CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

10 kN.m clockwise
Di / (N.m)
Participating institutes

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = -18.4 N.m and UCENAM = 1.5 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m anti-clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

10 kN.m anti-clockwise
Di / (N.m)
Participating institutes

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = 13.0 N.m and UCENAM = 0.7 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

20 kN.m clockwise
Di / (N.m)
Participating institutes

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = -12.7 N.m and UCENAM = 1.3 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m anti-clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

20 kN.m anti-clockwise
Di / (N.m)
Participating institutes

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = 4.5 N.m and UCENAM = 1.2 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

LABi Di Uneg,i Upos,i
N.m
N.m
N.m
NMIJ 0.4 0.6
MIKES -0.4 5.2
NIM -1.5 3.4
PTB -0.2 0.3
NMISA -0.4 3.2
Participating institutes
CCM.T-K2
CCM.T-K2.1

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = -18.4 N.m and UCENAM = 1.5 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m anti-clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

LABi Di Uneg,i Upos,i
N.m
N.m
N.m
NMIJ -0.4 0.6
MIKES -0.1 5.2
NIM 1.2 3.4
PTB 0.2 0.3
NMISA 0.3 3.2
Participating institutes
CCM.T-K2
CCM.T-K2.1

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = 13.0 N.m and UCENAM = 0.7 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

LABi Di Uneg,i Upos,i
N.m
N.m
N.m
NMIJ 0.9 1.2
MIKES -1.0 10.5
NIM -4.3 6.8
PTB -0.3 0.5
NMISA -3.7 6.3
Participating institutes
CCM.T-K2
CCM.T-K2.1

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = -12.7 N.m and UCENAM = 1.3 N.m

CCM.T-K2 and CCM.T-K2.1

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m anti-clockwise torque

 

Degrees of equivalence represented by Di  and its expanded uncertainty Ui (k = 2) at a 95 % level of confidence, both expressed in N·m.

LABi Di Uneg,i Upos,i
N.m
N.m
N.m
NMIJ -0.7 1.2
VTT -0.1 10.6
NIM 2.6 6.8
PTB 0.4 0.5
NMISA 4.4 6.3
Participating institutes
CCM.T-K2
CCM.T-K2.1

CENAM results were not included in the calculation of the KCRV (see page 2 of the CCM.T-K2 Final report)

where DCENAM = 4.5 N.m and UCENAM = 1.2 N.m

Comparison
Comparison close
CC comparison
Linked comparison
CCM.T-K2.1
Metrology area, Sub-field Mass and related quantities, Torque
Description Torque Measurements
Time of measurements 2021 - 2022
Status Approved for equivalence
Final Reports of the comparisons
References

Protocol

Measurand Torque
10 kN m, 20 kN m.
Parameters Temperature and relative humidity
Transfer device One torque transducer: TB2 / 10 kN m and a cable, connected to each amplifier, DMP40/41.
One bridge calibrator: BN100A.
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by CCM (Consultative Committee for Mass and Related Quantities)
Comments Results published 16 July 2024
Pilot institute NMIJ AIST
National Metrology Institute of Japan
Japan
Contact person Koji Ogushi

+81 29 861 4012
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

CCM.T-K2 and CCM.T-K2.1

 

CCM.T-K2

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m clockwise torque

 

Two different torque transducers, TB2 and TT1, were used for the comparison. The Key Comparison Reference Values xR and corresponding uncertainties uR were determined for each transducer and are listed below. However, the Degrees of Equivalence were calculated based on the mean value of the results obtained using both transducers.

Transducer xR / kN·m uR / N·m
TB2 clockwise 10.000 0.2
TT1 clockwise 10.000 0.3

 

For each torque the degree of equivalence of laboratory i with respect to the key comparison reference value xR is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in N·m. The computation of Di and Ui is explained in the CCM.T-K2 Final Report.

 

Linking CCM.T-K2.1 to CCM.T-K2

NMIJ provides the link between the key comparison CCM.T-K2.1 and CCM.T-K2, having participated in both comparisons. The linking process is described in Section 4.2 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

CCM.T-K2

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m anti-clockwise torque

 

Two different torque transducers, TB2 and TT1, were used for the comparison. The Key Comparison Reference Values xR and corresponding uncertainties uR were determined for each transducer and are listed below. However, the Degrees of Equivalence were calculated based on the mean value of the results obtained using both transducers.

Transducer xR / kN·m uR / N·m
TB2 anti-clockwise -10.000 0.2
TT1 anti-clockwise -10.000 0.3

 

For each torque the degree of equivalence of laboratory i with respect to the key comparison reference value xR is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in N·m. The computation of Di and Ui is explained in the CCM.T-K2 Final Report.

 

Linking CCM.T-K2.1 to CCM.T-K2

NMIJ provides the link between the key comparison CCM.T-K2.1 and CCM.T-K2, having participated in both comparisons. The linking process is described in Section 4.2 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

CCM.T-K2

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m clockwise torque

 

Two different torque transducers, TB2 and TT1, were used for the comparison. The Key Comparison Reference Values xR and corresponding uncertainties uR were determined for each transducer and are listed below. However, the Degrees of Equivalence were calculated based on the mean value of the results obtained using both transducers.

Transducer xR / kN·m uR / N·m
TB2 clockwise 20.000 0.3
TT1 clockwise 20.000 0.6

 

For each torque the degree of equivalence of laboratory i with respect to the key comparison reference value xR is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in N·m. The computation of Di and Ui is explained in the CCM.T-K2 Final Report.

 

Linking CCM.T-K2.1 to CCM.T-K2

NMIJ provides the link between the key comparison CCM.T-K2.1 and CCM.T-K2, having participated in both comparisons. The linking process is described in Section 4.2 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

CCM.T-K2

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m anti-clockwise torque

 

Two different torque transducers, TB2 and TT1, were used for the comparison. The Key Comparison Reference Values xR and corresponding uncertainties uR were determined for each transducer and are listed below. However, the Degrees of Equivalence were calculated based on the mean value of the results obtained using both transducers.

Transducer xR / kN·m uR / N·m
TB2 anti-clockwise -20.000 0.3
TT1 anti-clockwise -20.000 0.6

 

For each torque the degree of equivalence of laboratory i with respect to the key comparison reference value xR is given by a pair of terms: Di and its expanded uncertainty Ui (k = 2) both expressed in N·m. The computation of Di and Ui is explained in the CCM.T-K2 Final Report.

 

Linking CCM.T-K2.1 to CCM.T-K2

NMIJ provides the link between the key comparison CCM.T-K2.1 and CCM.T-K2, having participated in both comparisons. The linking process is described in Section 4.2 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m clockwise torque

 

CCM.T-K2

Two torque transducers were used as transfer standards for measurements at each nominal torque, clockwise and anti-clockwise: TB2 (SN 112-330-004) and TT1 (SN 37365-04). The individual measurement results are available on pp. 3 to 12 of the CCM.T-K2 Final Report.

 

CCM.T-K2.1

Measurement results are available in Section 3 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

MEASURAND: Torque

NOMINAL VALUE: 10 kN.m anti-clockwise torque

 

CCM.T-K2

Two torque transducers were used as transfer standards for measurements at each nominal torque, clockwise and anti-clockwise: TB2 (SN 112-330-004) and TT1 (SN 37365-04). The individual measurement results are available on pp. 3 to 12 of the CCM.T-K2 Final Report.

 

CCM.T-K2.1

Measurement results are available in Section 3 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m clockwise torque

 

CCM.T-K2

Two torque transducers were used as transfer standards for measurements at each nominal torque, clockwise and anti-clockwise: TB2 (SN 112-330-004) and TT1 (SN 37365-04). The individual measurement results are available on pp. 3 to 12 of the CCM.T-K2 Final Report.

 

CCM.T-K2.1

Measurement results are available in Section 3 of the CCM.T-K2.1 Final report.

CCM.T-K2 and CCM.T-K2.1

 

MEASURAND: Torque

NOMINAL VALUE: 20 kN.m anti-clockwise torque

 

CCM.T-K2

Two torque transducers were used as transfer standards for measurements at each nominal torque, clockwise and anti-clockwise: TB2 (SN 112-330-004) and TT1 (SN 37365-04). The individual measurement results are available on pp. 3 to 12 of the CCM.T-K2 Final Report.

 

CCM.T-K2.1

Measurement results are available in Section 3 of the CCM.T-K2.1 Final report.