CCQM-K48.2014

MEASURAND : Amount content of chloride in potassium chloride
NOMINAL VALUE : 13.4 mol/kg

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2), expressed in relative terms

CCQM-K48.2014

MEASURAND : Amount content of chloride in potassium chloride
NOMINAL VALUE : 13.4 mol/kg

Degrees of equivalence, Di and expanded uncertainty Ui (k = 2), expressed in relative terms

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
CC comparison
Linked comparison
CCQM-K48.2014
Metrology area, Sub-field Chemistry and Biology, Electrochemistry
Description Chloride in KCl
Time of measurements 2014
Status Approved for equivalence
Final Reports of the comparisons
Measurand Amount content of chloride in KCl
Transfer device Dedicated samples
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by CCQM (Consultative Committee for Amount of Substance)
Comments

CCQM-K48.2014, Chloride in KCl, is a repeat of CCQM-K48. In was originally registered as CCQM-K114.

Results published on 1 September 2016

Assay of potassium chloride

Pilot institute NIM
National Institute of Metrology
China
Contact person MA LIANDI

+86 10 642 23 987
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIM

National Institute of Metrology, China, APMP

CENAM

Centro Nacional de Metrologia, Mexico, SIM

GUM

Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

UNIIM

Ural Scientific and Research Institute for Metrology, Rosstandart, Russian Federation, COOMET

VNIIFTRI

Institute of Physical Technical and Radiotechnical Measurements, Rosstandart, Russian Federation, COOMET

CCQM-K48.2014

MEASURAND : Amount content of chloride in potassium chloride
NOMINAL VALUE : 13.4 mol/kg

The key comparison reference value, xR, is calculated as the median of all participant results, and its standard uncertainty, uR, is calculated as 1.858*[median(xi-xR)]*(n-1)-1/2 , where n represents the total number of participants and xi is the result of participant i.

xR = 13.41311 mol/kg and uR = 0.00083 mol/kg

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di = (xi - xR)/xR and Ui, its expanded uncertainty (k = 2), Ui = [2(ui2 + uR2)1/2]/xR, both expressed in relative terms.

 

 

 

CCQM-K48.2014

MEASURAND : Amount content of chloride in potassium chloride
NOMINAL VALUE : 13.4 mol/kg

xi : result of measurement carried out by laboratory i
ui : combined standard uncertainty of xi