Comparison
Comparison close

CCQM-K111, CCQM-K111.1, EURAMET.QM-K111, COOMET.QM-K111, APMP.QM-K111 and SIM.QM-K111

MEASURAND : Amount-of-substance fraction of Propane in Nitrogen
NOMINAL VALUE : 1000 µmol/mol

Degrees of equivalence, Di and its expanded uncertainty Ui (95 % level of confidence), expressed in µmol/mol

 

Comparison
Comparison close

CCQM-K111, CCQM-K111.1, EURAMET.QM-K111, COOMET.QM-K111, APMP.QM-K111 and SIM.QM-K111

MEASURAND : Amount-of-substance fraction of Propane in Nitrogen
NOMINAL VALUE : 1000 µmol/mol

Degrees of equivalence, Di and expanded uncertainty Ui (95% level of confidence), expressed in µmol/mol

Red : CCQM-K111
Green : CCQM-K111.1
Blue : EURAMET.QM-K111
Orange : COOMET.QM-K111
Violet : APMP.QM-K111
Pink: SIM.QM-K111

 

Results are presented under A4 printable format in Summary Results (PDF file).

SIM.QM-K111
Metrology area, Sub-field Chemistry and Biology, Gases
Description Propane in nitrogen
Time of measurements 2014
Status Approved for equivalence
Final Reports of the comparisons
Measurand Amount-of-substance fraction of propane in nitrogen
Transfer device Dedicated samples
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by SIM (Inter-American Metrology System)
Comments

 

 

Results published on 25 June 2018

Amount-of-substance fraction of propane in nitrogen

CCQM-K111.1, EURAMET.QM-K111, COOMET.QM-K111, APMP.QM-K111 and SIM.QM-K111 are linked to CCQM-K111.

 

 

Pilot institute INMETRO
Instituto Nacional de Metrologia, Qualidade e Tecnologia
Brazil
Contact person Valnei da Cunha

+55 21 2679 9069
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

CENAM

Centro Nacional de Metrologia, Mexico, SIM

Comparison
Comparison close

CCQM-K111, CCQM-K111.1, EURAMET.QM-K111, COOMET.QM-K111, APMP.QM-K111 and SIM.QM-K111

  • Key comparison CCQM-K111

MEASURAND : Amount-of-substance fraction of Propane in Nitrogen
NOMINAL VALUE : 1000 µmol/mol

The key comparison reference value for each cylinder is identical to xprep i with the standard uncertainty uref i

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in µmol/mol:
Di = xLab i - xprep i and its expanded uncertainty (95% level of confidence) Ui computed as explained in the Final Report

  •  Linking of CCQM-K111.1 to CCQM-K111

The degrees of equivalence determined from the CCQM-K111.1 comparison are linked to the CCQM-K111 via VSL who participated in both comparisons.

  • Linking of EURAMET.QM-K111 to CCQM-K111

The degrees of equivalence determined from the EURAMET.QM-K111 comparison are linked to the CCQM-K111 via VSL who participated in both comparisons.

  • Linking of COOMET.QM-K111 to CCQM-K111

The degrees of equivalence determined from the COOMET.QM-K111 comparison are linked to the CCQM-K111 via VNIIM who participated in both comparisons.

  • Linking of APMP.QM-K111 to CCQM-K111

The degrees of equivalence determined from the APMP.QM-K111 comparison are linked to the CCQM-K111 via CERI and KRISS who participated in both comparisons.

  • Linking of SIM.QM-K111 to CCQM-K111

The link is made using the results from INMETRO having participated in both comparisons (See section 2.8 in SIM.QM-K111 Final Report).

Comparison
Comparison close

CCQM-K111, CCQM-K111.1, EURAMET.QM-K111, COOMET.QM-K111, APMP.QM-K111 and SIM.QM-K111

MEASURAND : Amount-of-substance fraction of Propane in Nitrogen
NOMINAL VALUE : 1000 µmol/mol

  • Key comparison CCQM-K111

xprep i : amount-of-substance fraction from preparation
uprep i : standard uncertainty of xprep i
u
ver :
standard uncertainty from verification
uref i : standard uncertainty of reference value
xLab i : result of laboratory i
U
Lab i :
stated expanded uncertainty of xLab i at a confidence level at 95 %
kLab i : stated coverage factor

  • Key comparison CCQM-K111.1

The numerical results of CCQM-K111.1 are listed on p. 4 of the Final Report.

  • Key comparison EURAMET.QM-K111

The numerical results of EURAMET-K111 are listed in Teble 4 of the Final Report.

  • Key comparison COOMET.QM-K111

The numerical results of COOMET.QM-K111 are listed in Annex 4 of the Final Report.

  • Key comparison APMP.QM-K111

The numerical results of APMP.QM-K111 are listed in Table 4 of the Final Report.

  • Key comparison SIM.QM-K111

The numerical results of SIM.QM-K111 are listed in Table 5 of the Final Report.