CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
EURAMET.M.P-K7
Metrology area, Sub-field Mass and related quantities, Pressure
Description Pressure measurements in oil (gauge mode)
Time of measurements 2005 - 2007
Status Approved for equivalence
Final Reports of the comparisons
Measurand Pressure: 50 MPa to 500 MPa
Transfer device Pressure balance
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET Project No 881
Comments

Results published on 29 July 2019

APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7 results are linked to those of CCM.P-K13

Hydraulic gauge pressure

Pilot institute MIKES
VTT Technical Research Centre of Finland Ltd, Centre for Metrology/Mittatekniikan keskus
Finland
Contact person S. Saxholm

+358 50 4105 499
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
CMI

Czech Metrology Institute, Czechia, EURAMET

METAS

Federal Institute of Metrology, Switzerland, EURAMET

MIKES/VTT

Mittatekniikan Keskus, Centre for Metrology and Accreditation, Finland, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

SMD

Service de la Métrologie Belge, Belgium, EURAMET

SP

Technical Research Institute of Sweden from 2017 Research Institutes of Sweden AB, Sweden, EURAMET

VSL

VSL, Netherlands, EURAMET

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961152 mm2
uR / xR = 13.5 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 0.5 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

1.961187

17

Jan 2009

 CENAM 1.961171 85 Mar 2009
 NPLI 1.961104 44 May 2009
 NIM 1.961098 17 Jun 2009
 NMIJ 1.961152 17 Oct 2009
 LNE 1.961191 11 Feb 2010
 PTB 1.961148 15 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

We use cookies to optimize our website. View privacy policy