BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51
MEASURAND : Equivalent activity of Cr-51
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
Results in absolute values
Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Results in relative values
BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51
MEASURAND : Equivalent activity of Cr-51
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
MBq |
MBq |
MBq |
||
NMIJ | -0.9 | 3.1 | ||
NPL | -1.0 | 6.7 | ||
KRISS | 0.4 | 4.7 | ||
POLATOM | -4.4 | 5.9 | ||
VNIIM | 0.8 | 2.5 | ||
LNE-LNHB | -1.6 | 3.2 |
Results in absolute values
Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq/MBq |
kBq/MBq |
kBq/MBq |
||
NMIJ | -1.9 | 6.3 | ||
NPL | -2 | 14 | ||
KRISS | 0.7 | 9.7 | ||
POLATOM | -9 | 12 | ||
VNIIM | 1.6 | 5.1 | ||
LNE-LNHB | -3.2 | 6.6 |
Results in relative values
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | ACTIVITY OF RADIONUCLIDE Cr-51 |
Time of measurements | 2022 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References |
International comparison of activity measurements of a solution of 51Cr |
Measurand | Activity per mass kBq/g |
Transfer device | Radioactive 51Cr solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | EURAMET (European Association of National Metrology Institutes) |
RMO Internal Identifier | EURAMET Project 1554 |
Comments | The comparison is organized in a frame of EURAMET Mentoring Scheme Award MSA-002-2022_POLATOM-LNHB |
Pilot institute |
POLATOM
Radioisotope Centre Polatom Poland |
Contact person | Ryszard Broda +48 22 273 19 53 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51
MEASURAND : Equivalent activity of Cr-51
BIPM.RI(II)-K1.Cr-51
Key comparison reference value: the SIR key comparison reference value xR for this radionuclide is xR = 488.05 MBq with a standard uncertainty uR = 0.41 MBq
(see Section 4.1 of the BIPM.RI(II)-K1.I-Cr-51 Final Report). The value xi is taken as the equivalent activity for a laboratory i.
The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui its expanded uncertainty (k = 2), both expressed in MBq, and
Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.
The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Linking APMP.RI(II)-K2.Cr-51 to BIPM.RI(II)-K1.Cr-51
The value xi is the equivalent activity for laboratory i participant in APMP.RI(II)-K2.Cr-51 having been normalized using the value of the NMIJ as the linking laboratory.
The degree of equivalence of laboratory i participant in APMP.RI(II)-K2.Cr-51 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value.
Linking EURAMET.RI(II)-K2.Cr-51 to BIPM.RI(II)-K1.Cr-51
The value xi is the equivalent activity for laboratory i participant in EURAMET.RI(II)-K2.Cr-51 having been normalized to the value of the POLATOM as the linking laboratory.
The degree of equivalence of laboratory i participant in EURAMET.RI(II)-K2.Cr-511 with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used in the following table as none of these Di contributed to the key comparison reference value..
These statements make it possible to extend the BIPM.RI(II)-K1.Cr-51 matrices of equivalence to all participants in the APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51 comparisons
BIPM.RI(II)-K1.Cr-51, APMP.RI(II)-K2.Cr-51 and EURAMET.RI(II)-K2.Cr-51
MEASURAND : Equivalent activity of Cr-51
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
BIPM.RI(II)-K1.Cr-51
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
NMIJ | 487.1 | 1.6 | 2004-03-15 | |
NPL | 487.0 | 3.4 | 2004-05-28 | |
KRISS | 488.4 | 2.4 | 2012-01-27 | |
POLATOM | 483.6 | 3.0 | 2022-06-03 |
APMP.RI(II)-K2.Cr-51
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
VNIIM | 488.8 | 1.2 | 2004 |
EURAMET.RI(II)-K2.Cr-51
Labi | xi | ui | Date of measurement | |
kBq | kBq | |||
LNE-LNHB | 486.5 | 1.6 | 2022 |