BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5
MEASURAND Air kerma
Radiation quality Cs-137 beam
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2) expressed in mGy/Gy
BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5
MEASURAND Air kerma
Radiation quality Cs-137 beam
Degrees of equivalence, Di and expanded uncertainty Ui (k = 2) expressed in mGy/Gy
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mGy/Gy |
mGy/Gy |
mGy/Gy |
||
GUM | -0.5 | 5.8 | ||
IST/ITN | 1.3 | 4.2 | ||
LNE-LNHB | -1.6 | 5.2 | ||
MKEH | 5.3 | 5.0 | ||
VNIIM | 1.3 | 5.4 | ||
KRISS | -1.4 | 4.4 | ||
NIST | -1.0 | 7.0 | ||
NMIJ | -2.3 | 5.2 | ||
PTB | 2.4 | 5.4 | ||
NIM | -3.3 | 4.2 | ||
ININ | 4.8 | 4.0 | ||
VSL | -4.8 | 7.6 | ||
SMU | 5.1 | 5.4 | ||
BEV | 4.3 | 5.4 | ||
CIEMAT | -0.8 | 4.4 | ||
INER | -4.6 | 8.9 | ||
VINS | -3.3 | 20.2 |
Metrology area, Sub-field | Ionizing Radiation, Section I (x and gamma rays, electrons) |
Description | Measurement of air kerma for Caesium 137 |
Time of measurements | 1994 - |
Status | Continuous, approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Air kerma |
Parameters | Radiation type: Cs-137 gamma rays |
Transfer device | Ionization chambers |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | BIPM (Bureau International des Poids et Mesures) |
Comments | The Final Report for each bi-lateral comparsion is listed in the tab "Participants" First results published on 29 November 2012 Last update: 26 January 2024 |
Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
Contact person | C. Kessler +33 (0) 1 45 07 70 84 |
Pilot laboratory | |
---|---|
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
BEV |
Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET |
1995 | |
2009 | |
2023 |
CIEMAT |
Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET |
2023 |
ENEA-INMRI |
Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Italy, EURAMET |
1998 |
GUM |
Glowny Urzad Miar, Central Office of Measures, Poland, EURAMET |
2006 |
ININ |
Instituto Nacional de Investigaciones Nucleares, Mexico, SIM |
2015 |
ITN |
Instituto Tecnólogico e Nuclear, now IST-LPSR, Instituto Tecnológico e Nuclear, Laboratório de Proteção e Segurança Radiológica, Campus Tecnológico e Nuclear, Portugal, EURAMET |
2008 |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
2010 |
LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
1995 | |
2008 |
NIM |
National Institute of Metrology, China, APMP |
2014 |
NIST |
National Institute of Standards and Technology, United States, SIM |
1994 | |
2011 |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
2001 | |
2012 |
OMH |
Országos Mérésügyi Hivatal; Subsequently MKEH, now BKFH, Hungary, EURAMET |
1994 | |
2009 |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
2000 | |
2014 |
SMU |
Slovensky Metrologicky Ustav, Slovakia, EURAMET |
2017 |
VNIIM |
D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET |
1997 | |
2009 |
VSL |
VSL, Netherlands, EURAMET |
2016 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5
MEASURAND Air kerma
Radiation quality Cs-137 beam
The key comparison reference value is the BIPM evaluation of absrobed dose to water.
BIPM.RI(I)-K5
The degree of equivalence of each laboratory i with respect to the key comparison reference value is given by a pair of terms both expressed in mGy/Gy:
the relative difference Di and Ui, its expanded uncertainty (k = 2), with Ui = 2ui.
When required, the degree of equivalence between two laboratories i and j can be evaluated by a pair of terms both expressed in mGy/Gy:
Dij = Di - Dj and Uij , its expanded uncertainty (k = 2). The approximation for Uij is explained in section 5 of the Final Report.
Linking EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5
The linking of EURAMET.RI(I)-K5.1 to BIPM.RI(I)-K5 is made via BEV (Austria) who participated in both comparisons.
Linking APMP.RI(I)-K5 to BIPM.RI(I)-K5
The linking of APMP.RI(I)-K5 to BIPM.RI(I)-K5 is made via KRISS (Korea) and NMIJ AIST (Japan) who participated in both comparisons.
BIPM.RI(I)-K5, EURAMET.RI(I)-K5.1 and APMP.RI(I)-K5
MEASURAND Air kerma
Radiation quality Cs-137 beam
xi ratio of the measurement of laboratory i to that of the BIPM
ui combined standard uncertainty of the ratio xi
BIPM.RI(I)-K5
Labi | xi | ui | year of measurement | |
GUM | 0.9995 | 0.0029 | 2006 | |
IST/ITN | 1.0013 | 0.0021 | 2008 | |
LNE-LNHB | 0.9984 | 0.0026 | 2008 | |
MKEH | 1.0053 | 0.0025 | 2009 | |
VNIIM | 1.0013 | 0.0027 | 2009 | |
KRISS | 0.9986 | 0.0022 | 2010 | |
NIST | 0.9990 | 0.0035 | 2011 | |
NMIJ | 0.9977 | 0.0026 | 2012 | |
PTB | 1.0024 | 0.0027 | 2014 | |
NIM | 0.9967 | 0.0021 | 2014 | |
ININ | 1.0048 | 0.0020 | 2015 | |
VSL | 0.9952 | 0.0038 | 2016 | |
SMU | 1.0051 | 0.0027 | 2017 | |
BEV | 1.0043 | 0.0027 | 2023 | |
CIEMAT | 0.9992 | 0.0022 | 2023 |
EURAMET.RI(I)-K5.1
Labi | xi | ui | year of measurement | |
VINS | 0.9967 | 0.0101 | 2017 |
APMP.RI(I)-K5
Labi | xi | ui | year of measurement | |
KRISS | 0.9931 | 0.0041 | 2014 | |
NMIJ | 1.0033 | 0.0043 | 2014 | |
INER | 0.9954 | 0.0045 | 2014 | |
CIEMAT | 1.0042 | 0.0086 | 2014 | |
NIM | 1.0013 | 0.0044 | 2014 |