CCT-K10
MEASURAND : Temperature
TEMPERATURE : 962 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1100 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1300 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1500 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1700 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1800 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2000 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2200 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2400 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2500 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2600 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2800 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2900 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 3000 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1325 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1952 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 2747 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 962 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | -0.02 | 0.21 | ||
NIM | -0.11 | 0.41 | ||
KRISS | 0.03 | 0.24 | ||
NRC | -0.14 | 0.56 | ||
NIST | 0.05 | 0.59 | ||
LNE-Cnam | 0.01 | 0.22 | ||
PTB | -0.02 | 0.31 | ||
CEM | -0.04 | 0.40 | ||
NPL* | 0.07 | 0.50 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1100 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.00 | 0.24 | ||
NIM | -0.12 | 0.48 | ||
KRISS | 0.01 | 0.24 | ||
NRC | -0.14 | 0.62 | ||
NIST | 0.02 | 0.62 | ||
LNE-Cnam | 0.03 | 0.25 | ||
PTB | 0.03 | 0.41 | ||
CEM | -0.01 | 0.46 | ||
NPL* | 0.07 | 0.56 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1300 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | -0.01 | 0.29 | ||
NIM | -0.10 | 0.55 | ||
KRISS | 0.09 | 0.30 | ||
NRC | -0.15 | 0.75 | ||
NIST | -0.09 | 0.68 | ||
LNE-Cnam | -0.02 | 0.30 | ||
PTB | 0.08 | 0.54 | ||
CEM | -0.13 | 0.58 | ||
NPL* | 0.10 | 0.66 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1500 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.00 | 0.41 | ||
NIM | -0.09 | 0.68 | ||
KRISS | 0.07 | 0.44 | ||
NRC | -0.14 | 0.93 | ||
NIST | 0.06 | 0.81 | ||
LNE-Cnam | -0.03 | 0.41 | ||
PTB | 0.08 | 0.70 | ||
CEM | -0.39 | 0.75 | ||
NPL* | 0.10 | 0.78 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1700 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.08 | 0.47 | ||
NIM | 0.01 | 0.78 | ||
KRISS | 0.16 | 0.63 | ||
NRC | -0.02 | 1.13 | ||
NIST | 0.00 | 0.93 | ||
LNE-Cnam | -0.01 | 0.48 | ||
PTB | -0.30 | 0.82 | ||
CEM | -0.46 | 0.90 | ||
NPL* | 0.12 | 0.92 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1800 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.13 | 0.53 | ||
NIM | 0.09 | 0.83 | ||
KRISS | 0.20 | 0.66 | ||
NRC | 0.05 | 1.25 | ||
NIST | -0.08 | 1.01 | ||
LNE-Cnam | -0.04 | 0.53 | ||
PTB | -0.26 | 0.88 | ||
CEM | -0.72 | 0.99 | ||
NPL* | 0.09 | 1.02 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2000 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.09 | 0.65 | ||
NIM | 0.00 | 1.10 | ||
KRISS | 0.13 | 0.85 | ||
NRC | 0.08 | 1.53 | ||
NIST | -0.22 | 1.18 | ||
LNE-Cnam | -0.17 | 0.63 | ||
PTB | -0.51 | 0.94 | ||
CEM | -0.52 | 1.18 | ||
NPL* | 0.22 | 1.21 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2200 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.05 | 0.78 | ||
NIM | 0.10 | 1.22 | ||
KRISS | 0.09 | 1.11 | ||
NRC | 0.13 | 1.86 | ||
NIST | -0.49 | 1.36 | ||
LNE-Cnam | -0.33 | 0.74 | ||
PTB | -0.37 | 1.12 | ||
CEM | -0.57 | 1.60 | ||
NPL* | 0.03 | 1.43 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2400 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.26 | 1.08 | ||
NIM | 0.93 | 1.50 | ||
KRISS | 0.22 | 1.43 | ||
NRC | 0.39 | 2.30 | ||
NIST | -0.50 | 1.66 | ||
LNE-Cnam | -0.42 | 1.01 | ||
PTB | -0.56 | 1.42 | ||
CEM | -1.64 | 1.93 | ||
NPL* | 0.16 | 1.72 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2500 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.19 | 1.11 | ||
NIM | 1.46 | 1.55 | ||
KRISS | -0.05 | 1.53 | ||
NRC | 0.35 | 2.47 | ||
NIST | -0.48 | 1.74 | ||
PTB | -0.91 | 1.44 | ||
CEM | -0.38 | 2.01 | ||
NPL* | 0.04 | 1.86 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2600 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.17 | 1.15 | ||
NIM | 1.93 | 1.80 | ||
KRISS | -0.13 | 1.63 | ||
NRC | 0.38 | 2.67 | ||
NIST | -0.65 | 1.83 | ||
LNE-Cnam | -0.37 | 1.04 | ||
PTB | -1.10 | 1.49 | ||
NPL* | 0.14 | 1.93 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2800 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.12 | 1.45 | ||
NIM | 2.42 | 2.44 | ||
KRISS | -0.17 | 2.02 | ||
NRC | 0.24 | 3.19 | ||
NIST | -0.36 | 2.15 | ||
LNE-Cnam | -0.47 | 1.31 | ||
PTB | -1.60 | 1.81 | ||
NPL* | 0.94 | 2.35 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2900 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | -0.31 | 1.77 | ||
NIM | 2.69 | 3.19 | ||
KRISS | -0.34 | 2.42 | ||
NRC | 0.14 | 3.51 | ||
PTB | -2.24 | 2.11 | ||
NPL* | 0.60 | 3.17 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 3000 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | -0.68 | 1.65 | ||
KRISS | -0.56 | 2.37 | ||
NRC | 0.16 | 3.67 | ||
NPL* | 0.45 | 3.55 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1325 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.25 | 0.23 | ||
NIM | 0.01 | 0.28 | ||
KRISS | 0.09 | 0.25 | ||
NRC | 0.35 | 0.98 | ||
NIST | -0.26 | 0.74 | ||
LNE-Cnam | -0.04 | 0.33 | ||
PTB | -0.50 | 0.43 | ||
CEM | -0.36 | 0.49 | ||
NPL* | 0.01 | 0.43 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1952 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.32 | 0.47 | ||
NIM | -0.01 | 0.50 | ||
KRISS | 0.24 | 0.68 | ||
NRC | -0.11 | 1.82 | ||
NIST | -0.31 | 1.06 | ||
LNE-Cnam | -0.11 | 0.40 | ||
PTB | -0.13 | 0.68 | ||
CEM | -0.70 | 0.89 | ||
NPL* | -0.03 | 0.70 |
*Difference from the KCRV of the average NPL results.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 2747 °C
Degrees of equivalence represented by Di and its expanded uncertainty Ui both expressed in °C.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
°C |
°C |
°C |
||
NMIJ | 0.18 | 0.90 | ||
NIM | -0.05 | 0.96 | ||
KRISS | 0.20 | 1.64 | ||
NRC | -0.22 | 3.47 | ||
NIST | -0.94 | 1.83 | ||
LNE-Cnam | 0.13 | 0.73 | ||
CEM | 0.34 | 1.69 | ||
NPL* | -0.45 | 1.30 |
*Difference from the KCRV of the average NPL results.
Metrology area, Sub-field | Thermometry, Pyrometry |
Description | Realizations of the ITS-90 between 960 °C and 3000 °C |
Time of measurements | 2014 - 2016 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | Temperature: 960 °C, 1100 °C, 1300 °C, 1330 °C, 1500 °C, 1700 °C, 1800 °C, 1953 °C, 2000 °C, 2200 °C, 2400 °C, 2500 °C, 2600 °C, 2750 °C, 2800 °C, 2900 °C, 3000 °C |
Parameters | ITS-90 radiance temperature |
Transfer device | Two transfer radiation thermometers and a set of high temperature fixed-point blackbody cells |
Comparison type | Key Comparison |
Consultative Committee | CCT (Consultative Committee for Thermometry) |
Conducted by | CCT (Consultative Committee for Thermometry) |
Comments | Results published on 26 June 2024. |
Pilot institute |
NPL
National Physical Laboratory United Kingdom |
Contact person | Helen McEvoy +44 20 8943 6183 |
Pilot laboratory | |
---|---|
NPL |
National Physical Laboratory, United Kingdom, EURAMET |
CEM |
Centro Español de Metrologia, Spain, EURAMET |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
LNE-LCM/Cnam |
Conservatoire National des Arts et Métiers/Laboratoire Commun de Métrologie, France, EURAMET |
NIM |
National Institute of Metrology, China, APMP |
NIST |
National Institute of Standards and Technology, United States, SIM |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NRC |
National Research Council, Canada, SIM |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
VNIIM |
D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 962 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 4.19878 × 10-11 A | 0.04 °C | 0.06 °C |
Chino thermometer | 0.004871 V | 0.03 °C | 0.10 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui (at 95% level of confidence) are computed according to Section 11 of the CCT-K10 Final report, where for each participant at each temperature, Di has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values, from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1100 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 2.60581 ×10-10 A | 0.02 °C | 0.06 °C |
Chino thermometer | 0.030199 V | 0.04 °C | 0.11 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui (at 95% level of confidence) are computed according to Section 11 of the CCT-K10 Final report, where for each participant at each temperature, Di has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values, from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1300 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 2.01294 ×10-9 A | 0.07 °C | 0.10 °C |
Chino thermometer | 0.233136 V | 0.02 °C | 0.13 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1500 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 9.80505 ×10-9 A | 0.07 °C | 0.11 °C |
Chino thermometer | 1.135095 V | 0.09 °C | 0.18 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1700 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 3.46361 ×10-8 A | 0.08 °C | 0.13 °C |
Chino thermometer | 4.008140 V | 0.06 °C | 0.19 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1800 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 5.94185 ×10-8 A | 0.11 °C | 0.16 °C |
Chino thermometer | 6.874136 V | 0.08 °C | 0.21 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2000 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 1.51734 ×10-7 A | 0.10 °C | 0.17 °C |
Chino thermometer | 1.747312 V | 0.10 °C | 0.25 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2200 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 3.32893 ×10-7 A | 0.15 °C | 0.22 °C |
Chino thermometer | 3.835241 V | 0.12 °C | 0.30 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2400 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 6.49278 ×10-7 A | 0.21 °C | 0.28 °C |
Chino thermometer | 7.473960 V | 0.30 °C | 0.44 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2500 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 8.74685 ×10-7 A | 0.19 °C | 0.32 °C |
Chino thermometer | 1.008179 V | 0.27 °C | 0.43 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2600 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 1.15413 ×10-6 A | 0.19 °C | 0.34 °C |
Chino thermometer | 1.330654 V | 0.23 °C | 0.43 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2800 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 1.90405 ×10-6 A | 0.23 °C | 0.39 °C |
Chino thermometer | 2.195758 V | 0.39 °C | 0.57 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2900 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 2.39099 ×10-6 A | 0.25 °C | 0.42 °C |
Chino thermometer | 2.756808 V | 0.59 °C | 0.74 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 3000 °C
For each transfer thermometer, the key comparison reference value xR was calculated using the median of the reported results and its associated standard uncertainty uR calculated using the 'median of absolute deviations (MAD)' method as described in Section 10.3 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for drift, corrections and uncertainty in the median value.
Thermometer | xR | uR | uR-total |
LP3 thermometer | 2.95777 ×10-6 A | 0.24 °C | 0.43 °C |
Chino thermometer | 3.416361 V | 0.42 °C | 0.63 °C |
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di and its expanded uncertainty Ui are computed according to Section 11 of the CCT-K10 Final report, where for each participant, Di at each temperature has been calculated by taking the average (simple mean) of the differences between each participant's LP3 and Chino thermometer values from their respective reference values (medians). The uncertainty Ui at each temperature was taken to be, for that participant, the largest total uncertainty (laboratory uncertainty combined with the uncertainty in the reference value); that is, the larger of the total uncertainty for the LP3 and the total uncertainty for the Chino thermometer. Where data for one of the thermometers had been withdrawn from the comparison, the uncertainty was simply the total uncertainty related to the remaining thermometer.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1325 °C
The key comparison reference value xR and its associated standard uncertainty uR are calculated from the weighted mean with cut-off according to Section 12 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for the furnace, cell stability, corrections and uncertainty in the weighted mean value.
xR = 1324.75 °C uR = 0.04 °C uR-total = 0.16 °C
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di = (xi - xR) and Ui its expanded uncertainty (at 95%level of confidence) is computed according to equation (14) in Section 12 of the CCT-K10 Final report.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1952 °C
The key comparison reference value xR and its associated standard uncertainty uR are calculated from the weighted mean with cut-off according to Section 12 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for the furnace, cell stability, corrections and uncertainty in the weighted mean value.
xR = 1952.99 °C uR = 0.09 °C uR-total = 0.19 °C
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di = (xi - xR) and its expanded uncertainty Ui are computed according to section 12 of the CCT-K10 Final report.
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 2747 °C
The key comparison reference value xR and its associated standard uncertainty uR are calculated from the weighted mean with cut-off according to Section 12 of the CCT-K10 Final report. uR-total is the total uncertainty of the reference value, including components for the furnace, cell stability, corrections and uncertainty in the weighted mean value.
xR = 2747.25 °C uR = 0.18 °C uR-total = 0.35 °C
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in °C:
Di = (xi - xR) and Ui its expanded uncertainty (at 95%level of confidence) is computed according to equation (14) in Section 12 of the CCT-K10 Final report.
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 962 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 962 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 4.20018E-11 | 0.04 | 0.07 | |
NIM | 4.19529E-11 | 0.14 | 0.15 | |
KRISS | 4.19878E-11 | 0.07 | 0.09 | |
NRC | 4.19202E-11 | 0.27 | 0.27 | |
NIST | 4.20188E-11 | 0.28 | 0.29 | |
LNE-Cnam | 4.20025E-11 | 0.05 | 0.08 | |
PTB | 4.19539E-11 | 0.12 | 0.14 | |
CEM | 4.19262E-11 | 0.17 | 0.18 | |
NPL | 4.20726E-11 | 0.18 | 0.18 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 962 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 0.004867 | 0.05 | 0.11 | |
NIM | 0.004859 | 0.18 | 0.20 | |
KRISS | 0.004876 | 0.07 | 0.12 | |
NRC | 0.004860 | 0.27 | 0.28 | |
NIST | 0.004875 | 0.28 | 0.30 | |
LNE-Cnam | 0.004871 | 0.05 | 0.11 | |
PTB | 0.004872 | 0.12 | 0.15 | |
CEM | 0.004872 | 0.17 | 0.20 | |
NPL | 0.004871 | 0.23 | 0.25 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1100 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 1100 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 2.60644E-10 | 0.04 | 0.07 | |
NIM | 2.60475E-10 | 0.15 | 0.16 | |
KRISS | 2.60648E-10 | 0.04 | 0.08 | |
NRC | 2.60236E-10 | 0.29 | 0.30 | |
NIST | 2.60499E-10 | 0.29 | 0.29 | |
LNE-Cnam | 2.60581E-10 | 0.06 | 0.09 | |
PTB | 2.60735E-10 | 0.17 | 0.19 | |
CEM | 2.60156E-10 | 0.20 | 0.21 | |
NPL | 2.61033E-10 | 0.22 | 0.22 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 1100 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 0.030190 | 0.05 | 0.12 | |
NIM | 0.030123 | 0.21 | 0.24 | |
KRISS | 0.030196 | 0.05 | 0.12 | |
NRC | 0.030136 | 0.29 | 0.31 | |
NIST | 0.030220 | 0.29 | 0.31 | |
LNE-Cnam | 0.030219 | 0.06 | 0.13 | |
PTB | 0.030203 | 0.17 | 0.21 | |
CEM | 0.030243 | 0.20 | 0.23 | |
NPL | 0.030199 | 0.26 | 0.28 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1300 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 1300 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 2.01316E-09 | 0.06 | 0.12 | |
NIM | 2.01294E-09 | 0.17 | 0.20 | |
KRISS | 2.01358E-09 | 0.08 | 0.13 | |
NRC | 2.01043E-09 | 0.35 | 0.36 | |
NIST | 2.01103E-09 | 0.32 | 0.33 | |
LNE-Cnam | 2.01181E-09 | 0.09 | 0.13 | |
PTB | 2.01568E-09 | 0.23 | 0.25 | |
CEM | 2.00809E-09 | 0.25 | 0.28 | |
NPL | 2.01626E-09 | 0.27 | 0.29 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 1300 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 0.233086 | 0.07 | 0.14 | |
NIM | 0.232733 | 0.24 | 0.27 | |
KRISS | 0.233438 | 0.08 | 0.15 | |
NRC | 0.232796 | 0.35 | 0.37 | |
NIST | 0.232984 | 0.32 | 0.34 | |
LNE-Cnam | 0.233171 | 0.08 | 0.15 | |
PTB | 0.233136 | 0.24 | 0.27 | |
CEM | 0.233139 | 0.26 | 0.29 | |
NPL | 0.233182 | 0.30 | 0.33 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1500 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 1500 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 9.80554E-09 | 0.08 | 0.14 | |
NIM | 9.80629E-09 | 0.20 | 0.23 | |
KRISS | 9.80505E-09 | 0.14 | 0.18 | |
NRC | 9.79562E-09 | 0.43 | 0.44 | |
NIST | 9.80733E-09 | 0.37 | 0.38 | |
LNE-Cnam | 9.79790E-09 | 0.11 | 0.15 | |
PTB | 9.79782E-09 | 0.29 | 0.32 | |
CEM | 9.76223E-09 | 0.32 | 0.34 | |
NPL | 9.81852E-09 | 0.32 | 0.34 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 1500 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 1.134986 | 0.10 | 0.20 | |
NIM | 1.133486 | 0.28 | 0.34 | |
KRISS | 1.136242 | 0.13 | 0.22 | |
NRC | 1.133935 | 0.43 | 0.47 | |
NIST | 1.135770 | 0.37 | 0.41 | |
LNE-Cnam | 1.135364 | 0.11 | 0.21 | |
PTB | 1.137261 | 0.30 | 0.35 | |
CEM | 1.133876 | 0.32 | 0.37 | |
NPL | 1.135095 | 0.35 | 0.39 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1700 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 1700 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 3.46502E-08 | 0.11 | 0.17 | |
NIM | 3.46587E-08 | 0.24 | 0.28 | |
KRISS | 3.46465E-08 | 0.21 | 0.25 | |
NRC | 3.46243E-08 | 0.53 | 0.55 | |
NIST | 3.46361E-08 | 0.42 | 0.45 | |
LNE-Cnam | 3.46011E-08 | 0.14 | 0.19 | |
PTB | 3.45794E-08 | 0.36 | 0.39 | |
CEM | 3.44904E-08 | 0.39 | 0.42 | |
NPL | 3.46823E-08 | 0.40 | 0.42 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 1700 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 4.010009 | 0.14 | 0.24 | |
NIM | 4.005949 | 0.33 | 0.39 | |
KRISS | 4.014277 | 0.25 | 0.32 | |
NRC | 4.008684 | 0.53 | 0.56 | |
NIST | 4.008135 | 0.42 | 0.47 | |
LNE-Cnam | 4.011732 | 0.14 | 0.24 | |
PTB | 4.001250 | 0.36 | 0.41 | |
CEM | 4.004105 | 0.40 | 0.45 | |
NPL | 4.008140 | 0.42 | 0.46 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 1800 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 1800 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 5.94519E-08 | 0.13 | 0.21 | |
NIM | 5.94718E-08 | 0.26 | 0.31 | |
KRISS | 5.94351E-08 | 0.25 | 0.30 | |
NRC | 5.94185E-08 | 0.59 | 0.61 | |
NIST | 5.94024E-08 | 0.46 | 0.49 | |
LNE-Cnam | 5.93548E-08 | 0.15 | 0.22 | |
PTB | 5.93455E-08 | 0.40 | 0.44 | |
CEM | 5.90878E-08 | 0.44 | 0.47 | |
NPL | 5.94886E-08 | 0.46 | 0.49 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 1800 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 6.879800 | 0.16 | 0.27 | |
NIM | 6.874136 | 0.34 | 0.42 | |
KRISS | 6.886391 | 0.25 | 0.33 | |
NRC | 6.877912 | 0.59 | 0.62 | |
NIST | 6.870674 | 0.46 | 0.51 | |
LNE-Cnam | 6.878651 | 0.15 | 0.27 | |
PTB | 6.864309 | 0.33 | 0.40 | |
CEM | 6.861510 | 0.44 | 0.49 | |
NPL | 6.872750 | 0.46 | 0.51 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2000 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 2000 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 1.51782E-07 | 0.17 | 0.24 | |
NIM | 1.51832E-07 | 0.37 | 0.41 | |
KRISS | 1.51734E-07 | 0.34 | 0.39 | |
NRC | 1.51763E-07 | 0.72 | 0.74 | |
NIST | 1.51654E-07 | 0.53 | 0.56 | |
LNE-Cnam | 1.51517E-07 | 0.18 | 0.25 | |
PTB | 1.51449E-07 | 0.40 | 0.45 | |
CEM | 1.51250E-07 | 0.52 | 0.55 | |
NPL | 1.51978E-07 | 0.51 | 0.54 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 2000 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 1.748152 | 0.20 | 0.33 | |
NIM | 1.746161 | 0.47 | 0.55 | |
KRISS | 1.749278 | 0.34 | 0.43 | |
NRC | 1.748160 | 0.72 | 0.77 | |
NIST | 1.744942 | 0.53 | 0.59 | |
LNE-Cnam | 1.747312 | 0.18 | 0.31 | |
PTB | 1.742937 | 0.39 | 0.47 | |
CEM | 1.745093 | 0.52 | 0.59 | |
NPL | 1.747862 | 0.55 | 0.60 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2200 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 2200 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 3.33016E-07 | 0.21 | 0.30 | |
NIM | 3.33111E-07 | 0.41 | 0.47 | |
KRISS | 3.32893E-07 | 0.47 | 0.53 | |
NRC | 3.33113E-07 | 0.88 | 0.91 | |
NIST | 3.32619E-07 | 0.61 | 0.66 | |
LNE-Cnam | 3.32378E-07 | 0.22 | 0.31 | |
PTB | 3.32601E-07 | 0.47 | 0.54 | |
CEM | 3.31411E-07 | 0.73 | 0.76 | |
NPL | 3.33256E-07 | 0.62 | 0.65 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 2200 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 3.835241 | 0.26 | 0.39 | |
NIM | 3.835637 | 0.52 | 0.61 | |
KRISS | 3.837690 | 0.47 | 0.56 | |
NRC | 3.836444 | 0.88 | 0.93 | |
NIST | 3.824819 | 0.61 | 0.68 | |
LNE-Cnam | 3.832084 | 0.22 | 0.37 | |
PTB | 3.828337 | 0.47 | 0.56 | |
CEM | 3.836372 | 0.73 | 0.80 | |
NPL | 3.831870 | 0.65 | 0.71 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2400 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 2400 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 6.49658E-07 | 0.25 | 0.38 | |
NIM | 6.49890E-07 | 0.47 | 0.55 | |
KRISS | 6.49278E-07 | 0.56 | 0.64 | |
NRC | 6.49914E-07 | 1.06 | 1.10 | |
NIST | 6.48629E-07 | 0.71 | 0.76 | |
LNE-Cnam | 6.48153E-07 | 0.26 | 0.38 | |
PTB | 6.48145E-07 | 0.55 | 0.64 | |
CEM | 6.48478E-07 | 0.84 | 0.90 | |
NPL | 6.49917E-07 | 0.71 | 0.76 |
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 2400 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 7.481484 | 0.32 | 0.54 | |
NIM | 7.509864 | 0.59 | 0.75 | |
KRISS | 7.484104 | 0.56 | 0.71 | |
NRC | 7.484545 | 1.06 | 1.15 | |
NIST | 7.458087 | 0.71 | 0.83 | |
LNE-Cnam | 7.467360 | 0.25 | 0.51 | |
PTB | 7.461211 | 0.56 | 0.71 | |
CEM | 7.407262 | 0.85 | 0.96 | |
NPL | 7.473960 | 0.74 | 0.86 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2500 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 2500 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 8.75211E-07 | 0.28 | 0.43 | |
NIM | 8.75452E-07 | 0.49 | 0.60 | |
KRISS | 8.74053E-07 | 0.63 | 0.71 | |
NRC | 8.75697E-07 | 1.16 | 1.21 | |
NIST | 8.74159E-07 | 0.76 | 0.83 | |
LNE-Cnam | - | - | - | |
PTB | 8.72603E-07 | 0.60 | 0.70 | |
CEM | 8.73965E-07 | 0.89 | 0.96 | |
NPL | 8.75278E-07 | 0.76 | 0.83 |
'- Did not measure
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 2500 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 1.008692 | 0.35 | 0.56 | |
NIM | 1.015756 | 0.62 | 0.78 | |
KRISS | 1.008606 | 0.63 | 0.76 | |
NRC | 1.009021 | 1.16 | 1.24 | |
NIST | 1.005982 | 0.76 | 0.87 | |
LNE-Cnam | - | - | - | |
PTB | 1.005326 | 0.57 | 0.72 | |
CEM | 1.006790 | 0.89 | 1.00 | |
NPL | 1.007751 | 0.82 | 0.93 |
'- Did not measure
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2600 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 2600 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 1.15490E-06 | 0.30 | 0.46 | |
NIM | 1.15520E-06 | 0.66 | 0.75 | |
KRISS | 1.15334E-06 | 0.70 | 0.79 | |
NRC | 1.15600E-06 | 1.27 | 1.31 | |
NIST | 1.15336E-06 | 0.81 | 0.88 | |
LNE-Cnam | 1.15329E-06 | 0.32 | 0.47 | |
PTB | 1.15110E-06 | 0.61 | 0.72 | |
CEM | - | - | - | |
NPL | 1.15499E-06 | 0.82 | 0.89 |
'- Did not measure
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 2600 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 1.331014 | 0.38 | 0.57 | |
NIM | 1.343191 | 0.77 | 0.90 | |
KRISS | 1.330620 | 0.69 | 0.82 | |
NRC | 1.331247 | 1.26 | 1.33 | |
NIST | 1.326912 | 0.81 | 0.91 | |
LNE-Cnam | 1.328954 | 0.29 | 0.52 | |
PTB | 1.326265 | 0.60 | 0.74 | |
CEM | - | - | - | |
NPL | 1.330688 | 0.86 | 0.96 |
'- Did not measure
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2800 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 962 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 1.90520E-06 | 0.36 | 0.54 | |
NIM | 1.90539E-06 | 0.95 | 1.03 | |
KRISS | 1.90247E-06 | 0.84 | 0.94 | |
NRC | 1.90596E-06 | 1.50 | 1.55 | |
NIST | 1.90270E-06 | 0.92 | 1.00 | |
LNE-Cnam | 1.90290E-06 | 0.36 | 0.54 | |
PTB | 1.89782E-06 | 0.70 | 0.83 | |
CEM | - | - | - | |
NPL | 1.90680E-06 | 0.92 | 1.00 |
'- Did not measure
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 962 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 2.195663 | 0.45 | 0.73 | |
NIM | 2.219083 | 1.06 | 1.22 | |
KRISS | 2.195853 | 0.84 | 1.01 | |
NRC | 2.195996 | 1.49 | 1.60 | |
NIST | 2.193646 | 0.92 | 1.08 | |
LNE-Cnam | 2.192222 | 0.33 | 0.66 | |
PTB | 2.186457 | 0.70 | 0.90 | |
CEM | - | - | - | |
NPL | 2.202267 | 1.03 | 1.18 |
'- Did not measure
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 2900 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 2900 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 2.38995E-06 | 0.39 | 0.58 | |
NIM | 2.39309E-06 | 1.30 | 1.37 | |
KRISS | 2.38771E-06 | 0.96 | 1.06 | |
NRC | 2.39207E-06 | 1.60 | 1.66 | |
NIST | - | - | - | |
LNE-Cnam | - | - | - | |
PTB | 2.37990E-06 | 0.75 | 0.89 | |
CEM | - | - | - | |
NPL | 2.39202E-06 | 1.14 | 1.22 |
'- Did not measure
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 2900 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 2.754281 | 0.49 | 0.89 | |
NIM | 2.787050 | 1.40 | 1.59 | |
KRISS | 2.756414 | 0.96 | 1.21 | |
NRC | 2.757202 | 1.59 | 1.76 | |
NIST | - | - | - | |
LNE-Cnam | - | - | - | |
PTB | 2.742404 | 0.75 | 1.05 | |
CEM | - | - | - | |
NPL | 2.762893 | 1.40 | 1.59 |
'- Did not measure
CCT-K10
MEASURAND : Temperature
TEMPERATURE : 3000 °C
Transfer thermometer: LP3 thermometer
xi LP3 thermometer output photocurrent corresponding to the ITS-90 radiance temperature 3000 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
A | °C | °C | ||
NMIJ | 2.956931E-06 | 0.42 | 0.61 | |
NIM | - | - | - | |
KRISS | 2.952081E-06 | 1.00 | 1.10 | |
NRC | 2.958604E-06 | 1.73 | 1.78 | |
NIST | - | - | - | |
LNE-CNAM | - | - | - | |
PTB | - | - | - | |
CEM | - | - | - | |
NPL | 2.959621E-06 | 1.53 | 1.59 |
'- Did not measure
Transfer thermometer: Chino thermometer
xi Chino thermometer output voltage corresponding to the ITS-90 radiance temperature 3000 °C, submitted by laboratory i, Labi , following corrections according to Section 9.2 of the CCT-K10 Final report.
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Labi | xi | ui | ui-total | |
V | °C | °C | ||
NMIJ | 3.407657 | 0.53 | 0.82 | |
NIM | - | - | - | |
KRISS | 3.415004 | 1.00 | 1.18 | |
NRC | 3.417717 | 1.72 | 1.83 | |
NIST | - | - | - | |
LNE-Cnam | - | - | - | |
PTB | - | - | - | |
CEM | - | - | - | |
NPL | 3.420521 | 1.66 | 1.77 |
'- Did not measure
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1325 °C
xi measurement value submitted by laboratory i, Labi
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Co-C and Ni-C cells (linked)
Labi | xi | ui | ui-total | |
°C | °C | °C | ||
NMIJ | 1324.99 | 0.08 | 0.12 | |
NIM | 1324.76 | 0.12 | 0.14 | |
KRISS | 1324.84 | 0.09 | 0.12 | |
NRC | 1325.09 | 0.47 | 0.49 | |
NIST | 1324.49 | 0.34 | 0.37 | |
LNE-Cnam | 1324.71 | 0.06 | 0.17 | |
PTB | 1324.24 | 0.16 | 0.21 | |
CEM | 1324.39 | 0.20 | 0.24 | |
NPL | 1324.76 | 0.16 | 0.22 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 1952 °C
xi measurement value submitted by laboratory i, Labi
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
Ru-C cells
Labi | xi | ui | ui-total | |
°C | °C | °C | ||
NMIJ | 1953.31 | 0.18 | 0.23 | |
NIM | 1952.98 | 0.22 | 0.25 | |
KRISS | 1953.23 | 0.32 | 0.34 | |
NRC | 1952.88 | 0.90 | 0.91 | |
NIST | 1952.68 | 0.51 | 0.53 | |
LNE-Cnam | 1952.88 | 0.09 | 0.20 | |
PTB | 1952.85 | 0.32 | 0.34 | |
CEM | 1952.29 | 0.43 | 0.45 | |
NPL | 1952.96 | 0.33 | 0.35 |
CCT-K10
MEASURAND : Temperature
TEMPERATURE: 2747 °C
xi measurement value submitted by laboratory i, Labi
ui combined laboratory standard uncertainty of xi
ui-total combined standard uncertainty of xi including components for drift, corrections, and uncertainty of the median value.
WC-C cells
Labi | xi | ui | ui-total | |
°C | °C | °C | ||
NMIJ | 2747.42 | 0.40 | 0.45 | |
NIM | 2747.20 | 0.44 | 0.48 | |
KRISS | 2747.44 | 0.80 | 0.82 | |
NRC | 2747.03 | 1.73 | 1.74 | |
NIST | 2746.31 | 0.89 | 0.91 | |
LNE-Cnam | 2747.38 | 0.16 | 0.36 | |
PTB | - | - | - | |
CEM | 2747.59 | 0.82 | 0.84 | |
NPL | 2746.79 | 0.62 | 0.65 |