Comparison
Comparison close
Results
Comparison
Comparison close
Results
Comparison
Comparison close
RMO comparison
Linked comparison
EURAMET.AUV.V-S1
Metrology area, Sub-field Acoustics, Ultrasound, Vibration, Vibration
Description Secondary calibration of accelerometers following ISO 16063-21:2003
Time of measurements 2019 - 2021
Status Approved
Final Reports of the comparisons
References

EURAMET.AUV.V-S1 Technical Protocol

Measurand Accelerometer sensitivity magnitude in pC/(m/s2)
Phase in 1°

Nominal magnitude at 160 Hz: 0.13 pC/(m/s2) to 1 pC/(m/s2)
Nominal phase at 160 Hz: -0.7° to 0°
Parameters Frequency range: from 10 Hz to 10 kHz
Acceleration amplitude: admissible from 2 m/s2 to 200 m/s2
Transfer device Brϋel & Kjær 8305-001 (SN: 2423435) "single ended" (SE) type
Brϋel & Kjær 4371V (SN: 2046745) "single ended" (SE) type
Endevco 2270 (SN: 16198) "single ended" (SE) type
Comparison type Supplementary Comparison
Consultative Committee CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET Project 1481
Pilot institute BEV
Bundesamt für Eich- und Vermessungswesen
Austria
Contact person Peter Rosenkranz

+43121110826515
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BEV

Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET

BIM

Bulgarian Institute of Metrology, Bulgaria, EURAMET

CMI

Czech Metrology Institute, Czechia, EURAMET

IPQ

Instituto Portugues da Qualidade, Portugal, EURAMET

KEBS

Kenya Bureau of Standards, Kenya, AFRIMETS

METAS

Federal Institute of Metrology, Switzerland, EURAMET

MIKES

VTT Technical Research Centre of Finland Ltd, Centre for Metrology/Mittatekniikan keskus, Finland, EURAMET

Comparison
Comparison close
Results
Comparison
Comparison close
Results