IMT/LMT* -0.0027 0.0053
HMI/FSB-LPM* 0.0002 0.0108
IMBIH* -0.0018 0.0138
CEM* -0.0056 0.0107
KRISS* -0.0024 0.0077
INM(RO)* -0.0038 0.0098
FORCE* -0.0185 0.0174
IPQ* 0.0021 0.0104
MKEH* -0.0099 0.0086
MCCAA-SMI* -0.0022 0.0060

 

 

*Participants in EURAMET.M.P-K8

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 60 kPa

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value, expressed in mm2

 

* participants in APMP.M.P-K6 only
**participants in APMP.M.P-K6.1 only

Matrix of equivalence inside APMP.M.P-K6.1:

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0011 0.011
METAS -0.0219 0.023
PTB 0.0006 0.0032
NIST -0.0013 0.0023
NIM 0.0003 0.0028
NPL 0.0000 0.0020
NRC -0.0005 0.0021

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

 

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 70 kPa

Matrix of equivalence inside CCM.P-K6:

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0010 0.011
METAS -0.0158 0.018
PTB 0.0010 0.0037
NIST -0.0014 0.0026
NIM 0.0009 0.0032
NPL 0.0000 0.0022
NRC -0.0010 0.0024

 

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value, expressed in mm2 for the participants in EURAMET.M.P-K8.

Orange : Participants in EURAMET.M.P-K8

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

 

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 80 kPa

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value, expressed in mm2

 

* participants in APMP.M.P-K6 only
**participants in APMP.M.P-K6.1 only

Matrix of equivalence inside APMP.M.P-K6.1:

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0007 0.011
METAS -0.0113 0.023
PTB 0.0008 0.0038
NIST -0.0019 0.0024
NIM 0.0008 0.0031
NPL 0.0000 0.0019
NRC -0.0012 0.0022

 

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6  and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 90 kPa

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0006 0.011
METAS -0.0110 0.018
PTB 0.0009 0.0039
NIST -0.0017 0.0025
NIM 0.0010 0.0031
NPL 0.0000 0.0019
NRC -0.0008 0.0024

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and EURAMET.M.P-K8

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 100 kPa

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value, expressed in mm2

* participants in APMP.M.P-K6 only
**participants in APMP.M.P-K6.1 only
Orange: participants in EURAMET.M.P-K8

 

Matrix of equivalence inside APMP.M.P-K6.1:

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0005 0.011
METAS -0.0142 0.016
PTB 0.0008 0.0032
NIST -0.0016 0.0027
NIM 0.0012 0.0033
NPL 0.0000 0.0022
NRC -0.0014 0.0026

 

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 110 kPa

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0005 0.011
METAS -0.0117 0.019
PTB 0.0010 0.0027
NIST -0.0015 0.0025
NIM 0.0011 0.0032
NPL 0.0000 0.0019
NRC -0.0007 0.0024

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 120 kPa

 

 

     
Lab i Di Ui
  / mm2 / mm2
VSL 0.0009 0.011
METAS -0.0103 0.016
PTB 0.0011 0.0027
NIST -0.0012 0.0028
NIM 0.0022 0.0033
NRC -0.0009 0.0027

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

 

Comparison
Comparison close
CC comparison
Linked comparison
SIM.M.P-K6
Metrology area, Sub-field Mass and related quantities, Pressure
Description Pneumatic pressure measurements (gauge mode)
Time of measurements 2008 - 2011
Status Approved for equivalence
Final Reports of the comparisons
Measurand Effective area of a piston-cylinder unit
Parameters Pressure: 10 kPa to 120 kPa
Transfer device Piston-cylinder unit
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by SIM (Inter-American Metrology System)
RMO Internal Identifier SIM P-7.039
Pilot institute CENAM
Centro Nacional de Metrologia
Mexico
Contact person Jorge C. Torres-Guzman

+52 442 2110572
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
CENAM

Centro Nacional de Metrologia, Mexico, SIM

BSJ

Bureau of Standards Jamaica, Jamaica, SIM

CENAMEP AIP

Centro Nacional de Metrología de Panama AIP, Panama, SIM

INDECOPI

Instituto Nacional de Defensa de la Competencia y de la Protección de la Propiedad Intelectual, Peru, SIM

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

INTI

Instituto Nacional de Tecnologia Industrial, Argentina, SIM

LACOMET

Laboratorio Costarricense de Metrologia, Costa Rica, SIM

LCPN-P

Laboratorio Custodio de Patrones Nacionales de Presion, Chile, SIM

NIST

National Institute of Standards and Technology, United States, SIM

NRC

National Research Council, Canada, SIM

SIC

Superintendencia de Industria y Comercio, Colombia, SIM

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 10 kPa

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 2 mm2
uR = 0.001 5 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

 

The results issued by SIM.M.P-K6 are traceable to CCM.P-K6. 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 20 kPa

  • Key comparison CCM.P-K6

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 4 mm2
uR = 0.000 7 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).
  • Linking APMP.M.P-K6 to CCM.P-K6

PTB is the common participant to both key comparisons CCM.P-K6 and APMP.M.P-K6: its results are used to compute the linkage as explained in Sections 5.1 and 5.2 of the APMP.M.P-K6 Final Report.

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6 only, as explained in Section 5.3 of the APMP.M.P-K6 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6 are obtained as explained in Section 5.4 of the APMP.M.P-K6 Final Report. They are expressed in relative terms.
  • Linking APMP.M.P-K6.1 to CCM.P-K6

The key comparison APMP.M.P-K6.1 is linked to the CCM.P-K6 key comparison through the APMP.M.P-K6 key comparison.
A*STAR is the common participant to both key comparisons APMP.M.P-K6 and APMP.M.P-K6.1: its results are used to compute the linkage as explained in paragraph 4.1 the APMP.M.P-K6.1 Final Report.

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6.1 only, as explained in paragraph 4.2 of the APMP.M.P-K6.1 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6.1 are obtained as explained in paragraph 4.3 of the APMP.M.P-K6.1 Final Report. They are expressed in relative terms.
  • Linking EURAMET.M.P-K8 to CCM.P-K6

PTB and VSL provide the link between key comparisons CCM.P-K6 and EURAMET.M.P-K8 for the following nominal pressures:  25 kPa, 50 kPa, 75 kPa and 100 kPa.
The linkage process is described in Section 11 of the EURAMET.M.P-K8 Final Report.

The measurements of EURAMET.M.P-K8 at the nominal pressure 25 kPa are linked with the averaged value of the 20 kPa and 30 kPa measurements of the CCM comparison, and those at the nominal pressure 75 kPa are linked with the averaged value of the 70 kPa and 80 kPa measurements of the CCM comparison.

Therefore, the degrees of equivalence of the laboratories having participated in EURAMET.M.P-K8 for 25 kPa and 75 kPa are shown on a different graph from the ones for the participants in CCM.P-K6.

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 30 kPa

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 2 mm2
uR = 0.001 0 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

The results issued by SIM.M.P-K6 are traceable to NCCM.P-K6. 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 40 kPa

  • Key comparison CCM.P-K6

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 1 mm2
uR = 0.000 9 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

  • Linking APMP.M.P-K6 to CCM.P-K6

PTB is the common participant to both key comparisons CCM.P-K6 and APMP.M.P-K6: its results are used to compute the linkage as explained in Sections 5.1 and 5.2 of the APMP.M.P-K6 Final Report.

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6 only, as explained in Section 5.3 of the APMP.M.P-K6 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6 are obtained as explained in Section 5.4 of the APMP.M.P-K6 Final Report. They are expressed in relative terms.

 

  • Linking APMP.M.P-K6.1 to CCM.P-K6

The key comparison APMP.M.P-K6.1 is linked to the CCM.P-K6 key comparison through the APMP.M.P-K6 key comparison.
A*STAR is the common participant to both key comparisons APMP.M.P-K6 and APMP.M.P-K6.1: its results are used to compute the linkage as explained in paragraph 4.1 the APMP.M.P-K6.1 Final Report.

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6.1 only, as explained in paragraph 4.2 of the APMP.M.P-K6.1 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6.1 are obtained as explained in paragraph 4.3 of the APMP.M.P-K6.1 Final Report. They are expressed in relative terms.

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. 

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 50 kPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 3 mm2
uR = 0.000 8 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).
  • Linking EURAMET.M.P-K8 to CCM.P-K6

PTB and VSL provide the link between key comparisons CCM.P-K6 and EURAMET.M.P-K8 for the following nominal pressures: 25 kPa, 50 kPa, 75 kPa and 100 kPa.
The linkage process is described in Section 11 of the EURAMET.M.P-K8 Final Report.

The measurements of EURAMET.M.P-K8 at the nominal pressure 25 kPa are linked with the averaged value of the 20 kPa and 30 kPa measurements of the CCM comparison, and those at the nominal pressure 75 kPa are linked with the averaged value of the 70 kPa and 80 kPa measurements of the CCM comparison.

Therefore, the degrees of equivalence of the laboratories having participated in EURAMET.M.P-K8 for 25 kPa and 75 kPa are shown on a different graph from the ones for the participants in CCM.P-K6.

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 60 kPa

  • Key comparison CCM.P-K6

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 3 mm2
uR = 0.000 5 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

  • Linking APMP.M.P-K6 to CCM.P-K6

PTB is the common participant to both key comparisons CCM.P-K6 and APMP.M.P-K6: its results are used to compute the linkage as explained in Sections 5.1 and 5.2 of the APMP.M.P-K6 Final Report.

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6 only, as explained in Section 5.3 of the APMP.M.P-K6 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6 are obtained as explained in Section 5.4 of the APMP.M.P-K6 Final Report. They are expressed in relative terms.

 

  • Linking APMP.M.P-K6.1 to CCM.P-K6

The key comparison APMP.M.P-K6.1 is linked to the CCM.P-K6 key comparison through the APMP.M.P-K6 key comparison.
A*STAR is the common participant to both key comparisons APMP.M.P-K6 and APMP.M.P-K6.1: its results are used to compute the linkage as explained in paragraph 4.1 the APMP.M.P-K6.1 Final Report.

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6.1 only, as explained in paragraph 4.2 of the APMP.M.P-K6.1 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6.1 are obtained as explained in paragraph 4.3 of the APMP.M.P-K6.1 Final Report. They are expressed in relative terms.

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 70 kPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 3 mm2
uR = 0.000 8 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).
  • Linking EURAMET.M.P-K8 to CCM.P-K6

PTB and VSL provide the link between key comparisons CCM.P-K6 and EURAMET.M.P-K8 for the following nominal pressures: 25 kPa, 50 kPa, 75 kPa and 100 kPa.
The linkage process is described in Section 11 of the EURAMET.M.P-K8 Final Report.

The measurements of EURAMET.M.P-K8 at the nominal pressure 25 kPa are linked with the averaged value of the 20 kPa and 30 kPa measurements of the CCM comparison, and those at the nominal pressure 75 kPa are linked with the averaged value of the 70 kPa and 80 kPa measurements of the CCM comparison.

Therefore, the degrees of equivalence of the laboratories having participated in EURAMET.M.P-K8 for 25 kPa and 75 kPa are shown on a different graph from the ones for the participants in CCM.P-K6.

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 80 kPa

  • Key comparison CCM.P-K6

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 5 mm2
uR = 0.000 6 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

  • Linking APMP.M.P-K6 to CCM.P-K6

PTB is the common participant to both key comparisons CCM.P-K6 and APMP.M.P-K6: its results are used to compute the linkage as explained in Sections 5.1 and 5.2 of the APMP.M.P-K6 Final Report.

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6 only, as explained in Section 5.3 of the APMP.M.P-K6 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6 are obtained as explained in Section 5.4 of the APMP.M.P-K6 Final Report. They are expressed in relative terms.

 

  • Linking APMP.M.P-K6.1 to CCM.P-K6

The key comparison APMP.M.P-K6.1 is linked to the CCM.P-K6 key comparison through the APMP.M.P-K6 key comparison.
A*STAR is the common participant to both key comparisons APMP.M.P-K6 and APMP.M.P-K6.1: its results are used to compute the linkage as explained in paragraph 4.1 the APMP.M.P-K6.1 Final Report.

 

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6.1 only, as explained in paragraph 4.2 of the APMP.M.P-K6.1 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6.1 are obtained as explained in paragraph 4.3 of the APMP.M.P-K6.1 Final Report. They are expressed in relative terms.

 

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6.

CCM.P-K6  and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 90 kPa

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 5 mm2
uR = 0.000 7 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p ~ 100 kPa

  • Key comparison CCM.P-K6

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 5 mm2
uR = 0.000 9 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).
  • Linking APMP.M.P-K6 to CCM.P-K6

PTB is the common participant to both key comparisons CCM.P-K6 and APMP.M.P-K6: its results are used to compute the linkage as explained in Sections 5.1 and 5.2 of the APMP.M.P-K6 Final Report.

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6 only, as explained in Section 5.3 of the APMP.M.P-K6 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6 are obtained as explained in Section 5.4 of the APMP.M.P-K6 Final Report. They are expressed in relative terms.
  • Linking APMP.M.P-K6.1 to CCM.P-K6

The key comparison APMP.M.P-K6.1 is linked to the CCM.P-K6 key comparison through the APMP.M.P-K6 key comparison.
A*STAR is the common participant to both key comparisons APMP.M.P-K6 and APMP.M.P-K6.1: its results are used to compute the linkage as explained in paragraph 4.1 the APMP.M.P-K6.1 Final Report.

Degrees of equivalence relative to the CCM.P-K6 key comparison reference value are obtained for laboratories having participated in APMP.M.P-K6.1 only, as explained in paragraph 4.2 of the APMP.M.P-K6.1 Final Report. They are expressed in mm2.
Pair-wise degrees of equivalence among pairs of participants in APMP.M.P-K6.1 are obtained as explained in paragraph 4.3 of the APMP.M.P-K6.1 Final Report. They are expressed in relative terms.
  • Linking EURAMET.M.P-K8 to CCM.P-K6

PTB and VSL provide the link between key comparisons CCM.P-K6 and EURAMET.M.P-K8 for the following nominal pressures: 25 kPa, 50 kPa, 75 kPa and 100 kPa.
The linkage process is described in Section 11 of the EURAMET.M.P-K8 Final Report.

The measurements of EURAMET.M.P-K8 at the nominal pressure 25 kPa are linked with the averaged value of the 20 kPa and 30 kPa measurements of the CCM comparison, and those at the nominal pressure 75 kPa are linked with the averaged value of the 70 kPa and 80 kPa measurements of the CCM comparison.

Therefore, the degrees of equivalence of the laboratories having participated in EURAMET.M.P-K8 for 25 kPa and 75 kPa are shown on a different graph from the ones for the participants in CCM.P-K6.

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6.

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 110 kPa

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 4 mm2
uR = 0.000 7 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. 

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 120 kPa

 

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MAD / (n - 1)1/2, where MAD is the median of absolute deviations from the median, and n is the number of participants contributing to the reference value.

xR = 335.744 1 mm2
uR = 0.001 0 mm2

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms expressed in mm2:
Di = (xi - xR) and Ui = 2(ui2 + uR2)1/2, its expanded uncertainty (k = 2).
The degree of equivalence between two laboratories i and j is given by a pair of terms expressed in mm2:
Dij = Di - Dj = (xi - xj) and Uij = 2(ui2 + uj2)1/2, its expanded uncertainty (k = 2).

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6. However, no degrees of equivalence have been indicated expricitly.

CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 10 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

-

-

-

 METAS

335.699 2

110

Nov-Dec 1998

 PTB 335.744 4 7.0 Feb-Mar 1999
 NIST 335.744 0 3.2 Sep-Oct 1999
 NIM

335.747 1

4.3

Nov-Dec 1999

 NPL 335.742 1 15 Aug-Sep 2000
 NRC 335.745 6 4.2 Jul-Sep 2001

 

 

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 20 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 4

16

Aug-Sep 1998

 METAS

335.718 3

74

Nov-Dec 1998

 PTB 335.744 4 5.6 Feb-Mar 1999
 NIST 335.743 3 3.1 Sep-Oct 1999
 NIM

335.746 2

4.3

Nov-Dec 1999

 NPL 335.743 9 6.8 Aug-Sep 2000
 NRC 335.744 6 3.4 Jul-Sep 2001

 

 

  • Key comparison APMP.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 21.4 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NPLI

335.737 3

15.80

Oct 1998

 KRISS

335.742 1

11.10

Apr 1999

 NMIA 335.749 7 12.60 Feb 2001
 NMIJ 335.738 8 8.80 Sep 1999
 PTB

335.737 6

6.40

Mar 2001

 MSL 335.749 3 11.80 May 1999
 A*STAR 335.737 5 19.20 Nov 1998
 NML-SIRIM

335.744 8

11.40

Jan 1999

 SCL

335.747 7

18.00

Jun 1999

 NMISA

335.740 7

14.60

Oct 1999

  • Key comparison APMP.M.P-K6.1

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.75 mm2
NOMINAL PRESSURE : p = 21.4 kPa

xi : result obtained from the mean of measurements by laboratory i
ULabi : relative expanded uncertainty (k =2) of xi

 Lab i xi
/ mm2
ULabi
/ 10-6
Date of measurement
 CMS/ITRI

335.623 5

30.2

Apr 03 - Apr 04

 A*STAR

335.633 9

22.0

Apr 03 - Apr 04

 NIMT 335.635 1 18.6 Apr 03 - Apr 04

 

  • Key comparison EURAMET.M.P-K8

MEASURAND : Effective area of a piston-cylinder assembly determined in gauge pressure mode
NOMINAL VALUE : 980 mm2
NOMINAL PRESSURE : p = 25 kPa

The laboratory individual measurements of the participants in the key comparison EURAMET.M.P-K8 are given in Table 7 on page 26 of the EURAMET.M.P-K8 Final Report.

 

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 30 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 7

16

Aug-Sep 1998

 METAS

335.723 7

51

Nov-Dec 1998

 PTB 335.744 2 4.9 Feb-Mar 1999
 NIST 335.742 9 3.0 Sep-Oct 1999
 NIM

335.745 7

4.0

Nov-Dec 1999

 NPL 335.744 0 4.8 Aug-Sep 2000
 NRC 335.744 2 3.1 Jul-Sep 2001

 

 

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 40 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 8

16

Aug-Sep 1998

 METAS

335.720 7

49

Nov-Dec 1998

 PTB 335.744 6 4.1 Feb-Mar 1999
 NIST 335.743 0 3.0 Sep-Oct 1999
 NIM

335.745 3

4.2

Nov-Dec 1999

 NPL 335.744 1 3.7 Aug-Sep 2000
 NRC 335.743 9 3.0 Jul-Sep 2001

 

 

 

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 50 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 5

16

Aug-Sep 1998

 METAS

335.727 6

39

Nov-Dec 1998

 PTB 335.744 8 4.2 Feb-Mar 1999
 NIST 335.743 0 3.1 Sep-Oct 1999
 NIM

335.744 3

4.2

Nov-Dec 1999

 NPL 335.744 3 3.0 Aug-Sep 2000
 NRC 335.743 3 2.9 Jul-Sep 2001

 

 

  • Key comparison EURAMET.M.P-K8

MEASURAND : Effective area of a piston-cylinder assembly determined in gauge pressure mode
NOMINAL VALUE : 980 mm2
NOMINAL PRESSURE : p = 50 kPa

The laboratory individual measurements of the participants in the key comparison EURAMET.M.P-K8 are given in Table 7 on page 26 of the EURAMET.M.P-K8 Final Report.

 

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 60 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 4

16

Aug-Sep 1998

 METAS

335.722 4

34

Nov-Dec 1998

 PTB 335.744 9 4.5 Feb-Mar 1999
 NIST 335.743 0 3.1 Sep-Oct 1999
 NIM

335.744 6

4.0

Nov-Dec 1999

 NPL 335.744 3 2.6 Aug-Sep 2000
 NRC 335.743 8 2.8 Jul-Sep 2001

 

 

  • Key comparison APMP.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 61.3 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NPLI

335.737 8

15.80

Oct 1998

 KRISS

335.741 1

11.10

Apr 1999

 NMIA 335.742 4 12.60 Feb 2001
 NMIJ 335.738 3 8.80 Sep 1999
 PTB

335.738 7

6.20

Mar 2001

 MSL 335.740 6 11.70 May 1999
 A*STAR 335.735 0 19.20 Nov 1998
 NML-SIRIM

335.740 5

11.40

Jan 1999

 SCL

335.743 4

17.70

Jun 1999

 NMISA

335.732 2

14.60

Oct 1999

  • Key comparison APMP.M.P-K6.1

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.75 mm2
NOMINAL PRESSURE : p = 61.4 kPa

xi : result obtained from the mean of measurements by laboratory i
ULabi : relative expanded uncertainty (k =2) of xi

 Lab i xi
/ mm2
ULabi
/ 10-6
Date of measurement
 CMS/ITRI

335.624 2

30.0

Apr 03 - Apr 04

 A*STAR

335.632 9

21.4

Apr 03 - Apr 04

 NIMT 335.636 4 17.8 Apr 03 - Apr 04

 

The results issued by SIM.M.P-K6 are traceable to NIST and as such traceable to CCM.P-K6.

 

CCM.P-K6, EURAMET.M.P-K8 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 70 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 3

16

Aug-Sep 1998

 METAS

335.728 5

26

Nov-Dec 1998

 PTB 335.745 3 5.0 Feb-Mar 1999
 NIST 335.742 9 3.1 Sep-Oct 1999
 NIM

335.745 2

4.2

Nov-Dec 1999

 NPL 335.744 3 2.3 Aug-Sep 2000
 NRC 335.743 3 2.8 Jul-Sep 2001

 

 

  • Key comparison EURAMET.M.P-K8

MEASURAND : Effective area of a piston-cylinder assembly determined in gauge pressure mode
NOMINAL VALUE : 980 mm2
NOMINAL PRESSURE : p = 75 kPa

The laboratory individual measurements of the participants in the key comparison EURAMET.M.P-K8 are given in Table 7 on page 26 of the EURAMET.M.P-K8 Final Report.

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 80 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 2

16

Aug-Sep 1998

 METAS

335.733 2

34

Nov-Dec 1998

 PTB 335.745 3 5.3 Feb-Mar 1999
 NIST 335.742 6 3.1 Sep-Oct 1999
 NIM

335.745 3

4.2

Nov-Dec 1999

 NPL 335.744 5 2.1 Aug-Sep 2000
 NRC 335.743 3 2.8 Jul-Sep 2001

 

 

  • Key comparison APMP.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 81.3 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NPLI

335.737 7

15.80

Oct 1998

 KRISS

335.739 9

11.10

Apr 1999

 NMIA 335.741 1 12.60 Feb 2001
 NMIJ 335.737 8 8.80 Sep 1999
 PTB

335.739 3

5.90

Mar 2001

 MSL 335.740 3 11.70 May 1999
 A*STAR 335.734 8 19.20 Nov 1998
 NML-SIRIM

335.740 6

11.40

Jan 1999

 SCL

335.742 0

17.60

Jun 1999

 NMISA

335.731 4

14.60

Oct 1999

  •  

    Key comparison APMP.M.P-K6.1

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.75 mm2
NOMINAL PRESSURE : p = 81.4 kPa

xi : result obtained from the mean of measurements by laboratory i
ULabi : relative expanded uncertainty (k =2) of xi

 Lab i xi
/ mm2
ULabi
/ 10-6
Date of measurement
 CMS/ITRI

335.624 8

30.0

Apr 03 - Apr 04

 A*STAR

335.631 2

21.2

Apr 03 - Apr 04

 NIMT 335.636 8 17.8 Apr 03 - Apr 04

 

CCM.P-K6  and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 90 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 1

16

Aug-Sep 1998

 METAS

335.733 5

27

Nov-Dec 1998

 PTB 335.745 4 5.3 Feb-Mar 1999
 NIST 335.742 8 3.0 Sep-Oct 1999
 NIM

335.745 5

4.2

Nov-Dec 1999

 NPL 335.744 5 1.9 Aug-Sep 2000
 NRC 335.743 6 2.8 Jul-Sep 2001

 

 

CCM.P-K6, APMP.M.P-K6, APMP.M.P-K6.1, EURAMET.M.P-K8 and SIM.M.P-K6

  • Key comparison CCM.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 100 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 0

16

Aug-Sep 1998

 METAS

335.730 4

24

Nov-Dec 1998

 PTB 335.745 3 3.8 Feb-Mar 1999
 NIST 335.742 9 3.0 Sep-Oct 1999
 NIM

335.745 7

4.1

Nov-Dec 1999

 NPL 335.744 5 1.8 Aug-Sep 2000
 NRC 335.743 1 2.7 Jul-Sep 2001

 

 

  • Key comparison APMP.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 101.2 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NPLI

335.737 7

15.80

Oct 1998

 KRISS

335.739 7

11.10

Apr 1999

 NMIA 335.741 0 12.60 Feb 2001
 NMIJ 335.737 8 8.80 Sep 1999
 PTB

335.739 7

5.90

Mar 2001

 MSL 335.741 0 11.70 May 1999
 A*STAR 335.734 0 19.10 Nov 1998
 NML-SIRIM

335.742 0

11.40

Jan 1999

 SCL

335.742 0

17.60

Jun 1999

 NMISA

335.732 0

14.60

Oct 1999

  • Key comparison APMP.M.P-K6.1

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.75 mm2
NOMINAL PRESSURE : p = 101.4 kPa

xi : result obtained from the mean of measurements by laboratory i
ULabi : relative expanded uncertainty (k =2) of xi

 Lab i xi
/ mm2
ULabi
/ 10-6
Date of measurement
 CMS/ITRI

335.624 4

30.0

Apr 03 - Apr 04

 A*STAR

335.632 2

21.2

Apr 03 - Apr 04

 NIMT 335.636 7 17.8 Apr 03 - Apr 04
  • Key comparison EURAMET.M.P-K8

MEASURAND : Effective area of a piston-cylinder assembly determined in gauge pressure mode
NOMINAL VALUE : 980 mm2
NOMINAL PRESSURE : p = 100 kPa

The laboratory individual measurements of the participants in the key comparison EURAMET.M.P-K8 are given in Table 7 on page 26 of the EURAMET.M.P-K8 Final Report.

 

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 110 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 9

16

Aug-Sep 1998

 METAS

335.732 6

28

Nov-Dec 1998

 PTB 335.745 3 3.3 Feb-Mar 1999
 NIST 335.742 8 3.0 Sep-Oct 1999
 NIM

335.745 5

4.3

Nov-Dec 1999

 NPL 335.744 4 1.7 Aug-Sep 2000
 NRC 335.743 6 2.7 Jul-Sep 2001

 

 

CCM.P-K6 and SIM.M.P-K6

MEASURAND : Effective area of a piston-cylinder assembly
NOMINAL VALUE : 335.7 mm2
NOMINAL PRESSURE : p = 120 kPa

xi : result obtained from the mean of measurements by laboratory i
ui : combined standard uncertainty of xi

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 VSL

335.745 0

16

Aug-Sep 1998

 METAS

335.733 9

24

Nov-Dec 1998

 PTB 335.745 3 2.7 Feb-Mar 1999
 NIST 335.742 9 3.0 Sep-Oct 1999
 NIM

335.746 3

4.0

Nov-Dec 1999

 NPL - - -
 NRC 335.743 3 2.7 Jul-Sep 2001