MEASURAND SIR equivalent activity of Ac-225
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui (k =2), both expressed in kBq.
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui (k =2), both expressed in kBq/MBq.
MEASURAND SIR equivalent activity of Ac-225
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui (k =2), both expressed in kBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq |
kBq |
kBq |
||
PTB | -280 | 560 | ||
POLATOM | 280 | 560 |
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui (k =2), both expressed in kBq/MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq/MBq |
kBq/MBq |
kBq/MBq |
||
PTB | -3.7 | 7.5 | ||
POLATOM | 3.8 | 7.5 |
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Activity of radionuclide Ac-225 |
Time of measurements | 2019 - |
Status | Continuous |
Final Reports of the comparisons | |
Measurand | SIR equivalent activity |
Transfer device | Radioactive solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | BIPM (Bureau International des Poids et Mesures) |
Comments | SIR: Système International de Référence, International Reference System First and most recent results published on 15 December 2022. The status for not yet completed comparisons is indicated in "Participants". |
Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
Contact person | Carine Michotte +33 (0) 1 45 07 70 85 |
Pilot laboratory | |
---|---|
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
NIST |
National Institute of Standards and Technology, United States, SIM |
2023 |
Report in progress, draft A |
POLATOM |
Radioisotope Centre Polatom, Poland, EURAMET |
2021 |
|
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
2019 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
MEASURAND SIR equivalent activity of Ac-225
The SIR reference value for this radionuclide is xR = 74 800 kBq, with a standard uncertainty uR = 280 kBq.
xR is calculated by using the power-moderated weighted mean of the SIR results (see Section 4.1 of the Final Report). The value xi is taken as the equivalent activity for a laboratory i.
The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: Di = (xi - xR) and Ui its expanded uncertainty (k = 2), both expressed in kBq, and Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.
The Degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
MEASURAND SIR equivalent activity of Ac-225
xi equivalent activity measured by the SIR for a laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
PTB | 74520 | 200 | 2019-06-21 | |
POLATOM | 75080 | 210 | 2021-07-06 |