BIPM.RI(II)-K1.Ge-68 and CCRI(II)-K2.Ge-68
MEASURAND : Equivalent activity of Ge-68
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
Results in absolute values
Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Results in relative values
BIPM.RI(II)-K1.Ge-68 and CCRI(II)-K2.Ge-68
MEASURAND : Equivalent activity of Ge-68
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
MBq |
MBq |
MBq |
||
IRA | 0.02 | 0.16 | ||
LNE-LNHB | 0.08 | 0.15 | ||
TENMAK-NUKEN | 0.19 | 0.48 | ||
NIM | -0.11 | 0.12 | ||
ANSTO | -0.05 | 0.18 | ||
BARC | -0.12 | 0.10 | ||
CIEMAT | -0.09 | 0.13 | ||
IFIN-HH | -0.22 | 0.31 | ||
INER | -0.10 | 0.12 | ||
KRISS | 0.19 | 0.21 | ||
LNMRI-IRD | 0.00 | 0.18 | ||
NMIJ | 0.05 | 0.23 | ||
NPL | 0.09 | 0.23 | ||
POLATOM | 0.09 | 0.21 | ||
PTB | 0.11 | 0.27 | ||
SMU | 1.71 | 0.15 |
Results in absolute values
Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq/MBq |
kBq/MBq |
kBq/MBq |
||
NIST | 4 | 12 | ||
IRA | 2 | 10 | ||
LNE-LNHB | 5 | 9 | ||
TENMAK-NUKEN | 12 | 31 | ||
NIM | -7.0 | 7.7 | ||
ANSTO | -3 | 12 | ||
BARC | -7.5 | 6.6 | ||
CIEMAT | -5.8 | 8.1 | ||
IFIN-HH | -14 | 19 | ||
INER | -6.1 | 7.3 | ||
KRISS | 12 | 13 | ||
LNMRI-IRD | 0 | 11 | ||
NMIJ | 3 | 14 | ||
NPL | 5 | 14 | ||
POLATOM | 5 | 13 | ||
PTB | 7 | 17 | ||
SMU | 109 | 10 |
Results in relative values
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Activity of radionuclide Ge-68 |
Time of measurements | 2012 - |
Status | Continuous, approved for equivalence |
Final Reports of the comparisons | |
Measurand | SIR equivalent activity |
Transfer device | Radioactive solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | BIPM (Bureau International des Poids et Mesures) |
Comments | Most recent results published on 16 February 2024. The status for not yet completed comparisons is indicated in "Participants". |
Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
Contact person | Carine Michotte +33 (0) 1 45 07 70 85 |
Pilot laboratory | |
---|---|
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
IRA |
Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET |
2015 |
LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
2015 |
LNMRI/IRD |
Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM |
2013 |
NIM |
National Institute of Metrology, China, APMP |
2015 | |
2021 |
|
NIST |
National Institute of Standards and Technology, United States, SIM |
2014 |
TAEK |
Turkish Atomic Energy Authority, Türkiye, EURAMET |
2018 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(II)-K1.Ge-68 and CCRI(II)-K2.Ge-68
MEASURAND : Equivalent activity of Ge-68
BIPM.RI(II)-K1.Ge-68
Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is xR = 15773 kBq with a standard uncertainty uR = 30 kBq
(see Section 4.1 of the BIPM.RI(II)-K1.I-Ge-58 Final Report). The value xi is taken as the equivalent activity for a laboratory i.
The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui its expanded uncertainty (k = 2), both expressed in MBq, and
Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.
The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Linking CCRI(II)-K2.Ge-68 to BIPM.RI(II)-K1.Ge-68
The value xi is the equivalent activity for laboratory i participant in CCRI(II)-K2.Ge-68 having been normalized using the value of the NMIJ as the linking laboratory.
The degree of equivalence of laboratory i participant in CCRI(II)-K2.Ge-68 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.
These statements make it possible to extend the BIPM.RI(II)-K1.Ge-68 matrix of equivalence to all participants in the CRI(II)-K2.Ge-68 comparison.
BIPM.RI(II)-K1.Ge-68 and CCRI(II)-K2.Ge-68
MEASURAND : Equivalent activity of Ge-68
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
BIPM.RI(II)-K1.Ge-68
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
NIST | 15 829 | 98 | 2014-12-15 | |
IRA | 15 797 | 86 | 2015-04-22 | |
LNE-LNHB | 15 855 | 81 | 2015-05-28 | |
TENMAK-NÜKEN | 15 960 | 240 | 2018-01-04 | |
NIM | 15 663 | 37 | 2021-05-10 |
CCRI(II)-K2.Ge-68
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
ANSTO | 15 725 | 86 | 2018 | |
BARC | 15 655 | 42 | 2018 | |
CIEMAT | 15 682 | 56 | 2018 | |
IFIN-HH | 15 550 | 150 | 2018 | |
INER | 15 677 | 49 | 2018 | |
KRISS | 15 960 | 100 | 2018 | |
LNMRI-IRD | 15 777 | 84 | 2018 | |
NMIJ | 15 820 | 110 | 2018 | |
NPL | 15 860 | 110 | 2018 | |
POLATOM | 15 860 | 100 | 2018 | |
PTB | 15 880 | 130 | 2018 | |
SMU | 17 487 | 71 | 2018 |