BIPM.RI(II)-K1.Cs-134 and APMP.RI(II)-K2.Cs-134

MEASURAND   SIR equivalent activity of Cs-134

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq.

Cs-134
Di / kBq
Participating institutes

Results given in absolute values.

BIPM.RI(II)-K1.Cs-134 and APMP.RI(II)-K2.Cs-134

MEASURAND   SIR equivalent activity of Cs-134

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq.

Cs-134
Di / (kBq/MBq)
Participating institutes

Results given in relative values.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cs-134 and APMP.RI(II)-K2.Cs-134

MEASURAND   SIR equivalent activity of Cs-134

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq.

LABi Di Uneg,i Upos,i
kBq
kBq
kBq
NIST 18 62
JRC -76 77
BKFH 8 62
LNE-LNHB 1 42
NMIJ -19 41
BARC 20 95
CNEA 67 95
IFIN-HH 99 111
BEV -33 142
IRA -92 106
NMISA -22 58
POLATOM -16 77
NRC 8 85
LNMRI-IRD -36 77
PTB -42 53
INER 55 42
VNIIM -29 56
Participating institutes
BIPM.RI(II)-K1.Cs-134
APMP.RI(II)-K2.Cs-134

Results given in absolute values.

BIPM.RI(II)-K1.Cs-134 and APMP.RI(II)-K2.Cs-134

MEASURAND   SIR equivalent activity of Cs-134

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

LABi Di Uneg,i Upos,i
kBq/MBq
kBq/MBq
kBq/MBq
NIST 1.8 6.1
JRC -7.5 7.6
BKFH 0.8 6.1
LNE-LNHB 0.1 4.2
NMIJ -1.9 4.0
BARC 2 9
CNEA 7 9
IFIN-HH 10 11
BEV -3 14
IRA -9 10
NMISA -2.2 5.8
POLATOM -1.6 7.6
NRC 0.8 8.4
LNMRI-IRD -3.6 7.6
PTB -4.2 5.2
INER 5.4 4.1
VNIIM -2.9 5.5
Participating institutes
BIPM.RI(II)-K1.Cs-134
APMP.RI(II)-K2.Cs-134

Results given in relative values.

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Cs-134
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Cs-134
Time of measurements 1975 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
First results published on 08 July 2003
Most recent update: 13 January 2023


Earlier reports can be found in the participants list.
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

BARC

Bhabha Atomic Research Centre, India, APMP

BEV

Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET

CIEMAT

Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET

CNEA

National Commission of Atomic Energy, Argentina, SIM

IFIN-HH

"Horia Hulubei" National Institute of Research and Development for Physics and Nuclear Engineering, Romania, EURAMET

IRA

Institut Universitaire de Radiophysique Appliquée, Switzerland, EURAMET

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

LNE-LNHB

Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET

LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NPL

National Physical Laboratory, United Kingdom, EURAMET

Abandonned

NRC

National Research Council, Canada, SIM

POLATOM

Radioisotope Centre Polatom, Poland, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cs-134 and APMP.RI(II)-K2.Cs-134

MEASURAND   SIR equivalent activity of Cs-134

 

The key comparison reference value xR for this radionuclide is estimated to be

 

xR = 10 123 kBq with a standard uncertainty uR = 10 kBq

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence are both expressed in kBq, where xi is the equivalent activity for laboratory i.

The Degrees of equivalence are also given in relative terms by Di = (xI - xR) /xand its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Cs-134 and APMP.RI(II)-K2.Cs-134

MEASURAND   SIR equivalent activity of Cs-134

 

xi  measurement value submitted by laboratory iLabi

ui   combined standard uncertainty of xi

 

BIPM.RI(II)-K1.Cs-134

 

  Labi xi ui Date of measurement
    kBq kBq  
  NIST 10 141 31 2002-06-21
  JRC 10 047 39 2004-01-20
  BKFH 10 131 31 2004-12-08
  LNE-LNHB 10 124 20 2005-04-13
  NMIJ 10 104 19 2005-04-20
  BARC 10 143 48 2005-07-20
  CNEA 10 190 48 2005-09-12
  IFIN-HH 10 222 56 2006-11-21
  BEV 10 090 70 2008-11-24
  IRA 10 031 52 2009-02-25
  NMISA 10 101 29 2010-08-05
  POLATOM 10 107 39 2014-01-09
  NRC 10 131 43 2014-03-18
  LNMRI-IRD 10 087 39 2015-12-15
  PTB 10 081 26 2016-08-17

 

 

APMP.RI(II)-K2.Cs-134

         
  Labi xi u(xi) Date of measurement
  kBq kBq
  VNIIM 10 094 26 2005
  INER 10 178 18 2005