BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131
MEASURAND : Equivalent activity of I-131
Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
Results in absolute values
BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6 and EURAMET.RI(II)-K2
MEASURAND : Equivalent activity of I-131
Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value
Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Results in relative values
BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131
MEASURAND : Equivalent activity of I-131
Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
MBq |
MBq |
MBq |
||
CIEMAT | -0.22 | 0.37 | ||
PTB | -0.01 | 0.15 | ||
CMI | 0.00 | 0.23 | ||
IFIN-HH | 0.04 | 0.27 | ||
NMIJ | 0.06 | 0.19 | ||
NIST | -0.03 | 0.27 | ||
LNE-LNHB | -0.19 | 0.18 | ||
BEV | 0.34 | 0.35 | ||
POLATOM | -0.10 | 0.31 | ||
NMISA | 0.18 | 0.31 | ||
ANSTO | 0.08 | 0.45 | ||
BARC | -0.30 | 0.39 | ||
INER | 0.82 | 0.35 | ||
KRISS | 0.93 | 0.21 | ||
OAP | 0.5 | 1.3 | ||
PTKMR | -0.5 | 1.8 | ||
LNMRI/IRD | -0.2 | 1.1 | ||
ENEA-INMRI | -0.08 | 0.35 |
Results in absolute values
BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131
MEASURAND : Equivalent activity of I-131
Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value
Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq/MBq |
kBq/MBq |
kBq/MBq |
||
CIEMAT | -5.5 | 9.3 | ||
PTB | -0.3 | 3.7 | ||
CMI | 0.0 | 5.8 | ||
IFIN-HH | 1.0 | 6.8 | ||
NMIJ | 1.5 | 4.8 | ||
NIST | -0.8 | 6.8 | ||
LNE-LNHB | -4.6 | 4.4 | ||
BEV | 8.4 | 8.6 | ||
POLATOM | -2.5 | 7.8 | ||
NMISA | 4.4 | 7.8 | ||
ANSTO | 2 | 11 | ||
BARC | -7.5 | 10 | ||
INER | 20 | 9 | ||
KRISS | 23 | 5 | ||
OAP | 14 | 32 | ||
PTKMR | -12 | 45 | ||
LNMRI/IRD | -5 | 27 | ||
ENEA-INMRI | -2 | 8.6 |
Results in relative values
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Activity of radionuclide I-131 |
Time of measurements | 2009 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Activity concentration in kBq/g |
Transfer device | Radioactive solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | APMP (Asia Pacific Metrology Programme) |
Comments | Results published on 06 February 2014 APMP.RI(II)-K2.I-131 results are linked to those of BIPM.RI(II)-K1.I-131 SIR equivalent activity of the radionuclide 131I: |
Pilot institute |
NMIJ AIST
National Metrology Institute of Japan Japan |
Contact person | A. Yunoki +81 29 861 3470 |
Pilot laboratory | |
---|---|
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
ANSTO |
Australian Nuclear Science and Technology Organisation, Australia, APMP |
BARC |
Bhabha Atomic Research Centre, India, APMP |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
NRSL/INER |
Institute of Nuclear Energy Research, Chinese Taipei, APMP |
OAP |
Office of Atoms for Peace, Thailand, APMP |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131
BIPM.RI(II)-K1.I-131
MEASURAND : Equivalent activity of I-131
Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is xR = 40 371 kBq with a standard uncertainty uR = 34 kBq
(see Section 4.1 of the BIPM.RI(II)-K1.I-131Final Report). The value xi is taken as the equivalent activity for a laboratory i.
The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui its expanded uncertainty (k = 2), both expressed in MBq, and
Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.
The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Linking APMP.RI(II)-K2.I-131 to BIPM.RI(II)-K1.I-131
The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.I-131 having been normalized using the value of the NMIJ as the linking laboratory.
The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.
Linking CCRI(II)-S6.I-131 to BIPM.RI(II)-K1.I-131
The value xi is the equivalent activity for laboratory i participant in CCRI(II)-S6.I-131 having been normalized using the value of the CMI as the linking laboratory.
The degree of equivalence of laboratory i participant in CCRI(II)-S6.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..
Linking EURAMET.RI(II)-K2.I-131 to BIPM.RI(II)-K1.I-131
The value xi is the equivalent activity for laboratory i participant in EURAMET.RI(II)-K2.I-131 having been normalized to the value of the POLATOM as the linking laboratory.
The degree of equivalence of laboratory i participant in EURAMET.RI(II)-K2.I-131 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..
These statements make it possible to extend the BIPM.RI(II)-K1.I-131 matrices of equivalence to all participants in the APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131, and EURAMET.RI(II)-K2.I-131 comparisons
BIPM.RI(II)-K1.I-131, APMP.RI(II)-K2.I-131, CCRI(II)-S6.I-131 and EURAMET.RI(II)-K2.I-131
MEASURAND : Equivalent activity of I-131
Degrees of equivalence relative to the BIPM.RI(II)-K1.I-131 key comparison reference value
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
BIPM.RI(II)-K1.I-131
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
CIEMAT | 40 150 | 190 | 2002-06-26 | |
PTB | 40 357 | 77 | 2004-04-30 | |
CMI | 40 370 | 120 | 2006-08-22 | |
IFIN-HH | 40 410 | 140 | 2008-01-03 | |
NMIJ | 40 430 | 100 | 2009-03-09 | |
NIST | 40 340 | 140 | 2012-03-28 | |
LNE-LNHB | 40 184 | 92 | 2012-11-16 | |
BEV | 40 710 | 170 | 2015-06-02 | |
POLATOM | 40 270 | 160 | 2015-10-22 | |
NMISA | 40 550 | 160 | 2017-09-13 |
APMP.RI(II)-K2.I-131
Labi | xi | ui | Date of measurement |
kBq | kBq | ||
ANSTO | 40 450 | 220 | March 2009 |
BARC | 40 070 | 190 | March 2009 |
INER | 41 190 | 170 | March 2009 |
KRISS | 41 300 | 100 | March 2009 |
OAP | 40 920 | 640 | March 2009 |
PTKMR | 39 900 | 900 | March 2009 |
CCRI(II)-S6.I-131
Labi | xi | ui | Date of measurement |
kBq | kBq | ||
LNMRI/IRD | 40 160 | 540 | July 2006 |
|
EURAMET.RI(II)-K2.I-131
Labi | xi | ui | Date of measurement |
kBq | kBq | ||
ENEA-INMRI | 40 290 | 170 | December 2015 |