CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

10 mV 1 kHz
Di / (µV/V)
Participating institutes
CCEM-K11
CCEM-K11.1
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Ui
µV/V
µV/V
PTB -6 40
NPL -4 29
VSL -9 39
NMIA 0 11
INTI 13 25
NRC 8 35
A*STAR -1 33
NIM -41 78
SP -2 20
NIST 6 38
JV 5 31
INRIM -17 26
BEV 6 38
MKEH 1 26
IPQ 2 61
CEM -5 45
MIRS/SIQ 0 74
MIKES 13 33
Trescal -16 46
LNE 2 23
METAS -30 38
UME 5 51
NMISA -6 62
CMI 2 58
EIM 4 88
NPLI -7 26
VNIIM 2 29
BIM 7 66
Participating institutes
CCEM-K11
CCEM-K11.1
EUROMET.EM-K11
EURAMET.EM-K11.1
Comparison
Comparison close
CCEM-K11.1
Metrology area, Sub-field Electricity and Magnetism, AC Voltage, Current, Power, and AC/DC Transfer
Description AC/DC voltage transfer difference at low voltages
Time of measurements 2004 - 2005
Status Approved for equivalence
Final Reports of the comparisons
Measurand AC/DC transfer difference
Parameters Voltage: 10 mV and 100 mV
Frequency: 1 kHz to 1 MHz
Transfer device AC/DC transfer standard Fluke 792A
Comparison type Key Comparison
Consultative Committee CCEM (Consultative Committee for Electricity and Magnetism)
Conducted by CCEM (Consultative Committee for Electricity and Magnetism)
Comments

This is a bilateral follow-up of CCEM-K11. The same travelling standard and technical protocol are used.

CCEM-K11 and CCEM-K11.1 results published on 23 May 2007

EUROMET.EM-K11 results and SIM.EM-K11 results are linked to those of CCEM-K11 and CCEM-K11.1

AC/DC voltage transfer standards

Pilot institute RISE
Research Institutes of Sweden AB
Sweden
Contact person Karl-Erik Rydler

+46 105165401
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
RISE

Research Institutes of Sweden AB, Sweden, EURAMET

NIST

National Institute of Standards and Technology, United States, SIM

SP

Technical Research Institute of Sweden from 2017 Research Institutes of Sweden AB, Sweden, EURAMET

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = 1.6 µV/V                   2uR = 9.2 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1 and EURAMET.EM-K11.1, see final reports.