Comparison
Comparison close

CCPR-K3.a, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1, COOMET.PR-K3.a and APMP.PR-K3.a

MEASURAND       Luminous intensity

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-2.

Luminous intensity
Di
Participating institutes

Enlarged graph

Graph

Comparison
Comparison close

CCPR-K3.a, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1, COOMET.PR-K3.a and APMP.PR-K3.a

MEASURAND       Luminous intensity

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-2.

LABi Di Uneg,i Upos,i
10-2
10-2
10-2
LNE-INM/Cnam 0.89 0.60
NMISA 0.51 0.88
NMIA -0.07 0.60
NMIJ -0.09 0.56
INRIM -0.43 0.90
IO-CSIC -0.48 0.72
NIM -0.16 0.48
NIST 0.12 0.48
NPL 0.04 0.30
NRC 0.19 1.06
METAS -0.50 1.12
MKEH 0.05 0.94
PTB -0.31 0.40
SMU -0.36 1.40
VNIIOFI 0.33 0.92
INTI -0.25 0.74
BIPM 0.30 1.00
NMC, A*STAR 0.03 0.33
BIM 0.04 1.50
BelGIM -0.53 0.72
SP -0.61 0.93
UME 0.43 1.50
BEV -0.26 1.10
VSL -0.54 0.64
CMI -0.10 0.80
IPQ 0.57 1.60
DMDM 0.22 1.20
GUM 0.33 1.10
INM -0.21 1.40
SMU 0.42 0.72
MIKES -0.32 0.44
INM -0.72 0.90
BelGIM -0.57 0.94
NSC IM -0.66 1.28
KazStandard -0.22 0.94
CMS 1.93 0.86
SCL -0.48 1.30
SNSU-BSN -2.11 1.10
NPLI -0.61 0.88
NMIM -0.73 1.10
MSL -0.08 0.59
NIMT 0.61 0.87
NMC, A*STAR -0.09 0.71
VMI 0.28 1.00
NMISA -1.28 1.60
NIS -1.25 2.20
Participating institutes
CCPR-K3.a
CCPR-K3.a.1
EURAMET.PR-K3.a
EURAMET.PR-K3.a.1
COOMET.PR-K3.a
APMP.PR-K3.a
CCPR-K3.a.1
Metrology area, Sub-field Photometry and Radiometry, Photometry
Description Luminous intensity (Bilateral comparison)
Time of measurements 2003
Status Approved for equivalence
Final Reports of the comparisons
Measurand Luminous intensity
Transfer device Three travelling standard lamps
Comparison type Key Comparison
Consultative Committee CCPR (Consultative Committee for Photometry and Radiometry)
Conducted by CCPR (Consultative Committee for Photometry and Radiometry)
Comments Results published on 1 March 2005.
Pilot institute NMIA
National Measurement Institute, Australia
Australia
Contact person M. Ballico

+61 2 9413 7300
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
CSIRO-NML

Commonwealth Scientific and Industrial Research Organization - National Measurement Laboratory; became National Measurement Institute, Australia, NMIA in 2004, Australia, APMP

SPRING Singapore

Singapore Standards, Productivity and Innovation Board (became NMC, A*STAR in 2008), Singapore, APMP

Comparison
Comparison close

CCPR-K3.a, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1, COOMET.PR-K3.a and APMP.PR-K3.a

MEASURAND       Luminous intensity

 

The individual measurements, xi, of the participating laboratories take the form of ratios which depend on the reference used at the pilot laboratory. This is cancelled out by normalization using the key comparison reference value, xR. This procedure leads to normalized individual measurements: xi / xR.

 

CCPR-K3.a

The key comparison reference value, xR, is calculated as the weighted average of the invidual results xi, weighted by the inverse square of the individual standard uncertainties, ui, with the application of a minimum cutoff of 0.25%. The INTI and the BIPM are excluded from the calculation of xR. The standard uncertainty of xR is uR = 0.09 %. It is negligible compared to the ui values.

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: Di = (xi - xR) / xR and Ui, its expanded uncertainty
(k = 2), both expressed in relative units. Ui = 2ui.

 

Linking CCPR-K3.a.1 to CCPR-K3.a

The link to the key comparison CCPR-K3.a is made though CSIRO-NML (now NMIA) that participated in both comparisons, see Section 9 of the CCPR-K3.a.1 Final Report.

 

Linking EURAMET.PR-K3.a to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realized through the common participation of LNE-INM/Cnam, INRiM (named "BNM-INM" and "IEN" at the time of the CCPR key comparison) and PTB that participated in both comparisons, see EURAMET.PR-K3.a Final Report.

 

Linking EURAMET.PR-K3.a.1 to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realized through the common participation of LNE see EURAMET.PR-K3.a.1 Final Report.

 

Linking COOMET.PR-K3.a to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realized through the common participation of SMU and VNIIOFI, see COOMET.PR-K3.a Final Report.

 

Linking APMP.PR-K3.a to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realised through the common participation of NMIJ and NIM, see Section 7 of the APMP.PR-K3.a Final Report.

 

Comparison
Comparison close

CCPR-K3.a, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1, COOMET.PR-K3.a and APMP.PR-K3.a

MEASURAND       Luminous intensity

 

For individual measurements see the Final Reports.