CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Zinc freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

Zinc freezing point
Di / mK
Participating institutes

Enlarged graph

Graph_Zn_FP

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Tin freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

Tin freezing point
Di / mK
Participating institutes

Enlarged graph

Graph_Sn_FP

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Indium freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

Indium freezing point
Di / mK
Participating institutes

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Gallium melting point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

Gallium melting point
Di / mK
Participating institutes

Enlarged graph

Graph_Ga_MP

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Mercury triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

Mercury triple point
Di / mK
Participating institutes

Enlarged graph

Graph_Hg_Tp

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Argon triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

Argon triple point
Di / mK
Participating institutes

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Zinc freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

LABi Di Uneg,i Upos,i
mK
mK
mK
INMETRO -0.69 1.37
INRIM -0.61 0.91
INTI -1.56 3.37
KRISS 0.61 0.99
LNE-LCM/Cnam 1.01 0.96
NIM 0.13 0.74
NIST 0.02 0.61
NMIA -1.07 0.64
NMIJ 0.19 0.78
NPL -1.78 0.64
NRC 0.30 0.75
PTB 1.69 1.23
VSL 2.53 1.04
TTBS -1.9 11.9
BEV -0.80 1.62
BFKH 3.74 1.08
BIM -0.59 3.27
BOM 3.86 1.71
CEM 1.43 0.98
CMI 0.85 1.41
DMDM 0.11 1.71
DTI 0.73 1.37
EIM -1.47 1.69
FTMC 1.01 1.80
GUM 3.69 1.35
FSB-LPM -2.39 4.35
IMBIH 1.97 2.77
INM -0.85 2.44
IPQ 2.48 2.22
JV 1.19 1.19
MIKES 1.45 2.11
MIRS/UL-FE/LMK -0.35 0.89
NSAI NML -0.45 1.92
RISE 2.52 1.33
METAS 0.41 1.59
SMD 4.55 2.82
SMU 3.72 1.39
UME 1.42 1.69
AS METROSERT 1.92 6.60
Participating institutes
CCT-K9
SIM.T-K9.3
EURAMET.T-K9
EURAMET.T-K9.1

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Tin freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

LABi Di Uneg,i Upos,i
mK
mK
mK
INMETRO -0.91 0.84
INRIM 0.09 0.59
INTI -1.10 1.88
KRISS 0.22 0.75
LNE-LCM/Cnam 0.29 0.64
NIM 0.33 0.54
NIST -0.11 0.37
NMIA -0.21 0.46
NMIJ -0.03 0.53
NPL 0.09 0.50
NRC 0.65 0.47
PTB 0.28 0.79
VSL 0.54 0.56
TTBS -3.9 10.8
BEV -0.05 1.81
BFKH 1.33 1.35
BIM 0.87 1.14
BOM 0.43 0.91
CEM 1.10 0.96
CMI 0.02 1.26
DMDM 0.58 1.24
DTI -0.42 0.46
EIM -0.05 1.22
FTMC -0.13 1.21
GUM 0.29 0.51
FSB-LPM -0.22 1.84
IMBIH -0.78 1.52
INM 0.01 2.10
IPQ 1.17 1.40
JV 0.34 0.64
MIKES 0.29 1.74
MIRS/UL-FE/LMK -0.31 0.64
NSAI NML -0.53 1.55
RISE -0.02 0.72
METAS -0.16 1.22
SMD 0.45 1.27
SMU -0.27 0.98
UME 0.20 1.01
AS METROSERT 0.10 4.90
Participating institutes
CCT-K9
SIM.T-K9.3
EURAMET.T-K9
EURAMET.T-K9.1

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Indium freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

LABi Di Uneg,i Upos,i
mK
mK
mK
INMETRO -0.65 0.61
INRIM -0.02 0.58
INTI 0.45 2.35
KRISS 0.27 0.97
LNE-LCM/Cnam 0.64 0.49
NIM 0.13 0.43
NIST 0.13 0.25
NMIA -0.04 0.39
NMIJ 0.16 0.36
NPL -0.29 0.29
NRC 0.12 0.29
PTB 0.06 0.75
VSL -0.02 0.40
BEV 0.21 1.58
BFKH 0.69 1.17
BIM 0.56 0.86
CEM 0.48 0.75
CMI 0.02 0.89
DMDM 0.46 1.76
DTI -0.13 1.06
EIM 0.59 1.63
GUM 0.54 0.84
INM 0.13 1.79
IPQ 1.30 1.80
JV -0.07 0.48
MIRS/UL-FE/LMK 0.08 0.58
NSAI NML -1.13 1.98
RISE -0.19 0.56
METAS 0.44 1.72
SMD 0.11 0.96
SMU 0.64 0.92
UME -0.19 1.12
Participating institutes
CCT-K9
EURAMET.T-K9

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Gallium melting point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

LABi Di Uneg,i Upos,i
mK
mK
mK
BIPM 0.11 0.32
INMETRO 0.29 0.34
INRIM 0.04 0.35
INTI 0.15 0.87
KRISS -0.08 0.41
LNE-LCM/Cnam -0.36 0.38
NIM 0.26 0.38
NIST -0.22 0.23
NMIA -0.33 0.35
NMIJ -0.24 0.30
NPL -0.06 0.26
NRC 0.12 0.27
PTB -0.05 0.30
VSL -0.14 0.36
TTBS -0.7 6.4
BEV 0.14 0.70
BFKH 0.20 0.42
BIM 1.66 0.44
BOM 0.42 0.66
CEM 0.07 0.48
CMI -0.22 0.40
DMDM 0.32 0.61
DTI -0.19 0.41
EIM -0.10 0.49
FTMC -1.08 0.62
GUM -0.16 0.48
FSB-LPM -0.44 0.84
IMBiH -0.25 0.64
INM 0.23 0.49
IPQ -0.39 0.68
JV -0.30 0.56
MIKES -0.53 0.76
MIRS/UL-FE/LMK 0.18 0.34
NSAI NML -0.12 0.62
RISE -0.32 0.46
METAS 0.23 0.69
SMD -0.28 0.49
SMU -0.27 0.36
UME -0.24 0.40
AS METROSERT -0.67 1.85
Participating institutes
CCT-K9
SIM.T-K9.3
EURAMET.T-K9
EURAMET.T-K9.1

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Mercury triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

LABi Di Uneg,i Upos,i
mK
mK
mK
INMETRO -0.11 0.41
INRIM 0.18 0.30
INTI -0.08 0.66
KRISS -0.19 0.40
LNE-LCM/Cnam 0.76 0.64
NIM 0.24 0.31
NIST -0.32 0.25
NMIA -0.04 0.29
NMIJ 0.18 0.59
NPL 0.06 0.24
NRC 0.01 0.24
PTB 0.12 0.29
VSL -0.07 0.25
TTBS 1.0 5.9
BEV 0.00 0.87
BFKH 1.30 0.33
BIM -0.02 0.55
BOM -0.20 0.67
CEM 0.65 0.78
CMI 0.31 0.44
DMDM -0.17 0.68
DTI -0.11 0.47
EIM 0.03 0.56
FTMC -0.09 0.67
GUM -0.27 0.47
FSB-LPM -0.05 0.27
IMBIH -0.28 0.55
INM -0.05 0.37
IPQ 0.48 1.00
JV 0.60 0.74
MIKES 0.55 0.80
MIRS/UL-FE/LMK 0.00 0.31
NSAI NML 0.07 0.60
RISE -0.20 0.45
METAS 0.07 0.82
SMD 0.63 1.03
SMU 1.16 0.59
UME 0.55 0.57
AS METROSERT 0.91 3.48
Participating institutes
CCT-K9
SIM.T-K9.3
EURAMET.T-K9
EURAMET.T-K9.1

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Argon triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

Degrees of equivalence represented by Di = (TI - TR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mK.

LABi Di Uneg,i Upos,i
mK
mK
mK
INMETRO 0.21 0.79
INRIM 0.22 1.08
INTI 0.95 1.74
KRISS 0.28 1.03
LNE-LCM/Cnam -0.17 0.80
NIM -0.67 0.66
NIST -1.41 0.54
NMIA -0.36 0.99
NMIJ 0.20 1.30
NPL 0.35 0.66
NRC -0.06 0.65
PTB 0.10 0.73
VSL -0.82 0.63
BEV -0.66 1.53
CEM -0.60 1.20
CMI 0.84 1.03
DTI -3.89 2.57
EIM 0.86 0.88
GUM -0.82 1.55
INM -0.98 1.06
IPQ -0.47 1.68
JV 0.45 1.09
MIKES 0.04 1.52
MIRS/UL-FE/LMK -0.17 1.16
RISE 0.06 1.57
METAS -0.60 1.32
SMD -1.49 1.16
UME 0.42 1.05
Participating institutes
CCT-K9
EURAMET.T-K9
CCT-K9.3
Metrology area, Sub-field Thermometry, Standard Platinum Resistance Thermometers
Description Realizations of the ITS-90 from 83.805 8 K to 1 234.93 K
Time of measurements 2023 - 2024
Status Measurements in progress
Measurand Temperature: from the triple point of Ar (83.805 8 K) to the freezing point of Ag (1 234.93 K)
Transfer device Standard Platinum Resistance Thermometers
Comparison type Key Comparison
Consultative Committee CCT (Consultative Committee for Thermometry)
Conducted by CCT (Consultative Committee for Thermometry)
Pilot institute CEM
Centro Español de Metrologia
Spain
Contact person Raúl Caballero

+34918074822
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
CEM

Centro Español de Metrologia, Spain, EURAMET

CESMEC

Centro de Estudios, Medición y Certificación de Calidad, Chile, SIM

GEOSTM

Georgian National Agency for Standards and Metrology, Georgia, EURAMET

LACOMET

Laboratorio Costarricense de Metrologia, Costa Rica, SIM

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Zinc freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

CCT-K9

Key comparison reference value:

Resistance ratios are used with calibration equations for each thermometer to obtain temperature Ti.
The key comparison reference temperature, TR, is calculated from the mean of the average temperatures from the participants' results, excluded SRPT measurements that fail the cutoff criteria described in section 3.5 of the
Final Report, and its associated  standard uncertainty uR is  computed as explain in section 2.4 of the Final Report.

 

        TR = -0.20 mK     uR = 0.11 mK

 

The degree of equivalence of laboratory i relative to the key comparison reference value TR is given by a pair of terms, both expressed in mK:

 

Di = (Ti - TR) and its expanded uncertainty Ui at a 95 % level of confidence is computed as described in section 2.3 of the Final Report.

 

Linking SIM.T-K9.3 to CCT-K9

The INMETRO provides the link between the key comparison CCT-K9 and SIM.T-K9.3, having participated in both comparisons. The linking process is described in Section 7 of the the SIM.T-K9.3 Final report.

 

Linking EURAMET.T-K9 to CCT-K9

LNE-LCM/Cnam, INRIM, NPL, PTB and VSL provide the link between the key comparison CCT-K9 and EURAMET.T-K9, having participated in both comparisons. The comparison was organised in five separate colapsed stars, each having one CCT-K9 participant as pilot, which provides the link to the KCRV (Section 1, 2.1 and 2.4 of EURAMET.T-K9 Final report).

 

Linking EURAMET.T-K9.1 to CCT-K9

JV provides the link between the key comparison CCT-K9 and EURAMET.T-K9.1, having participated in both comparisons. The linking process is described in Section 3 of the EURAMET.T-K9.1 Final report.

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Tin freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Key comparison reference value:

Resistance ratios are used with calibration equations for each thermometer to obtain temperature Ti.
The key comparison reference temperature, TR, is calculated from the mean of the average temperatures from the participants' results, excluded SRPT measurements that fail the cutoff criteria described in section 3.5 of the
Final Report, and its associated  standard uncertainty uR is  computed as explain in section 2.4 of the Final Report.

 

        TR = 0.06 mK     uR = 0.08 mK

 

The degree of equivalence of laboratory i relative to the key comparison reference value TR is given by a pair of terms, both expressed in mK:

 

Di = (Ti - TR) and its expanded uncertainty Ui at a 95 % level of confidence is computed as described in section 2.3 of the Final Report.

 

Linking SIM.T-K9.3 to CCT-K9

The INMETRO provides the link between the key comparison SIM.T-K9.3 to CCT-K9, having participated in both comparisons. The linking process is described in Section 7 of the the SIM.T-K9.3 Final report.

 

Linking EURAMET.T-K9 to CCT-K9

LNE-LCM/Cnam, INRIM, NPL, PTB and VSL provide the link between the key comparison CCT-K9 and EURAMET.T-K9, having participated in both comparisons. The comparison was organised in five separate colapsed stars, each having one CCT-K9 participant as pilot, which provides the link to the KCRV (Section 1, 2.1 and 2.4 of EURAMET.T-K9 Final report).

 

Linking EURAMET.T-K9.1 to CCT-K9

JV provides the link between the key comparison CCT-K9 and EURAMET.T-K9.1, having participated in both comparisons. The linking process is described in Section 3 of the EURAMET.T-K9.1 Final report.

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Indium freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Key comparison reference value:

Resistance ratios are used with calibration equations for each thermometer to obtain temperature Ti.
The key comparison reference temperature, TR, is calculated from the mean of the average temperatures from the participants' results, excluded SRPT measurements that fail the cutoff criteria described in section 3.5 of the
Final Report, and its associated  standard uncertainty uR is  computed as explain in section 2.4 of the Final Report.

 

        TR = -0.08 mK     uR = 0.05 mK

 

The degree of equivalence of laboratory i relative to the key comparison reference value TR is given by a pair of terms, both expressed in mK:

 

Di = (Ti - TR) and its expanded uncertainty Ui at a 95 % level of confidence is computed as described in section 2.3 of the Final Report.

 

Linking EURAMET.T-K9 to CCT-K9

LNE-LCM/Cnam, INRIM, NPL, PTB and VSL provide the link between the key comparison CCT-K9 and EURAMET.T-K9, having participated in both comparisons. The comparison was organised in five separate colapsed stars, each having one CCT-K9 participant as pilot, which provides the link to the KCRV (Section 1, 2.1 and 2.4 of EURAMET.T-K9 Final report).

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Gallium melting point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Key comparison reference value:

Resistance ratios are used with calibration equations for each thermometer to obtain temperature Ti.
The key comparison reference temperature, TR, is calculated from the mean of the average temperatures from the participants' results, excluded SRPT measurements that fail the cutoff criteria described in section 3.5 of the
Final Report, and its associated  standard uncertainty uR is  computed as explain in section 2.4 of the Final Report.

 

        TR = 0.23 mK     uR = 0.03 mK

 

The degree of equivalence of laboratory i relative to the key comparison reference value TR is given by a pair of terms, both expressed in mK:

 

Di = (Ti - TR) and its expanded uncertainty Ui at a 95 % level of confidence is computed as described in section 2.3 of the Final Report.

 

Linking SIM.T-K9.3 to CCT-K9

The INMETRO provides the link between the key comparison SIM.T-K9.3 to CCT-K9, having participated in both comparisons. The linking process is described in Section 7 of the the SIM.T-K9.3 Final report.

 

Linking EURAMET.T-K9 to CCT-K9

LNE-LCM/Cnam, INRIM, NPL, PTB and VSL provide the link between the key comparison CCT-K9 and EURAMET.T-K9, having participated in both comparisons. The comparison was organised in five separate colapsed stars, each having one CCT-K9 participant as pilot, which provides the link to the KCRV (Section 1, 2.1 and 2.4 of EURAMET.T-K9 Final report).

 

Linking EURAMET.T-K9.1 to CCT-K9

JV provides the link between the key comparison CCT-K9 and EURAMET.T-K9.1, having participated in both comparisons. The linking process is described in Section 3 of the EURAMET.T-K9.1 Final report.

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Mercury triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Key comparison reference value:

Resistance ratios are used with calibration equations for each thermometer to obtain temperature Ti.
The key comparison reference temperature, TR, is calculated from the mean of the average temperatures from the participants' results, excluded SRPT measurements that fail the cutoff criteria described in section 3.5 of the
Final Report, and its associated  standard uncertainty uR is  computed as explain in section 2.4 of the Final Report.

 

        TR = 0.28 mK     uR = 0.04 mK

 

The degree of equivalence of laboratory i relative to the key comparison reference value TR is given by a pair of terms, both expressed in mK:

 

Di = (Ti - TR) and its expanded uncertainty Ui at a 95 % level of confidence is computed as described in section 2.3 of the Final Report.

 

Linking SIM.T-K9.3 to CCT-K9

The INMETRO provides the link between the key comparison CCT-K9 and SIM.T-K9.3, having participated in both comparisons. The linking process is described in Section 7 of the the SIM.T-K9.3 Final report.

 

Linking EURAMET.T-K9 to CCT-K9

LNE-LCM/Cnam, INRIM, NPL, PTB and VSL provide the link between the key comparison CCT-K9 and EURAMET.T-K9, having participated in both comparisons. The comparison was organised in five separate colapsed stars, each having one CCT-K9 participant as pilot, which provides the link to the KCRV (Section 1, 2.1 and 2.4 of EURAMET.T-K9 Final report).

 

Linking EURAMET.T-K9.1 to CCT-K9

JV provides the link between the key comparison CCT-K9 and EURAMET.T-K9.1, having participated in both comparisons. The linking process is described in Section 3 of the EURAMET.T-K9.1 Final report.

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Argon triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Key comparison reference value:

Resistance ratios are used with calibration equations for each thermometer to obtain temperature Ti.
The key comparison reference temperature, TR, is calculated from the mean of the average temperatures from the participants' results, excluded SRPT measurements that fail the cutoff criteria described in section 3.5 of the
Final Report, and its associated  standard uncertainty uR is  computed as explain in section 2.4 of the Final Report.

 

        TR = 0.80 mK     uR = 0.08 mK

 

The degree of equivalence of laboratory i relative to the key comparison reference value TR is given by a pair of terms, both expressed in mK:

 

Di = (Ti - TR) and its expanded uncertainty Ui at a 95 % level of confidence is computed as described in section 2.3 of the Final Report.

 

Linking EURAMET.T-K9 to CCT-K9

LNE-LCM/Cnam, INRIM, NPL, PTB and VSL provide the link between the key comparison CCT-K9 and EURAMET.T-K9, having participated in both comparisons. The comparison was organised in five separate colapsed stars, each having one CCT-K9 participant as pilot, which provides the link to the KCRV (Section 1, 2.1 and 2.4 of EURAMET.T-K9 Final report).

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Zinc freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Wi   dimensionless resistance ratio measurement results of thermometer from laboratory i

ui   combined standard uncertainty of measurement made using thermometer from laboratory i are expressed in mK

 

The laboratory measurement results are listed in Appendix B of the CCT-K9 Final report.

 

SIM.T-K9.3

The measurement results are listed in Table 3 and 4 of the SIM.T-K9.3 Final report.

 

EURAMET.T-K9

The measurement results are listed in Section 4 of the EURAMET.T-K9 Final report.

 

EURAMET.T-K9.1

The measurement results are listed in Section 4 of the EURAME.T-K9.1 Final report.

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Tin freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Wi   dimensionless resistance ratio measurement results of thermometer from laboratory i

ui   combined standard uncertainty of measurement made using thermometer from laboratory i are expressed in mK

The laboratory measurement results are listed in Appendix B of the CCT-K9 Final report.

 

SIM.T-K9.3

The measurement results are listed in Table 3 and 4 of the SIM.T-K9.3 Final report.

 

EURAMET.T-K9

The measurement results are listed in Section 4 of the EURAMET.T-K9 Final report.

 

EURAMET.T-K9.1

The measurement results are listed in Section 4 of the EURAME.T-K9.1 Final report.

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Indium freezing point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Wi   dimensionless resistance ratio measurement results of thermometer from laboratory i

ui   combined standard uncertainty of measurement made using thermometer from laboratory i are expressed in mK

The laboratory measurement results are listed in Appendix B of the CCT-K9 Final report.

 

EURAMET.T-K9

The measurement results are listed in Section 4 of the EURAMET.T-K9 Final report.

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Gallium melting point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Wi   dimensionless resistance ratio measurement results of thermometer from laboratory i

ui   combined standard uncertainty of measurement made using thermometer from laboratory i are expressed in mK

The laboratory measurement results are listed in Appendix B of the CCT-K9 Final report.

 

SIM.T-K9.3

The measurement results are listed in Table 3 and 4 of the SIM.T-K9.3 Final report.

 

EURAMET.T-K9

The measurement results are listed in Section 4 of the EURAMET.T-K9 Final report.

 

EURAMET.T-K9.1

The measurement results are listed in Section 4 of the EURAME.T-K9.1 Final report.

CCT-K9, SIM.T-K9.3, EURAMET.T-K9 AND EURAMET.T-K9.1

 

CCT-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Mercury triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Wi   dimensionless resistance ratio measurement results of thermometer from laboratory i

ui   combined standard uncertainty of measurement made using thermometer from laboratory i are expressed in mK

The laboratory measurement results are listed in Appendix B of the CCT-K9 Final report.

 

SIM.T-K9.3

The measurement results are listed in Table 3 and 4 of the SIM.T-K9.3 Final report.

 

EURAMET.T-K9

The measurement results are listed in Section 4 of the EURAMET.T-K9 Final report.

 

EURAMET.T-K9.1

The measurement results are listed in Section 4 of the EURAME.T-K9.1 Final report.

CCT-K9 AND EURAMET.T-K9

 

MEASURAND  Resistance Ratio, W, at nominal temperature

NOMINAL TEMPERATURE  Argon triple point, a defined fixed point on the International Temperature Scale of 1990 (ITS-90)

 

Wi   dimensionless resistance ratio measurement results of thermometer from laboratory i

ui   combined standard uncertainty of measurement made using thermometer from laboratory i are expressed in mK

 

CCT-K9

The laboratory measurement results are listed in Appendix B of the CCT-K9 Final Report.

 

EURAMET.T-K9

The measurement results are listed in Section 4 of the EURAMET.T-K9 Final report.