Comparison
Comparison close
Results

BIPM.RI(II)-K1.Zn-65

 

MEASURAND           Equivalent activity of Zn-65

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Zn-65
Di / MBq
Participating institutes

Results given in absolute values

MEASURAND           Equivalent activity of Zn-65

 

Degrees of equivalence represented by Di = (xI - xR) / xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in KBq/MBq.

Zn-65
Di / xR / (kBq/MBq)
Participating institutes

Results given in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Zn-65

 

MEASURAND           Equivalent activity of Zn-65

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
BARC -0.60 0.61
LNMRI-IRD 0.16 0.40
LNE-LNHB -0.01 0.18
Participating institutes
BIPM.RI(II)-K1.Zn-65

Results given in absolute values

MEASURAND           Equivalent activity of Zn-65

 

Degrees of equivalence represented by Di = (xI - xR) / xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in KBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
BARC -20 21
LNMRI-IRD 5 13
LNE-LNHB -0.5 6.0
Participating institutes
BIPM.RI(II)-K1.Zn-65

Results given in relative values

Comparison
Comparison close
BIPM comparison
Linked comparison
SIM.RI(II)-K2.Zn-65
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of Zn-65 solution
Time of measurements 2021 - 2022
Status Report in progress, draft A
References

Protocol_SIM.RI(II)K2-Zn-65_ Aug2022

Measurand Activity
0.5 MBq
Transfer device Sample of Zn-65
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by SIM (Inter-American Metrology System)
Pilot institute LNMRI
Laboratorio Nacional de Metrologia das Radiaçoes Ionizantes, IRD
Brazil
Contact person KARLA PATRÃO

+5521993858835
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
LNMRI

Laboratorio Nacional de Metrologia das Radiaçoes Ionizantes, IRD, Brazil, SIM

BARC

Bhabha Atomic Research Centre, India, APMP

BFKH

Government Office of the Capital City Budapest, Hungary, EURAMET

CIEMAT

Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET

CNEA

National Commission of Atomic Energy, Argentina, SIM

FTMC

Center for Physical Sciences and Technology, Lithuania, EURAMET

ININ

Instituto Nacional de Investigaciones Nucleares, Mexico, SIM

NIST

National Institute of Standards and Technology, United States, SIM

NRC

National Research Council, Canada, SIM

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

VINS

Vinca Institute of Nuclear Sciences, Serbia, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Zn-65

 

MEASURAND           Equivalent activity of Zn-65

 

Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is  xR = 29 730 kBq with a standard uncertainty uR = 39 kBq

(see Section 4.1 of the Final report : Metrologia 2024 61 Tech. Supp. 06004).  The value xi is taken as the equivalent activity for a laboratory i.

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in MBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Zn-65

 

MEASURAND           Equivalent activity of Zn-65

 

xi   result of measurement carried out in the SIR for the sample submitted by laboratory i, Labi

ui   combined standard uncertainty of xi

 

  Labi xi ui Date of measurement
    kBq kBq  
  BARC 29130 310 2006-11-29
  LNMRI-IRD 29890 200 2015-12-15
  LNE-LNHB 29715 88 2018-12-14