CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

10 mV 1 kHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     20 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

10 mV 20 kHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     100 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

10 mV 100 kHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 MHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

10 mV 1 MHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

100 mV 1 kHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     20 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

100 mV 20 kHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     100 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

100 mV 100 kHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 MHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

100 mV 1 MHz
Di / (µV/V)
Participating institutes

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -6 40
NPL -4 29
VSL -9 39
NMIA 0 11
INTI 13 25
NRC 8 35
A*STAR -1 33
NIM -41 78
SP -2 20
NIST 6 38
JV 5 31
INRIM -17 26
BEV 6 38
MKEH 1 26
IPQ 2 61
CEM -5 45
MIRS/SIQ 0 74
MIKES 13 33
Trescal -16 46
LNE 2 23
METAS -30 38
UME 5 51
NMISA -6 62
CMI 2 58
EIM 4 88
NPLI -7 26
VNIIM 2 29
BIM 7 66
Participating institutes
CCEM-K11
CCEM-K11.1
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     20 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -13 41
NPL -4 29
VSL -7 39
NMIA 2 14
INTI 11 28
NRC 2 31
A*STAR -2 36
NIM -28 78
SP -1 21
NIST 25 40
JV 11 32
INRIM -18 25
BEV 0 38
MKEH -3 26
IPQ -6 60
CEM -4 45
MIRS/SIQ -12 73
MIKES 8 32
Trescal -26 46
LNE 6 26
METAS -42 37
UME 7 51
NMISA 17 60
CMI 1 62
EIM -3 83
NPLI -8 25
VNIIM 20 30
BIM -2 69
Participating institutes
CCEM-K11
CCEM-K11.1
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     100 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -9 43
NPL 4 40
VSL -5 48
NMIA 7 39
INTI 13 44
NRC 7 59
A*STAR -9 42
NIM 1 95
SP -3 30
NIST -3 53
JV 19 73
INRIM -32 45
BEV -3 46
MKEH -9 37
IPQ -10 71
CEM -2 65
MIRS/SIQ -10 123
MIKES -1 45
Trescal -32 67
LNE -5 33
METAS -44 58
UME -19 60
NMISA 2 139
CMI 1 84
EIM 39 90
NPLI -22 30
VNIIM 62 69
BIM -10 79
Participating institutes
CCEM-K11
CCEM-K11.1
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 MHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -44 100
NPL -46 281
VSL 26 298
NMIA 3 102
INTI -29 105
NRC 42 84
A*STAR -42 369
NIM 138 184
SP -24 94
NIST 19 151
JV 8 269
INRIM -141 142
BEV -49 148
MKEH -172 233
IPQ 79 287
CEM 2 168
MIRS/SIQ/Metrology 31 754
MIKES -38 147
Trescal -126 207
METAS -151 129
UME -51 132
NMISA -579 758
CMI 40 243
EIM 126 190
NPLI -2 83
VNIIM 53 178
Participating institutes
CCEM-K11
CCEM-K11.1
EUROMET.EM-K11

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB 0.3 2.5
NPL -5.1 8.9
VSL -1.7 7.9
NMIA -0.1 5.7
INTI -0.1 9.9
NRC 0.7 4.8
A*STAR -3.3 18.0
NIM 3.7 13.9
SP 0.5 6.5
NIST -0.9 12.3
UTE 5.2 18.1
INMETRO -0.8 12.8
CENAM -1.3 25.1
JV 0.8 14.3
INRIM -10.0 10.0
BEV 2.4 12.7
MKEH 1.6 8.0
IPQ 9.5 41.2
CEM 3.4 10.8
MIRS/SIQ 4.4 30.3
MIKES 1.4 6.2
Trescal -8.7 8.5
LNE 2.6 7.5
METAS -11.1 13.3
UME 1.1 13.6
NMISA 6.9 18.0
CMI 2.3 19.4
EIM 1.6 5.6
NPLI -7.7 11.5
VNIIM 1.4 10.4
BIM 2.4 27.0
Participating institutes
CCEM-K11
CCEM-K11.1
SIM.EM-K11
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     20 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -0.3 4.3
NPL -3.2 8.7
VSL 0.9 11.8
NMIA 2.9 6.7
INTI 0.9 9.8
NRC -0.2 5.6
A*STAR -4.2 23.9
NIM -3.2 15.8
SP 0.3 6.5
NIST -2.8 12.7
UTE -1.1 18.2
INMETRO -1.1 12.5
CENAM -5.1 24.9
JV 0.3 15.6
INRIM -8.4 10.4
BEV 0.4 14.1
MKEH 0.5 9.9
IPQ 6.1 48.3
CEM -7.9 11.4
MIRS/SIQ 1.1 30.5
MIKES 0.2 7.8
Trescal -9.0 9.1
LNE 0.5 9.1
METAS -12.1 10.0
UME 5.0 15.1
NMISA 6.5 18.7
CMI 3.4 21.7
EIM 2.8 6.6
NPLI -1.8 12.2
VNIIM 18.2 14.1
BIM -0.6 27.7
Participating institutes
CCEM-K11
CCEM-K11.1
SIM.EM-K11
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     100 kHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -1.5 6.0
NPL -2.6 12.7
VSL -0.6 14.7
NMIA -1.1 15.8
INTI -0.2 20.8
NRC 3.5 6.6
A*STAR -5.8 26.9
NIM 8.1 29.9
SP -1.6 12.8
NIST 2.7 23.1
UTE -2.5 38.0
INMETRO -2.5 17.8
CENAM 1.1 26.0
JV 1.9 28.7
INRIM -19.2 13.7
BEV -2.6 18.9
MKEH -5.3 18.7
IPQ 6.6 61.5
CEM -11.4 31.1
MIRS/SIQ -7.5 50.7
MIKES -2.7 10.2
Trescal -14.6 14.5
LNE -4.5 15.4
METAS -16.0 15.4
UME -8.7 18.0
NMISA 9.9 35.8
CMI -4.2 30.1
EIM -4.9 9.4
NPLI -19.4 14.4
VNIIM 66.6 34.0
BIM -10.6 35.7
Participating institutes
CCEM-K11
CCEM-K11.1
SIM.EM-K11
EUROMET.EM-K11
EURAMET.EM-K11.1

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 MHz

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in µV/V.

LABi Di Uneg,i Upos,i
µV/V
µV/V
µV/V
PTB -28 50
NPL -18 67
VSL -1 101
NMIA 13 77
INTI -30 62
NRC 13 32
A*STAR -6 88
NIM 72 99
SP 7 63
NIST 1 77
UTE -34 111
INMETRO 19 105
CENAM 0 57
JV 11 111
INRIM -59 71
BEV -29 76
MKEH -44 155
IPQ -37 92
CEM -47 92
MIRS/SIQ -56 403
MIKES -31 56
Trescal -58 62
METAS -57 74
UME -23 61
NMISA 194 193
CMI -57 92
EIM -40 50
NPLI 106 58
VNIIM 229 127
Participating institutes
CCEM-K11
CCEM-K11.1
SIM.EM-K11
EUROMET.EM-K11
Comparison
Comparison close
EURAMET.EM-K11.1
Metrology area, Sub-field Electricity and Magnetism, AC Voltage, Current, Power, and AC/DC Transfer
Description AC/DC voltage transfer difference at low voltages
Time of measurements 2020
Status Approved for equivalence
Final Reports of the comparisons
Measurand AC-DC voltage transfer difference of traveling standard
Nominal voltages: 100 mV, 10 mV and 2 mV
Parameters Frequencies: 1 kHz, 20 kHz and 100 kHz
Transfer device AC-DC transfer standard Fluke 792A, provided by BEV
Comparison type Key Comparison
Consultative Committee CCEM (Consultative Committee for Electricity and Magnetism)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET project 1500
Comments Results published on 14 January 2022.
Pilot institute BEV
Bundesamt für Eich- und Vermessungswesen
Austria
Contact person Wolfgang Waldmann

+43 1 21110 826350
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BEV

Bundesamt für Eich- und Vermessungswesen, Austria, EURAMET

BIM

Bulgarian Institute of Metrology, Bulgaria, EURAMET

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = 1.6 µV/V                   2uR = 9.2 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     20 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = 1.0 µV/V                   2uR = 10.4 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     100 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = 2.8 µV/V                   2uR = 15.8 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 MHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = 24 µV/V                   2uR = 49 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = -0.5 µV/V                   2uR = 2.3 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking SIM.EM-K11 to CCEM-K11

INTI (Argentina) and NRC (Canada) provide the link between the key comparison SIM.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     20 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = -0.3 µV/V                   2uR = 2.9 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking SIM.EM-K11 to CCEM-K11

INTI (Argentina) and NRC (Canada) provide the link between the key comparison SIM.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     100 kHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = 1.6 µV/V                   2uR = 4.5 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking SIM.EM-K11 to CCEM-K11

INTI (Argentina) and NRC (Canada) provide the link between the key comparison SIM.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 MHz

 

The key comparison reference value xR is calculated as the weighted mean of the results of all participants in CCEM-K11. The value of the mean of the seven results of SP is used for the computation of xR, see equations 10 and 11 of the Final Report. The standard uncertainty uR of xR is computed as explained in section 6.1.4 of the Final Report.

                                xR = -7 µV/V                   2uR = 27 µV/V

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in µV/V:

 

Di = (xi - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 ui.

 

 

Linking CCEM-K11.1 to CCEM-K11

SP (Sweden) provides the link between the key comparison CCEM-K11.1 and CCEM-K11, having participated in both comparisons.

 

Linking EUROMET.EM-K11 to CCEM-K11

SP (Sweden), PTB (Germany) and VSL (Netherlands) provide the link between the key comparison EUROMET.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking SIM.EM-K11 to CCEM-K11

INTI (Argentina) and NRC (Canada) provide the link between the key comparison SIM.EM-K11 and CCEM-K11, having participated in both comparisons.

 

Linking EURAMET.EM-K11.1 to CCEM-K11

BEV (Austria) provides the link between the key comparison EURAMET.EM-K11.1 and CCEM-K11, via their participation in EUROMET.EM-K11.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     20 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     100 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      10 mV

FREQUENCY                     1 MHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1, SIM.EM-K11 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     20 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1, SIM.EM-K11 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     100 kHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1, SIM.EM-K11 and EURAMET.EM-K11.1, see final reports.

CCEM-K11, CCEM-K11.1, EUROMET.EM-K11, SIM.EM-K11 and EURAMET.EM-K11.1

MEASURAND                    AC/DC voltage transfer difference

NOMINAL VOLTAGE      100 mV

FREQUENCY                     1 MHz

 

xi          measurement result as obtained by laboratory iLabi

ULabi  expanded uncertainty of xi at 95 % level of confidence (= 2)

 

CCEM-K11

 

 

CCEM-K11.1

 

 

For initial results obtained in EUOMET.EM-K1, SIM.EM-K11 and EURAMET.EM-K11.1, see final reports.