CCQM-K90

MEASURAND : Formaldehyde (HCHO) in nitrogen

Degrees of equivalence Di and expanded uncertainty Ui (k = 2)

 

 

 

CCQM-K90

MEASURAND : Formaldehyde (HCHO) in nitrogen

Degrees of equivalence Di and expanded uncertainty Ui (k = 2)

 

 

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
CC comparison
Linked comparison
APMP.QM-K90
Metrology area, Sub-field Chemistry and Biology, Gases
Description Formaldehyde in Nitrogen
Time of measurements 2018 - 2019
Status Measurements in progress
References

Technical Protocol

Measurand Amount-of-substance fraction of Formaldehyde in Nitrogen
2 μmolmol
Transfer device Gas cylinders
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by APMP (Asia Pacific Metrology Programme)
RMO Internal Identifier APMP.QM-K90
Comments APMP.QM-K90 is a key comparison (model 1) of formaldehyde in nitrogen at 2 µmol/mol linked to CCQM-K90
Pilot institute NMIJ AIST
National Metrology Institute of Japan
Japan
Contact person Takuya Shimosaka

+81-29-861-6851
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

CMS

ITRI Center for Measurement Standards, Chinese Taipei, APMP

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NPL

National Physical Laboratory, United Kingdom, EURAMET

CCQM-K90

MEASURAND : Formaldehyde (HCHO) in nitrogen

The key comparison reference value for each cylinder, xRi, is equal to the determination made by the BIPM, using their largest uncertainty.

 

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = xi - xRi and its expanded uncertainty (k = 2), Ui both expressed in µmol/mol. Ui2 = 4 [u(xi)2 + u(xRi)2].

CCQM-K90

MEASURAND : Formaldehyde (HCHO) in nitrogen

NOMINAL VALUES : 2 µmol/mol

t          time in days
a1       loss rate
x
0       HCHO mole fraction extrapolated to 14 August 2014 (t = 0)
xi        reported values at time t
u(xi)   standard uncertainty of xi
xRi      reference value at time t
uRi         standard uncertainty of xRi