Comparison
Comparison close
Results

MEASURAND   Mass fraction of GE

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/g.

Mass fraction of GE
Di
Participating institutes

MEASURAND   Mass fraction of total peptide related impurities

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/g.

Mass fraction of overall peptide-related impurities
Di
Participating institutes
Comparison
Comparison close
Results

MEASURAND   Mass fraction of GE

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/g.

LABi Di Uneg,i Upos,i
mg/g
mg/g
mg/g
INMETRO -30.8 72.4
LNE -10.0 73.6
NIM 55.0 67.9
BIPM 2.5 72.9
LGC -43.6 67.2
NMIJ -3.2 69.3
NRC 15.4 73.4
PTB 31.7 70.6
UME -24.9 76.4
HSA 0.5 70.3
Participating institutes
CCQM-K115.c GE

MEASURAND   Mass fraction of total peptide related impurities

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/g.

LABi Di Uneg,i Upos,i
mg/g
mg/g
mg/g
INMETRO -4.8 22.6
LNE 6.1 22.2
NIM -5.1 16.3
BIPM -8.9 16.5
LGC 35.6 19.5
NMIJ -5.1 17.0
NRC 11.2 26.8
PTB -11.7 16.6
UME 3.1 16.7
HSA -3.1 22.5
Participating institutes
CCQM-K115.c GE
Comparison
Comparison close
CC comparison
Linked comparison
CCQM-K115.c
Metrology area, Sub-field Chemistry and Biology, Protein Analysis
Description Peptide purity determination - Glycated hexapeptide of HbA1c (glcVHLTPE or GE)
Time of measurements 2018 - 2020
Status Approved for equivalence
Final Reports of the comparisons
Measurand Mass fraction of glycated hexapeptide of HbA1c in the material (in mg/g)
Transfer device Dedicated samples
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by CCQM (Consultative Committee for Amount of Substance)
Comments
 
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person R. Josephs

+33 (0) 1 45 07 70 55
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

HSA

Health Science Authority, Singapore, APMP

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

LGC

Laboratory of the Government Chemist, United Kingdom, EURAMET

LNE

Laboratoire national de métrologie et d'essais, France, EURAMET

NIM

National Institute of Metrology, China, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NRC

National Research Council, Canada, SIM

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

Comparison
Comparison close
Results

MEASURAND   Mass fraction of GE

The key comparison reference value xR and its associated standard uncertainty uR  are based on the DerSimonian-Laird variance-weighted mean (DSL) of the participants' results.

        xR = 628 mg/g      uR = 12 mg/g

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/g:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 (uR2 + ui2)1/2.

MEASURAND   Mass fraction of total peptide related impurities

The key comparison reference value xR and its associated standard uncertainty uR  are based on the DerSimonian-Laird variance-weighted mean (DSL) of the participants' results.

        xR = 45.4 mg/g      uR = 4.2 mg/g

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/g:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 (uR2 + ui2)1/2.

Comparison
Comparison close
Results

MEASURAND   Mass fraction of GE

 

xi    measurement value reported by participant laboratory i, Lab i

ui   combined standard uncertainty of xi

Labi xi ui
  mg/g mg/g
INMETRO 597.6 15.4
LNE 618.444 17.624
NIM 683.41 7.13
BIPM 630.9 14.5
LGC 584.8 6.4
NMIJ 625.2 6.3
NRC 643.8 14.9
PTB 660.1 7.5
UME 603.5 18.4
HSA 628.9 9.5

 

MEASURAND   Mass fraction of total peptide related impurities

 

xi    measurement value reported by participant laboratory i, Lab i

ui   combined standard uncertainty of xi

Labi xi ui
  mg/g mg/g
INMETRO 40.6 8.1
LNE 51.530 8.042
NIM 40.25 2.105
BIPM 36.5 1.8
LGC 81.0 6.1
NMIJ 40.3 2.7
NRC 56.6 11.1
PTB 33.7 1.3
UME 48.5 1.8
HSA 42.30 8