CCM.FF-K4.2.2011, EURAMET.M.FF-K4.2.2014, and AFRIMETS.M.FF-K4.2.2015
MEASURAND Water volume
NOMINAL VALUE 100 µL
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µL.
Five micropipettes, serial numbers 354828Z, 354853Z, 354864Z, 354868Z and 354872Z, were used in CCM.FF-K4.2.2011.
CCM.FF-K4.2.2011 and EURAMET.M.FF-K4.2.2014: Data obtained for 354828Z are listed here, used in both comparisons. For other micro- pipettes, see the Final Reports.
CCM.FF-K4.2.2011and AFRIMETS.M.FF-K4.2.2015: Data obtained for 354868Z are listed here for AFRIMETS.M.FF-K4.2.2015, used in CCM.FF-K4.2.2011and AFRIMETS.M.FF-K4.2.2015. For other micro- pipettes, see the Final Reports.
CCM.FF-K4.2.2011, EURAMET.M.FF-K4.2.2014, and AFRIMETS.M.FF-K4.2.2015
MEASURAND Water volume
NOMINAL VALUE 100 µL
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µL.
Five micropipettes, serial numbers 354828Z, 354853Z, 354864Z, 354868Z and 354872Z, were used in CCM.FF-K4.2.2011.
CCM.FF-K4.2.2011 and EURAMET.M.FF-K4.2.2014: Data obtained for 354828Z are listed here, used in both comparisons. For other micro- pipettes, see the Final Reports.
CCM.FF-K4.2.2011and AFRIMETS.M.FF-K4.2.2015: Data obtained for 354868Z are listed here for AFRIMETS.M.FF-K4.2.2015, used in CCM.FF-K4.2.2011and AFRIMETS.M.FF-K4.2.2015. For other micro- pipettes, see the Final Reports.
Metrology area, Sub-field | Mass and related quantities, Fluid Flow |
Description | Liquid volume |
Time of measurements | 2011 - 2012 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Liquid volume: 100 µL |
Transfer device | Five 100 µL micropipettes |
Comparison type | Key Comparison |
Consultative Committee | CCM (Consultative Committee for Mass and Related Quantities) |
Conducted by | CCM (Consultative Committee for Mass and Related Quantities) |
Comments | Results published on 01 February 2013 Results of EURAMET.M.FF-K4.2.2014 are linked to those of CCM.FF-K4.2.2011 Liquid volume using five 100 µL micropipettes
|
Pilot institute |
IPQ
Instituto Portugues da Qualidade Portugal |
Contact person | E. Batista +35 1 212 948 167 |
Pilot laboratory | |
---|---|
IPQ |
Instituto Portugues da Qualidade, Portugal, EURAMET |
CENAM |
Centro Nacional de Metrologia, Mexico, SIM |
INMETRO |
Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM |
LNE |
Laboratoire national de métrologie et d'essais, France, EURAMET |
NIM |
National Institute of Metrology, China, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
NMO |
National Measurement Office, now National Measurement and Regulation Office, NMRO, United Kingdom, EURAMET |
UME |
TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCM.FF-K4.2.2011, EURAMET.M.FF-K4.2.2014, and AFRIMETS.M.FF-K4.2.2015
MEASURAND Water volume
NOMINAL VALUE 100 µL
The key comparison reference value xR and the associated expanded uncertainty (k = 2) UR for each pipette is basically obtained from the weighted means of the results.
For each micropipette, the degree of equivalence of laboratory i with respect to the key comparison reference value is computed according to equations 8, 9 and 10 on page 11 of the Final Report, and is given by a pair of terms, both expressed in µL: Di and Ui, its expanded uncertainty (k = 2).
Linking EURAMET.M.FF-K4.2.2014 to CCM.FF-K4.2.2011
TRANFERT INSTRUMENTS: Two micropipettes, serial numbers 354828Z and 354853Z
EURAMET.M.FF-K4.2.2014 is linked to CCM.FF-K4.2.2011 via IPQ, having participated in both comparisons.
Linking AFRIMETS.M.FF-K4.2.2015 to CCM.FF-K4.2.2011
TRANFERT INSTRUMENTS: Two micropipettes, serial numbers 354868Z and 354872Z
AFRIMETS.M.FF-K4.2.2015 is linked to CCM.FF-K4.2.2011 via IPQ, having participated in both comparisons.
CCM.FF-K4.2.2011, EURAMET.M.FF-K4.2.2014, and AFRIMETS.M.FF-K4.2.2015
MEASURAND Water volume
NOMINAL VALUE 100 µL
CCM.FF-K4.2.2011
V Lab i volume measured by laboratory i after correction to the standard atmospheric pressure
U Lab i expanded uncertainty (k = 2) of V Lab i
EURAMET.M.FF-K4.2.2014
TRANFERT INSTRUMENTS: Two micropipettes, serial numbers 354828Z and 354853Z
The measurement results obtained in EURAMET.M.FF-K4.2.2014 are listed in table 10 on page 11 of the EURAMET.M.FF-K4.2.2014 Final report.
AFRIMETS.M.FF-K4.2.2015
TRANFERT INSTRUMENTS: Two micropipettes, serial numbers 354868Z and 354872Z
The measurement results obtained in AFRIMETS.M.FF-K4.2.2015 are listed in the AFRIMETS.M.FF-K4.2.2015 Final report.