CCAUV.V-K3,  EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1 

MEASURAND:  Magnitude

FREQUENCY :  0.1 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2). 

Magnitude of complex sensitivity at 0.1 Hz
Di / (mV/(m/s2))
Participating institutes
CCAUV.V-K3
EURAMET.AUV.V-K3
APMP.AUV.V-K3.1

CCAUV.V-K3,  EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1 

MEASURAND:  Magnitude

FREQUENCY :  0.1 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mV / (m/s2). 

LABi Di Ui
(mV/(m/s2))
(mV/(m/s2))
NIM -0.62 0.66
PTB -0.20 0.67
BKSV-DPLA 0.34 0.67
NMISA 4.21 0.86
CENAM 0.22 0.95
NMIA 0.28 0.23
NMIJ -0.10 0.14
A*STAR -0.45 0.75
CMI 2.73 2.42
CMS -0.74 0.51
KRISS 0.64 1.9
NMISA -0.53 1.2
NPLI -0.57 1.5
Participating institutes
CCAUV.V-K3
EURAMET.AUV.V-K3
APMP.AUV.V-K3.1
CCAUV.V-K3
Metrology area, Sub-field Acoustics, Ultrasound, Vibration, Vibration
Description Acceleration complex sensitivity
Time of measurements 2013 - 2015
Status Approved for equivalence
Final Reports of the comparisons
References

CCAUV.V-K3 Technical Protocol (and Annex 1)

Measurand Acceleration complex sensitivity
Modulus, nominal value: 124 mV/(m/s2); and phase, nominal value: 0 degree

Parameters Different frequencies from 0.1 Hz to 40 Hz with specified acceleration amplitudes
Transfer device Low-frequency vibration comparison transfer standard
Comparison type Key Comparison
Consultative Committee CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Conducted by CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Comments Results published on 07 December 2016
Updated on 24 June 2019
Pilot institute NIM
National Institute of Metrology
China
Contact person Qiao SUN

+86 10 645 24 623
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIM

National Institute of Metrology, China, APMP

CENAM

Centro Nacional de Metrologia, Mexico, SIM

GUM

Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET

HBK-DPLA

Hottinger Brüel & Kjaer A/S - Danish Primary Laboratory for Acoustics Section, Denmark, EURAMET

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

LNE

Laboratoire national de métrologie et d'essais, France, EURAMET

METAS

Federal Institute of Metrology, Switzerland, EURAMET

NMC, A*STAR

National Metrology Centre, Agency for Science, Technology and Research, Singapore, APMP

NMIA

National Measurement Institute, Australia, Australia, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

CCAUV.V-K3,  EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1 

MEASURAND:  Magnitude

FREQUENCY :  0.1 Hz

 

CCAUV.V-K3

The computation of the key comparison reference value at each frequency, xR, and of its expanded uncertainty, UR (k = 2), is explained in Section 8 of the CCAUV.V-K3 Final Report and is based on a weighted mean.

 

        xR = 136.75 mV / (m/s2)      UR = 0.15 mV / (m/s2)

 

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di where Di = xi - xR, and its expanded uncertainty Ui (k = 2), computed as explained in Section 8 of the CCAUV.V-K3 Final Report.

 

Linking of EURAMET.AUV.V-K3 to CCAUV.V-K3

The linking was made via BKSV-DPLA, GUM, LNE, METAS and PTB, having participated in both comparisons. The linking process is described in Section 2 of the linkage report of EURAMET.AUV.V-K3 to CCAUV.V-K3.

 

Linking of APMP.AUV.V-K3.1 to CCAUV.V-K3

The linking was made via NIM,  having participated in both comparisons. The linking process is described in Section 9 of the APMP.AUV.V-K3.1 Final Report.

CCAUV.V-K3,  EURAMET.AUV.V-K3 and APMP.AUV.V-K3.1 

MEASURAND:  Magnitude

FREQUENCY :  0.1 Hz

 

CCAUV.V-K3

The laboratories' individual measurement may be found in Section 7 of the CCAUV.V-K3 Final Report.

EURAMET.AUV.V-K3

The laboratories' individual measurement may be found in Section 7 of the EURAMET.AUV.V-K3 Final Report.

APMP.AUV.V-K3.1

The laboratories' individual measurement may be found in Section 7 of the APMP.AUV.V-K3.1 Final Report.

 

 