Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ho-166m and EURAMET.RI(II)-K2.Ho-166m

 

MEASURAND   Equivalent activity of 166mHo

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

Equivalent activity of 166mHo
Di / MBq
Participating institutes

Results in absolute values

BIPM.RI(II)-K1.Ho-166m and EURAMET.RI(II)-K2.Ho-166m

 

MEASURAND   Equivalent activity of 166mHo

 

Degrees of equivalence represented by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Equivalent activity of 166mHo
Di / kBq/MBq
Participating institutes

Results in relative values

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ho-166m and EURAMET.RI(II)-K2.Ho-166m

 

MEASURAND   Equivalent activity of 166mHo

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.

LABi Di Uneg,i Upos,i
MBq
MBq
MBq
IRA -0.05 0.06
NPL -0.03 0.07
PTB -0.05 0.06
NMISA 0.01 0.08
CIEMAT -0.06 0.07
LNE-LNHB -0.07 0.08
CMI-IIR -0.02 0.14
IRMM -0.05 0.06
Participating institutes
BIPM.RI(II)-K1-Ho-166m
EURAMET.RI(II)-K2.Ho-166m

Results in absolute values

BIPM.RI(II)-K1.Ho-166m and EURAMET.RI(II)-K2.Ho-166m

 

MEASURAND   Equivalent activity of 166mHo

 

Degrees of equivalence represented by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
NMIJ 9.04 7.03
IRA -5.02 6.02
NPL -3.01 7.03
PTB -5.02 6.02
NMISA 1.00 8.03
CIEMAT -6.02 7.03
LNE-LNHB -7.03 8.03
CMI-IIR -2.01 14.1
IRMM -5.02 6.02
Participating institutes
BIPM.RI(II)-K1-Ho-166m
EURAMET.RI(II)-K2.Ho-166m

Results in relative values

CCRI.RI(II)-K2.Ho-166m.2024
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Ho-166m
Time of measurements 2024
Status Measurements in progress
Measurand Activity of Ho-166m per unit mass
Transfer device Ampoules of Holmium chloride
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by CCRI (Consultative Committee for Ionizing Radiation)
Comments Ho-166m is long-lived isotope useful for various instrument calibration. The solution provided by IRA is free of gamma emitting impurities, and can be used for both activity determination and half-life measurement. Participants will be provided with ampoules of Holmium chloride of about 1 MBq.
Pilot institute IRA
Institut Universitaire de Radiophysique Appliquée
Switzerland
Contact person Frederic Juget

+41 21 31 48 162
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
IRA

Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP

BARC

Bhabha Atomic Research Centre, India, APMP

BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

CIEMAT

Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET

IFIN-HH

"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET

ININ

Instituto Nacional de Investigaciones Nucleares, Mexico, SIM

JRC

Joint Research Centre, European Union, EURAMET

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NPL

National Physical Laboratory, United Kingdom, EURAMET

NRC

National Research Council, Canada, SIM

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

TENMAK-NUKEN

Nuclear Energy Research Institute, Türkiye, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ho-166m and EURAMET.RI(II)-K2.Ho-166m

 

MEASURAND   Equivalent activity of 166mHo

 

  • Key comparison BIPM.RI(II)-K1.Ho-166m

Key comparison reference value xR is the SIR reference value for this radionuclide where xR = 9.959 MBq, with a standard uncertainty uR = 0.022 MBq (see section 4.1 of the Final Report). xi is the equivalent activity for laboratory i.

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. With n the number of laboratories, Ui = 2[(1 - 2/n)ui2 + (1/n2ui2]1/2 when each laboratory has contributed to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xi - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

       

  • Linking EURAMET.RI(II)-K2.Ho-166m to BIPM.RI(II)-K1.Ho-166m

The value xi is the equivalent activity for laboratory i participant in EURAMET.RI(II)-K2.Ho-166m, having been normalized to the value of PTB as the linking laboratory.

The degree of equivalence of laboratory i participant in EURAMET.RI(II)-K2.Ho-166m with respect to the key comparison reference value is given by a pair of terms:

Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Ui = 2(ui2 + uR2)1/2 is used as none of these laboratories contributed to the computation of the key comparison reference value xR.

These statements make it possible to extend the BIPM.RI(II)-K1.Ho-166m table of equivalence to participants in EURAMET.RI(II)-K2.Ho-166m.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Ho-166m and EURAMET.RI(II)-K2.Ho-166m

 

MEASURAND   Equivalent activity of 166mHo

 

  • Key comparison BIPM.RI(II)-K1.Ho-166m

xi : result of measurement carried out in the SIR for the sample submitted by laboratory i
ui : combined standard uncertainty of xi

  Labi xi ui Date of measurement
    kBq kBq  
  IRA 9 913 26 2006-12-22
  NPL 9 932 30 2009-03-02
  PTB 9 912 28 2013-03-01
  NMISA 9 973 40 2017-09-15

 

 

  • Key comparison EURAMET.RI(II)-K2.Ho-166m

xi : result of measurement carried out by laboratory i converted to the equivalent activity through the PTB
ui : combined standard uncertainty of xi

  Labi xi ui Date of measurement
    kBq kBq  
  CIEMAT 9 901 29 2013-06-25
  LNE-LNHB 9 890 32 2013-06-26
  CMI-IIR 9 944 66 2013-07-11
  IRMM 9 910 20 2013-08-23