Comparison
Comparison close
Results

BIPM.RI(II)-K1.Co-60

MEASURAND : Equivalent activity of Co-60

Degrees of equivalence relative to the BIPM.RI(II)-K1.Co-60 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq.

Co-60
Di / kBq
Participating institutes

Results in absolute values

BIPM.RI(II)-K1.Co-60

MEASURAND : Equivalent activity of Co-60

Degrees of equivalence relative to the BIPM.RI(II)-K1.Co-60 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Co-60
Di / kBq/MBq
Participating institutes

Results in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Co-60

MEASURAND : Equivalent activity of Co-60

Degrees of equivalence relative to the BIPM.RI(II)-K1.Co-60 key comparison reference value

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq.

LABi Di Uneg,i Upos,i
kBq
kBq
kBq
NMIJ -12 16
JRC -23 34
IFIN-HH 39 48
BEV -5 34
CNEA 8 52
NRC 3 18
NIM -10 38
TENMAK-NUKEN -14 178
VNIIM 0 14
PTB 7 36
NIST 0 36
SMU -15 54
BARC -13 42
POLATOM 14 52
LNE-LNHB 8 24
ENEA-INMRI 34 60
NPL -4 20
LNMRI-IRD -4 46
NMISA 6 42
ANSTO 0 18
Participating institutes
BIPM.RI(II)-K1.Co-60

Results in absolute values

Degrees of equivalence represented by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
NMIJ -1.7 2.2
JRC -3.3 4.8
IFIN-HH 5.5 6.8
BEV -0.7 4.9
CNEA 1.1 7.3
NRC 0.4 2.5
NIM -1.4 5.3
TENMAK-NUKEN -2 25
VNIIM -0.0 1.9
PTB 1.0 5.1
NIST -0.0 5.1
SMU -2.1 7.6
BARC -1.8 5.9
POLATOM 2.0 7.3
LNE-LNHB 1.1 3.3
ENEA-INMRI 4.8 8.5
NPL -0.6 2.8
LNMRI-IRD -0.6 6.5
NMISA 0.8 5.9
ANSTO -0.0 2.5
Participating institutes
BIPM.RI(II)-K1.Co-60

Results in relative values

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Co-60
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Co-60
Time of measurements 1975 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
Most recent results published on 29 September 2021, revised on 20 March 2023.
The status for not yet completed comparisons is indicated in "Participants".
 
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP


BARC

Bhabha Atomic Research Centre, India, APMP


BEV

Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET

BFKH

Government Office of the Capital City Budapest, Hungary, EURAMET

comparison carried out with OMH (Országos Mérésügyi Hivatal)

CIEMAT

Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET

CMI

Czech Metrology Institute, Czechia, EURAMET

comparison carried out with CMI-IIR (Czech Metrology Institute - Inspectorate for Ionizing Radiation)

CNEA

National Commission of Atomic Energy, Argentina, SIM

ENEA-INMRI

Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Italy, EURAMET


IFIN-HH

"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET

IRA

Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET

JRC

Joint Research Centre, European Union, EURAMET

comparison carried out with IRMM (Institute for Reference Materials and Measurements)

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET


LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM


NIM

National Institute of Metrology, China, APMP


NIST

National Institute of Standards and Technology, United States, SIM


NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS


comparison carried out with CSIR-NML (Council for Scientific and Industrial Research, National Metrology Laboratory)

NPL

National Physical Laboratory, United Kingdom, EURAMET


NRC

National Research Council, Canada, SIM

Report in progress, Draft A

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET


PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET


PTKMR-BATAN

Center for Technology of Radiation Safety and Metrology, National Nuclear Energy Agency, Indonesia, APMP

comparison carried out with P3KRBiN (Pusat Penelitian & Pengembangan Keselamatan Radiasi & Biomedika Nuklir)

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET


TENMAK-NUKEN

Nuclear Energy Research Institute, Türkiye, EURAMET

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET


Comparison
Comparison close
Results

BIPM.RI(II)-K1.Co-60

MEASURAND : Equivalent activity of Co-60

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 7 062 kBq with a standard uncertainty uR = 2.3 kBq

(see Section 4.1 of the BIPM.RI(II)-K1.Co-60 Final Report).  The value xi is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in kBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Co-60

MEASURAND : Equivalent activity of Co-60

xi        result of measurement carried out in the SIR for the sample submitted by laboratory iLabi

ui    combined standard uncertainty of xi

   
  Labi xi ui Date of measurement
    kBq kBq  
  NMIJ 7050 8 2004-03-17
  JRC 7039 17 2005-01-27
  IFIN-HH 7101 24 2007-05-10
  BEV 7057 17 2007-09-27
  CNEA 7070 26 2011-10-24
  NRC 7065 9 2012-08-29
  NIM 7052 19 2014-07-01
  TENMAK-NUKEN 7048 89 2018-01-08
  VNIIM 7062 7 2019-06-28
  PTB 7069 18 2020-07-30
  NIST 7062 18 2020-11-09
  SMU 7047 27 2020-11-09
  BARC 7049 21 2021-01-07
  POLATOM 7076 26 2021-03-03
  LNE-LNHB 7070 12 2021-03-04
  ENEA-INMRI 7096 30 2021-04-15
  NPL 7058 10 2021-05-03
  LNMRI-IRD 7058 23 2021-09-01
  NMISA 7068 21 2022-04-08
  ANSTO 7062 9 2022-08-18