CCQM-K92 and SIM.QM-K92
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 0.05 S/m
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2), expressed in S/m
Red diamonds: participants in CCQM-K92
Blue squares: participants in SIM.QM-K92 only
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 20 S/m
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2), expressed in S/m
UINMETRO = 0.845 S/m
CCQM-K92 and SIM.QM-K92
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 0.05 S/m
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2), expressed in S/m
Dark blue: participants in CCQM-K92
Blue: participants in SIM.QM-K92 only
Results are presented under A4 printable format in Summary Results (PDF file).
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 20 S/m
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2), expressed in S/m
Results are presented under A4 printable format in Summary Results (PDF file).
Metrology area, Sub-field | Chemistry and Biology, Electrochemistry |
Description | Measurement of electrolytic conductivity |
Time of measurements | 2011 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Conductivity: 0.05 S/m and 20 S/m |
Transfer device | Dedicated samples |
Comparison type | Key Comparison |
Consultative Committee | CCQM (Consultative Committee for Amount of Substance) |
Conducted by | CCQM (Consultative Committee for Amount of Substance) |
Comments | Results published on 30 July 2013 SIM.QM-K92 results are linked to those of CCQM-K92 Electrolytic conductivity
|
Pilot institute |
SMU
Slovensky Metrologicky Ustav Slovakia |
Contact person | L. Vyskocil +421 2 602 94 225 |
Pilot laboratory | |
---|---|
SMU |
Slovensky Metrologicky Ustav, Slovakia, EURAMET |
CENAM |
Centro Nacional de Metrologia, Mexico, SIM |
CMI |
Czech Metrology Institute, Czechia, EURAMET |
DFM |
Danish Fundamental Metrology A/S, Denmark, EURAMET |
GUM |
Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET |
INMETRO |
Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM |
INPL |
National Physical Laboratory of Israel, Israel, N/A |
INRIM |
Istituto Nazionale di Ricerca Metrologica, Italy, EURAMET |
MKEH |
Hungarian Trade Licensing Office; now BFKH, Hungary, EURAMET |
NIM |
National Institute of Metrology, China, APMP |
NIST |
National Institute of Standards and Technology, United States, SIM |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
SE "Ukrmetrteststandard" |
State Enterprise "Ukrmetrteststandard", Ukraine, EURAMET |
SP |
Technical Research Institute of Sweden from 2017 Research Institutes of Sweden AB, Sweden, EURAMET |
VNIIFTRI |
Institute of Physical Technical and Radiotechnical Measurements, Rosstandart, Russian Federation, COOMET |
VNIIM |
D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCQM-K92 and SIM.QM-K92
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 0.05 S/m
The key comparison reference value, xR, is calculated as the weighted mean of the participants' results excluding outliers, and its standard uncertainty, uR, is calculated as explained on page 15 of the Final Report. |
xR = 0.0500304 S/m and uR = 0.0000061 S/m
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), |
There are no pair-wise degrees of equivalence computed for this key comparison. |
INMETRO and CENAM provide the link between key comparisons CCQM-K92 and SIM.QM-K92 as described on page 8 of the SIM.QM-K92 Final Report. |
The degree of equivalence of laboratory i participant in SIM.QM-K92 only relative to the CCQM-K92 key comparison reference value is given by a pair of terms: Di and Ui, its expanded uncertainty (k = 2), both expressed in S/m and computed according to Equations 2 and 3 of the SIM.QM-K92 Final Report. |
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 20 S/m
The key comparison reference value, xR, is calculated as the weighted mean of the participants' results excluding outliers, and its standard uncertainty, uR, is calculated as explained on page 15 of the Final Report. |
xR = 20.2964 S/m and uR = 0.0039 S/m
The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), |
There are no paire-wise degrees of equivalence computed for this key comparison. |
CCQM-K92 and SIM.QM-K92
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 0.05 S/m
xi : result of measurement carried out by laboratory i
ui : combined standard uncertainty of xi
MEASURAND : Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 0.05 S/m
xi : result of measurement carried out by laboratory i
ui : combined standard uncertainty of xi
MEASURAND :Â Electrolytic conductivity of a KCl solution
NOMINAL VALUE : 20 S/m
xi : result of measurement carried out by laboratory i
ui : combined standard uncertainty of xi