Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ga-67

 

MEASURAND: Equivalent activity of  67Ga

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Ga-67
Di / MBq
Participating institutes

Results given in absolute values.

BIPM.RI(II)-K1.Ga-67

 

MEASURAND: Equivalent activity of  67Ga

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Ga-67
Di / (kBq/MBq)
Participating institutes

Results given in relative values.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ga-67

 

MEASURAND: Equivalent activity of  67Ga

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
LNE-LNHB -0.96 0.91
PTB 0.7 1.2
NIST 0.3 1.1
CMI 0.3 2.0
Participating institutes
BIPM.RI(II)-K1.Ga-67

Results given in absolute values.

BIPM.RI(II)-K1.Ga-67

 

MEASURAND: Equivalent activity of  67Ga

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
LNE-LNHB -8.3 8.0
PTB 6 10
NIST 3 10
CMI 3 18
Participating institutes
BIPM.RI(II)-K1.Ga-67

Results given in relative values.

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Ga-67
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Ga-67
Time of measurements 1978 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
First results published 28 April 2003.
Most recent results published on 11 September 2024.
The status for not yet completed comparisons is indicated in "Participants".
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

BFKH

Government Office of the Capital City Budapest, Hungary, EURAMET

CIEMAT

Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET

CMI

Czech Metrology Institute, Czechia, EURAMET

DTU

Technical University of Denmark, Denmark, EURAMET

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NPL

National Physical Laboratory, United Kingdom, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ga-67

 

MEASURAND: Equivalent activity of  67Ga

 

Key comparison reference value: the SIR reference value xR for this radionuclide is 114 780 kBq, with a standard uncertainty uequal to 420 kBq 

(see section 4.1 of the BIPM.RI(II)-K1.Ga-67 Final report). The value x is taken as the equivalent activity for a laboratory i.

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in MBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ga-67

 

MEASURAND: Equivalent activity of  67Ga

 

xi   result of measurement carried out in the SIR for the sample submitted by laboratory iLabi 

ui   combined standard uncertainty of xi

  Labi xi ui Date of measurement
  kBq kBq
  LNE-LNHB 113 820 320 2005-10-20
  PTB 115 510 600 2010-03-11
  NIST 115 110 530 2010-05-04
  CMI 115 100 1 100 2023-09-18